320-row CT coronary angiography: effect of 100-kV tube voltages on image quality, contrast volume, and radiation dose

被引:0
|
作者
Chuanchen Zhang
Zhaoqi Zhang
Zixu Yan
Lei Xu
Wei Yu
Rui Wang
机构
[1] Capital Medical University,Department of Radiology, Beijing Anzhen Hospital
来源
The International Journal of Cardiovascular Imaging | 2011年 / 27卷
关键词
320-Row computed tomography; Contrast material; Coronary angiography; Coronary vessels; Radiation dosage;
D O I
暂无
中图分类号
学科分类号
摘要
To prospectively evaluate image quality parameters, contrast volume and radiation dose at the 100-kilovolt (kV) setting during coronary computed tomographic angiography (CCTA) on a 320-row computed tomography scanner. We enrolled 107 consecutive patients with a heart rate <65 beats per minute (bpm) undergoing prospective electrocardiogram (ECG)-triggered CCTA. Forty patients with a body mass index (BMI) <25 kg/m2 were scanned using 100-kV tube voltage settings, while 67 patients were scanned using 120-kV protocols. Image quality was assessed by two readers unaware of patient information and scan parameters. Attenuation in the aorta and perivascular fat tissue and image noise were measured. Contrast-to-noise ratios (CNRs) and contrast material volumes were calculated. The effective radiation doses were estimated using a chest conversion coefficient (0.017). Diagnostic image quality was achieved in 98.2% of coronary segments with 100-kV CCTA and 98.6% of coronary segments with 120-kV CCTA, with no significant differences in image quality scores for each coronary segment. Vessel attenuation, image noise, and CNR were not significantly different between the 100- and 120-kV protocols. Mean contrast injection rate and mean material volume were significantly lower for the 100-kV CCTA (4.35 ± 0.28 ml/s and 53.13 ± 3.77 ml, respectively) than for the 120-kV CCTA (5.16 ± 0.21 ml/s and 62.40 ± 3.66 ml respectively; P < 0.001). The effective radiation dose was 2.12 ± 0.19 mSv for 100-kV CCTA, a reduction of 54% compared to 4.61 ± 0.82 mSv for 120-kV CCTA. A 100-kV CCTA can be implemented in patients with a BMI < 25 kg/m2. The 100-kV setting allows significant reductions in contrast material volume and effective radiation dose while maintaining adequate diagnostic image quality.
引用
收藏
页码:1059 / 1068
页数:9
相关论文
共 36 条
  • [31] Single-versus multi-phase acquisition protocol for prospective-triggered sequential dual-source CT coronary angiography: comparison of image quality and radiation dose
    Huang, Wei
    Xu, Yiming
    Lu, Daoyan
    Shi, Yuzhen
    Lu, Guangming
    CLINICAL IMAGING, 2015, 39 (04) : 597 - 602
  • [32] Retrospective Gating vs. Prospective Triggering for Noninvasive Coronary Angiography: Assessment of Image Quality and Radiation Dose Using a 256-Slice CT Scanner with 270 ms Gantry Rotation
    Law, Wei-Yip
    Yang, Ching-Ching
    Chen, Liang-Kuang
    Huang, Tzung-Chi
    Lu, Kun-Mu
    Wu, Tung-Hsin
    Mok, Greta S. P.
    ACADEMIC RADIOLOGY, 2011, 18 (01) : 31 - 39
  • [33] Adenosine-stress dynamic myocardial perfusion imaging using 128-slice dual-source CT in patients with normal body mass indices: effect of tube voltage, tube current, and iodine concentration on image quality and radiation dose
    Kim, Sung Mok
    Cho, Young Kwon
    Choe, Yeon Hyeon
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2014, 30 : 95 - 103
  • [34] The effect of tube voltage combination on image artefact and radiation dose in dual-source dual-energy CT: comparison between conventional 80/140 kV and 80/150 kV plus tin filter for gout protocol
    Ji Young Jeon
    Sheen-Woo Lee
    Yu Mi Jeong
    Han Joo Baek
    European Radiology, 2019, 29 : 1248 - 1257
  • [35] The effect of tube voltage combination on image artefact and radiation dose in dual-source dual-energy CT: comparison between conventional 80/140 kV and 80/150 kV plus tin filter for gout protocol
    Jeon, Ji Young
    Lee, Sheen-Woo
    Jeong, Yu Mi
    Baek, Han Joo
    EUROPEAN RADIOLOGY, 2019, 29 (03) : 1248 - 1257
  • [36] Adenosine-stress dynamic myocardial perfusion imaging using 128-slice dual-source CT in patients with normal body mass indices: effect of tube voltage, tube current, and iodine concentration on image quality and radiation dose
    Sung Mok Kim
    Young Kwon Cho
    Yeon Hyeon Choe
    The International Journal of Cardiovascular Imaging, 2014, 30 : 95 - 103