Optimization of a centrifugal blood pump designed using an industrial method through experimental and numerical study

被引:1
作者
Yazdanpanah-Ardakani, Kohyar [1 ]
Niroomand-Oscuii, Hanieh [1 ]
Kanan, Reza Sahebi-Kuzeh [1 ]
Shokri, Nasim [1 ]
机构
[1] Sahand Univ Technol, Dept Biomed Engn, Tabriz, Iran
关键词
Ventricular assist device (VAD); Point-by-point method; Centrifugal pump; Blade profile; VENTRICULAR ASSIST DEVICE; COMPUTATIONAL FLUID-DYNAMICS; BLADE OUTLET ANGLE; INTERAGENCY REGISTRY; CIRCULATORY SUPPORT; FLOW; PERFORMANCE; HEMOLYSIS; MODELS; PROTOTYPE;
D O I
10.1038/s41598-024-57019-9
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
With improved treatment of coronary artery disease, more patients are surviving until heart failure occurs. This leads to an increase in patients needing devices for struggling with heart failure. Ventricular assist devices are known as the mainstay of these devices. This study aimed to design a centrifugal pump as a ventricular assist device. In order to design the pump, firstly, the geometrical parameters of the pump, including the gap distance, blade height, and position of the outlet relative to the blade, were investigated. Finally, the selected configuration, which had all the appropriate characteristics, both hydraulically and physiologically, was used for the rest of the study. The study of the blade, as the main component in energy transfer to the blood, in a centrifugal pump, has been considered in the present study. In this regard, the point-to-point design method, which is used in industrial applications, was implemented. The designer chooses the relationship between the blade angles at each radius in the point-to-point method. The present study selected logarithmic and second-order relations for designing the blade's profile. In total, 58 blades were examined in this study, which differed regarding blade inlet and outlet angles and the relationship between angle and radial position. ANSYS CFX 17.0 software was utilized to simulate blades' performances, and a benchmark pump provided by the US Food and Drug Administration (FDA) was used to validate the numerical simulations. Then, the selected impeller from the numerical investigation was manufactured, and its performance was compared experimentally with the FDA benchmark pump. A hydraulic test rig was also developed for experimental studies. The results showed that among the blades designed in this study, the blade with an input angle of 45 degrees and an output angle of 55 degrees, which is designed to implement a logarithmic relationship, has the best performance. The selected impeller configuration can increase the total head (at least by 20%) at different flow rates compared to the FDA pump.
引用
收藏
页数:17
相关论文
共 70 条
[21]   Impeller (straight blade) design variations and their influence on the performance of a centrifugal blood pump [J].
Fang, Peng ;
Du, Jianjun ;
Yu, Shunzhou .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2020, 43 (12) :782-795
[22]   Hybrid Continuous-Flow Total Artificial Heart [J].
Fox, Carson ;
Chopski, Steven ;
Murad, Nohra ;
Allaire, Paul ;
Mentzer, Robert, Jr. ;
Rossano, Joseph ;
Arabia, Francisco ;
Throckmorton, Amy .
ARTIFICIAL ORGANS, 2018, 42 (05) :500-509
[23]   The use of computational fluid dynamics in the development of ventricular assist devices [J].
Fraser, Katharine H. ;
Taskin, M. Ertan ;
Griffith, Bartley P. ;
Wu, Zhongjun J. .
MEDICAL ENGINEERING & PHYSICS, 2011, 33 (03) :263-280
[24]   Incidence and Patterns of Adverse Event Onset During the First 60 Days After Ventricular Assist Device Implantation [J].
Genovese, Elizabeth A. ;
Dew, Mary Amanda ;
Teuteberg, Jeffrey J. ;
Simon, Marc A. ;
Kay, Joy ;
Siegenthaler, Michael P. ;
Bhama, Jay K. ;
Bermudez, Christian A. ;
Lockard, Kathleen L. ;
Winowich, Steve ;
Kormos, Robert L. .
ANNALS OF THORACIC SURGERY, 2009, 88 (04) :1162-1170
[25]   Shape optimization of a centrifugal blood pump by coupling CFD with metamodel-assisted genetic algorithm [J].
Ghadimi, Behnam ;
Nejat, Amir ;
Nourbakhsh, Seyed Ahmad ;
Naderi, Nasim .
JOURNAL OF ARTIFICIAL ORGANS, 2019, 22 (01) :29-36
[26]   Multi-Objective Genetic Algorithm Assisted by an Artificial Neural Network Metamodel for Shape Optimization of a Centrifugal Blood Pump [J].
Ghadimi, Behnam ;
Nejat, Amir ;
Nourbakhsh, Seyed Ahmad ;
Naderi, Nasim .
ARTIFICIAL ORGANS, 2019, 43 (05) :E76-E93
[27]  
Gregory S., 2017, Mechanical Circulatory and Respiratory Support, P25
[28]   Early and mid-term outcomes of left ventricular assist device implantation and future prospects [J].
Hata H. ;
Fujita T. ;
Shimahara Y. ;
Sato S. ;
Yanase M. ;
Seguchi O. ;
Sato T. ;
Nakatani T. ;
Kobayashi J. .
General Thoracic and Cardiovascular Surgery, 2015, 63 (10) :557-564
[29]   Numerical study of the effect of LVAD inflow cannula positioning on thrombosis risk [J].
He, Ge ;
Han, Lu ;
Zhang, Jiafeng ;
Shah, Aakash ;
Kaczorowski, David J. ;
Griffith, Bartley P. ;
Wu, Zhongjun .
COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2022, 25 (08) :852-860
[30]   Durability of left ventricular assist devices: Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS) 2006 to 2011 [J].
Holman, William L. ;
Naftel, David C. ;
Eckert, Chad E. ;
Kormos, Robert L. ;
Goldstein, Daniel J. ;
Kirklin, James K. .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2013, 146 (02) :437-+