Evaluation of the stress distribution in CFR-PEEK dental implants by the three-dimensional finite element method

被引:0
|
作者
João Rodrigo Sarot
Cintia Mussi Milani Contar
Ariadne Cristiane Cabral da Cruz
Ricardo de Souza Magini
机构
[1] Universidade Federal do Paraná,Department of Pos
[2] Universidade Tuiuti do Paraná,Graduation Course of Implantology
[3] Universidade Federal de Santa Catarina,Department of Oral Radiology
[4] Universidade Federal de Santa Catarina,Biotecnology Post
来源
Journal of Materials Science: Materials in Medicine | 2010年 / 21卷
关键词
Carbon Fiber; Cortical Bone; Zinc Phosphate; Titanium Implant; Poly Ether Ether Ketone;
D O I
暂无
中图分类号
学科分类号
摘要
CFR-PEEK (carbon fiber reforced—poly ether ether ketone) has been demonstrated to be excellent substitute titanium in orthopedic applications and can be manufactured with many physical, mechanical, and surface properties, in several shapes. The aim of this study was to compare, using the three-dimensional finite element method, the stress distribution in the peri-implant support bone of distinct models composed of PEEK components and implants reinforced with 30% carbon fiber (30% CFR-PEEK) or titanium. In simulations with a perfect bonding between the bone and the implant, the 30% CFR-PEEK presented higher stress concentration in the implant neck and the adjacent bone, due to the decreased stiffness and higher deformation in relation to the titanium. However, 30% CFR-PEEK implants and components did not exhibit any advantages in relation to the stress distribution compared to the titanium implants and components.
引用
收藏
页码:2079 / 2085
页数:6
相关论文
共 50 条
  • [1] Evaluation of the stress distribution in CFR-PEEK dental implants by the three-dimensional finite element method
    Sarot, Joao Rodrigo
    Milani Contar, Cintia Mussi
    Cabral da Cruz, Ariadne Cristiane
    Magini, Ricardo de Souza
    JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2010, 21 (07) : 2079 - 2085
  • [2] Comparative analysis of stress distribution around CFR-PEEK implants and titanium implants with different prosthetic crowns: A finite element analysis
    Tamrakar, Swapnil Kumar
    Mishra, Sunil Kumar
    Chowdhary, Ramesh
    Rao, Srinivas
    DENTAL AND MEDICAL PROBLEMS, 2021, 58 (03) : 359 - 367
  • [3] Evaluation of Stress Distribution during Insertion of Tapered Dental Implants in Various Osteotomy Techniques: Three-Dimensional Finite Element Study
    Mahendra, Jaideep
    Chand, Yemineni Bhavan
    Mahendra, Little
    Fageeh, Hytham N.
    Fageeh, Hammam Ibrahim
    Ibraheem, Wael
    Alzahrani, Khaled M.
    Alqahtani, Nasser M.
    Alahmari, Nasser Mesfer
    Almagbol, Mohammad
    Robaian, Ali
    Jigeesh, Nasina
    Varadarajan, Saranya
    Balaji, Thodur Madapusi
    Patil, Shankargouda
    MATERIALS, 2021, 14 (24)
  • [4] Stress distribution of two commercial dental implant systems: A three-dimensional finite element analysis
    Chang, Hao-Sheng
    Chen, Yi-Chin
    Hsieh, Yao-Dung
    Hsu, Ming-Lun
    JOURNAL OF DENTAL SCIENCES, 2013, 8 (03) : 261 - 271
  • [5] Prediction of Wear on Tibial Inserts Made of UHMWPE, PEEK, and CFR-PEEK in Total Knee Arthroplasty Using Finite-Element Analysis
    Koh, Yong-Gon
    Lee, Jin-Ah
    Kang, Kyoung-Tak
    LUBRICANTS, 2019, 7 (04)
  • [6] The influence of framework material on stress distribution in maxillary complete-arch fixed prostheses supported by four dental implants: a three-dimensional finite element analysis
    Dayan, Suleyman Cagatay
    Geckili, Onur
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2021, 24 (14) : 1606 - 1617
  • [7] Evaluating Angled Abutments: Three-Dimensional Finite Element Stress Analysis of Anterior Maxillary Implants
    Chethan, K. N.
    Eram, Afiya
    Shetty, Nisha
    Shetty, Divya D.
    Futane, Mohan
    Keni, Laxmikant G.
    PROSTHESIS, 2024, 6 (02): : 315 - 328
  • [8] Finite element analysis of the biomechanical effects of titanium and Cfr-peek additively manufactured subperiosteal jaw implant (AMSJI) on maxilla
    Altiparmak, Nur
    Polat, Serhat
    Onat, Selen
    JOURNAL OF STOMATOLOGY ORAL AND MAXILLOFACIAL SURGERY, 2023, 124 (01)
  • [9] Stress Distribution in Obliquely Inserted Orthodontic Miniscrews Evaluated by Three-Dimensional Finite-Element Analysis
    Kuroda, Shingo
    Inoue, Masahide
    Kyung, Hee-Moon
    Koolstra, Jan Harm
    Tanaka, Eiji
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2017, 32 (02) : 344 - 349
  • [10] Effect of different restorative crown and customized abutment materials on stress distribution in single implants and peripheral bone: A three-dimensional finite element analysis study
    Kaleli, Necati
    Sarac, Duygu
    Kulunk, Safak
    Ozturk, Ozgur
    JOURNAL OF PROSTHETIC DENTISTRY, 2018, 119 (03) : 437 - 445