Application of the Bjerrum function to the potentiometric tritration of aluminum by fluoride ions

被引:0
作者
V. I. Golovanov
机构
[1] GUOVPO Chelyabinsk State University,
来源
Journal of Analytical Chemistry | 2006年 / 61卷
关键词
Aluminum; Analytical Chemistry; Titration; Fluoride; Systematic Error;
D O I
暂无
中图分类号
学科分类号
摘要
The theoretical grounds and analytical characteristics of two analytical methods, titration to a given potential and the linearized graph method, are compared using the titration of aluminum by fluoride ions in the presence of a LaF3 electrode as an example. It is shown that there is no fundamental difference between these two methods. The linearized graph method is simpler and more accurate than the titration to a given potential. An equation is proposed for processing the results obtained using the linearized graph. This equation eliminates systematic errors due to the retarded kinetics of complexation.
引用
收藏
页码:369 / 374
页数:5
相关论文
empty
未找到相关数据