Influence of 320-detector-row volume scanning and AAPM report 111 CT dosimetry metrics on size-specific dose estimate: a Monte Carlo study

被引:0
作者
Tomonobu Haba
Shuji Koyama
Yutaka Kinomura
Yoshihiro Ida
Masanao Kobayashi
机构
[1] Fujita Health University Hospital,Department of Radiology
[2] Nagoya University,Brain & Mind Research Center
[3] Fujita Health University,Faculty of Radiological Technology
来源
Australasian Physical & Engineering Sciences in Medicine | 2016年 / 39卷
关键词
Size-specific dose estimate; 320-detector-row volume CT; AAPM task group 111; Monte Carlo simulation;
D O I
暂无
中图分类号
学科分类号
摘要
The American Association of Physicists in Medicine (AAPM) task group 204 has recommended the use of size-dependent conversion factors to calculate size-specific dose estimate (SSDE) values from volume computed tomography dose index (CTDIvol) values. However, these conversion factors do not consider the effects of 320-detector-row volume computed tomography (CT) examinations or the new CT dosimetry metrics proposed by AAPM task group 111. This study aims to investigate the influence of these examinations and metrics on the conversion factors reported by AAPM task group 204, using Monte Carlo simulations. Simulations were performed modelling a Toshiba Aquilion ONE CT scanner, in order to compute dose values in water for cylindrical phantoms with 8–40-cm diameters at 2-cm intervals for each scanning parameter (tube voltage, bow-tie filter, longitudinal beam width). Then, the conversion factors were obtained by applying exponential regression analysis between the dose values for a given phantom diameter and the phantom diameter combined with various scanning parameters. The conversion factors for each scanning method (helical, axial, or volume scanning) and CT dosimetry method (i.e., the CTDI100 method or the AAPM task group 111 method) were in agreement with those reported by AAPM task group 204, within a percentage error of 14.2 % for phantom diameters ≥11.2 cm. The results obtained in this study indicate that the conversion factors previously presented by AAPM task group 204 can be used to provide appropriate SSDE values for 320-detector-row volume CT examinations and the CT dosimetry metrics proposed by the AAPM task group 111.
引用
收藏
页码:697 / 703
页数:6
相关论文
共 61 条
[11]  
Brady SL(2003)A new look at CT dose measurement: beyond CTDI Med Phys 30 1272-3413
[12]  
Mirro AE(2007)Experimental validation of a versatile system of CT dosimetry using a conventional ion chamber: beyond CTDI Med Phys 34 3399-6495
[13]  
Kaufman RA(2007)Determination of the weighted CT dose index in modern multi-detector CT scanners Phys Med Biol 52 6485-2164
[14]  
Silverman JD(2009)A method to generate equivalent energy spectra and filtration models based on measurement for multidetector CT monte carlo dosimetry simulations Med Phys 36 2154-2597
[15]  
Paul NS(2007)Estimating radiation doses from multidetector CT using monte carlo simulations: effects of different size voxelized patient models on magnitudes of organ and effective dose Phys Med Biol 52 2583-1038
[16]  
Siewerdsen JH(2009)Variablity of surface and center position radiation dose in MDCT: monte carlo simulation using CTDI and anthropomorphic phantoms Med Phys 36 1025-undefined
[17]  
Kroft LJ(undefined)undefined undefined undefined undefined-undefined
[18]  
Roelofs JJ(undefined)undefined undefined undefined undefined-undefined
[19]  
Geleijns J(undefined)undefined undefined undefined undefined-undefined
[20]  
Orrison WW(undefined)undefined undefined undefined undefined-undefined