Biological verification of heavy ion treatment planning

被引:0
作者
A. Mitaroff
W. Kraft-Weyrather
O. B. Geiß
G. Kraft
机构
[1] Gesellschaft für Schwerionenforschung – Biophysik,
[2] Planckstrasse 1,undefined
[3] D-64291 Darmstadt,undefined
[4] Germany,undefined
来源
Radiation and Environmental Biophysics | 1998年 / 37卷
关键词
Treatment Planning; Target Volume; Dose Distribution; Chinese Hamster Ovary Cell; Chinese Hamster Ovary;
D O I
暂无
中图分类号
学科分类号
摘要
Chinese hamster ovary (CHO) cells and thermoluminescent detectors (TLD-700) were used for physical and biological verification of heavy ion treatment planning. Experiments were performed in a cylindrical water phantom, in some cases with lung and bone equivalent material in front of the target volume. The results confirm the possibility of using thermoluminescent detectors for a quantitative verification of dose distributions. CHO cells can be used at least for qualitative dose verification.
引用
收藏
页码:47 / 51
页数:4
相关论文
empty
未找到相关数据