Investigation of thermal and temporal responses of ionization chambers in radiation dosimetry

被引:3
作者
Hussein AlMasri
Akira Funyu
Yasumasa Kakinohana
Sadayuki Murayama
机构
[1] Department of Radiology, Graduate School of Medical Science, University of the Ryukyu, Nishihara-cho, Okinawa 903-0215
关键词
Dosimetry; Environmental conditions; Ion chambers; Temporal response; Thermal response;
D O I
10.1007/s12194-012-0151-8
中图分类号
学科分类号
摘要
The ionization chamber is a primary dosimeter that is used in radiation dosimetry. Generally, the ion chamber response requires temperature/pressure correction according to the ideal gas law. However, this correction does not consider the thermal volume effect of chambers. The temporal and thermal volume effects of various chambers (CC01, CC13, NACP parallel-plate, PTW) with different wall and electrode materials have been studied in a water phantom. Measurements were done after heating the water with a suitable heating system, and chambers were submerged for a sufficient time to allow for temperature equilibrium. Temporal results show that all chambers equilibrate quickly in water. The equilibration time was between 3 and 5 min for all chambers. Thermal results show that all chambers expanded in response to heating except for the PTW, which contracted. This might be explained by the differences in the volumes of all chambers and also by the difference in wall material composition of PTW from the other chambers. It was found that the smallest chamber, CC01, showed the greatest expansion. The magnitude of the expansion was ̃1, 0.8, and 0.9 % for CC01, CC13, and parallel-plate chambers, respectively, in the temperature range of 295-320 K. The magnitude of the detected contraction was<0.3 % for PTW in the same temperature range. For absolute dosimetry, it is necessary to make corrections for the ion chamber response, especially for small ion chambers like the CC01. Otherwise, room and water phantom temperatures should remain within a close range. © Japanese Society of Radiological Technology and Japan Society of Medical Physics 2012.
引用
收藏
页码:172 / 177
页数:5
相关论文
共 35 条
  • [1] Investigation of thermal and temporal responses of ionization chambers in radiation dosimetry
    AlMasri, Hussein
    Funyu, Akira
    Kakinohana, Yasumasa
    Murayama, Sadayuki
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2012, 5 (02) : 172 - 177
  • [2] Thermal and temporal response of ionization chambers in radiation dosimetry
    Das, IJ
    Zhu, TC
    MEDICAL PHYSICS, 2004, 31 (03) : 573 - 578
  • [3] Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties
    Solimanian, A.
    Ghafoori, M.
    IRANIAN JOURNAL OF RADIATION RESEARCH, 2010, 8 (03): : 195 - 199
  • [4] Development of novel ionization chambers for reference dosimetry in electron flash radiotherapy
    Liu, Kevin
    Holmes, Shannon
    Khan, Ahtesham Ullah
    Hooten, Brian
    DeWerd, Larry
    Schuler, Emil
    Beddar, Sam
    MEDICAL PHYSICS, 2024, : 9275 - 9289
  • [5] An evaluation of ionization chambers for the relative dosimetry of kilovoltage x-ray beams
    Hill, Robin
    Mo, Zhao
    Haque, Mamoon
    Baldock, Clive
    MEDICAL PHYSICS, 2009, 36 (09) : 3971 - 3981
  • [6] Measurement methods of ionization current and electric charges in radiation dosimetry
    Knyziak, Adrian Bozydar
    Rzodkiewicz, Witold
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2016, 822 : 1 - 8
  • [7] A hybrid radiation detector for simultaneous spatial and temporal dosimetry
    C. Poole
    J. V. Trapp
    J. Kenny
    T. Kairn
    K. Williams
    M. Taylor
    R. Franich
    C. M. Langton
    Australasian Physical & Engineering Sciences in Medicine, 2011, 34 : 327 - 332
  • [8] A hybrid radiation detector for simultaneous spatial and temporal dosimetry
    Poole, C.
    Trapp, J. V.
    Kenny, J.
    Kairn, T.
    Williams, K.
    Taylor, M.
    Franich, R.
    Langton, C. M.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2011, 34 (03) : 327 - 332
  • [9] Absolute dosimetry on a dynamically scanned sample for synchrotron radiotherapy using graphite calorimetry and ionization chambers
    Lye, J. E.
    Harty, P. D.
    Butler, D. J.
    Crosbie, J. C.
    Livingstone, J.
    Poole, C. M.
    Ramanathan, G.
    Wright, T.
    Stevenson, A. W.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (11) : 4201 - 4222
  • [10] Design and Construction of a Cylindrical Ionization Chamber for Reference Dosimetry in Radiation Protection
    Imani-Shirvanehdeh, Hamed
    Khorshidi, Abdollah
    Soltani-Nabipour, Jamshid
    Alipour, Anita
    Arbabi, Kourosh