Usefulness of subcutaneous implantable cardioverter-defibrillator therapy in patients with Brugada syndrome

被引:0
作者
Tetsuji Shinohara
Ichitaro Abe
Kei Hirota
Hidekazu Kondo
Akira Fukui
Hidefumi Akioka
Yasushi Teshima
Kunio Yufu
Mikiko Nakagawa
Naohiko Takahashi
机构
[1] Oita University,Department of Cardiology and Clinical Examination, Faculty of Medicine
来源
Heart and Vessels | 2021年 / 36卷
关键词
Brugada syndrome; Inappropriate shock; SMART pass; Subcutaneous implantable cardioverter-defibrillator; T-wave oversensing;
D O I
暂无
中图分类号
学科分类号
摘要
Implantable cardioverter-defibrillators (ICDs) are the main therapy to prevent sudden cardiac death in patients with Brugada syndrome (BrS). The subcutaneous ICD (S-ICD) can eliminate lead-associated complications compared with transvenous ICD (TV-ICD). However, S-ICD is susceptible to T-wave oversensing (TWOS) and may result in more frequent inappropriate shocks in patients with BrS. This study aimed to compare inappropriate shocks between TV-ICD and S-ICD in patients with BrS. We enrolled 32 patients with BrS (including one woman; mean age 52 ± 18 years) who were implanted with ICD (23 TV-ICDs and 9 S-ICDs) between January 2002 and November 2018 in Oita University Hospital. We carried out a standard surface electrocardiogram (ECG) screening tests in both supine and standing positions prior to S-ICD implantation. The patients received routine clinical review every month and device monitoring every 4 months. The period of follow-up was 129 ± 51 months. Six patients with BrS and TV-ICDs experienced inappropriate shocks (26%) with their ICD therapy. In contrast, two patients with BrS and S-ICDs experienced inappropriate shocks (22%). There was no significant difference between the two groups (P = 0.82). Although one case in the S-ICD group experienced TWOS-induced inappropriate shock, SMART Pass (new high-pass filter) prevented the subsequent recurrence of inappropriate shocks during ICD therapy. Our results suggest that S-ICD is not inferior to TV-ICD in the incidence of inappropriate shocks. SMART Pass may be a useful tool to prevent inappropriate ICD shocks by TWOS in patients with BrS.
引用
收藏
页码:260 / 266
页数:6
相关论文
共 360 条
  • [1] Priori SG(2013)Executive summary: Hrs/ehra/aphrs expert consensus statement on the diagnosis and management of patients with inherited primary arrhythmia syndromes Europace 15 1389-1406
  • [2] Wilde AA(2010)An entirely subcutaneous implantable cardioverter-defibrillator N Engl J Med 363 36-44
  • [3] Horie M(2020)Questionnaire in patients with aborted sudden cardiac death due to coronary spasm in Japan Heart Vessels 34 851-859
  • [4] Cho Y(2019)Prevalence of subcutaneous implantable cardioverter-defibrillator based on template ECG screening and ineligible surface ECG predicting factors in patients with hypertrophic cardiomyopathy in China Heart Vessels 20 1188-1193
  • [5] Behr ER(2018)High rate of subcutaneous implantable cardioverter-defibrillator sensing screening failure in patients with brugada syndrome: a comparison with other inherited primary arrhythmia syndromes Europace 15 1515-1522
  • [6] Berul C(2018)Prospective blinded evaluation of a novel sensing methodology designed to reduce inappropriate shocks by the subcutaneous implantable cardioverter-defibrillator Heart Rhythm 20 f240-f248
  • [7] Blom N(2018)Propensity score matched comparison of subcutaneous and transvenous implantable cardioverter-defibrillator therapy in the simple and effortless studies Europace 36 2793-2867
  • [8] Brugada J(2015)2015 esc guidelines for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: The task force for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death of the european society of cardiology (esc). Endorsed by: Association for european paediatric and congenital cardiology (aepc) Eur Heart J 111 659-670
  • [9] Chiang CE(2005)Brugada syndrome: report of the second consensus conference: endorsed by the heart rhythm society and the european heart rhythm association Circulation 10 e85-108
  • [10] Huikuri H(2013)Executive summary: Hrs/ehra/aphrs expert consensus statement on the diagnosis and management of patients with inherited primary arrhythmia syndromes Heart Rhythm 117 2721-2723