Investigation of the hydrodynamic properties of a new MRI-resistant programmable hydrocephalus shunt

被引:6
作者
Allin D.M. [1 ]
Czosnyka M. [1 ]
Richards H.K. [1 ]
Pickard J.D. [1 ]
Czosnyka Z.H. [1 ]
机构
[1] Shunt Evaluation Laboratory, Academic Neurosurgical Unit, Addenbrooke's Hospital, Cambridge CB2 2QQ, Hills Road
来源
Cerebrospinal Fluid Research | / 5卷 / 1期
基金
英国医学研究理事会;
关键词
Hydrodynamic Property; Hydrodynamic Resistance; Closing Pressure; Distal Catheter; Programmable Valve;
D O I
10.1186/1743-8454-5-8
中图分类号
学科分类号
摘要
Background: The Polaris valve is a newly released hydrocephalus shunt that is designed to drain cerebrospinal fluid (CSF) from the brain ventricles or lumbar CSF space. The aim of this study was to bench test the properties of the Polaris shunt, independently of the manufacturer. Methods: The Polaris Valve is a ball-on-spring valve, which can be adjusted magnetically in vivo. A special mechanism is incorporated to prevent accidental re-adjustment by an external magnetic field. The performance and hydrodynamic properties of the valve were evaluated in the UK Shunt Evaluation Laboratory, Cambridge, UK. Results: The three shunts tested showed good mechanical durability over the 3-month period of testing, and a stable hydrodynamic performance over 45 days. The pressure-flow performance curves, operating, opening and closing pressures were stable. The drainage rate of the shunt increased when a negative outlet pressure (siphoning) was applied. The hydrodynamic parameters fell within the limits specified by the manufacturer and changed according to the five programmed performance levels. Hydrodynamic resistance was dependant on operating pressure, changing from low values of 1.6 mmHg/ml/min at the lowest level to 11.2 mmHg/ml/min at the highest performance level. External programming proved to be easy and reliable. Even very strong magnetic fields (3 Tesla) were not able to change the programming of the valve. However, distortion of magnetic resonance images was present. Conclusion: The Polaris Valve is a reliable, adjustable valve. Unlike other adjustable valves (except the Miethke ProGAV valve), the Polaris cannot be accidentally re-adjusted by an external magnetic field. © 2008 Allin et al; licensee BioMed Central Ltd.
引用
收藏
相关论文
共 58 条
[21]  
Jandial R(undefined)undefined undefined undefined undefined-undefined
[22]  
Aryan HE(undefined)undefined undefined undefined undefined-undefined
[23]  
Hughes SA(undefined)undefined undefined undefined undefined-undefined
[24]  
Levy M(undefined)undefined undefined undefined undefined-undefined
[25]  
Schneider T(undefined)undefined undefined undefined undefined-undefined
[26]  
Knauff U(undefined)undefined undefined undefined undefined-undefined
[27]  
Nitsch J(undefined)undefined undefined undefined undefined-undefined
[28]  
Firsching R(undefined)undefined undefined undefined undefined-undefined
[29]  
Inoue T(undefined)undefined undefined undefined undefined-undefined
[30]  
Kuzu Y(undefined)undefined undefined undefined undefined-undefined