High-resolution entry and exit surface dosimetry in a 1.5 T MR-linac

被引:0
|
作者
E. Patterson
P. Stokes
D. Cutajar
A. Rosenfeld
J. Baines
P. Metcalfe
M. Powers
机构
[1] University of Wollongong,Centre of Medical and Radiation Physics
[2] University of Wollongong,Illawarra Health Medical Research Institute
[3] Ingham Institute for Applied Medical Research,Townsville Cancer Centre
[4] Townsville Hospital and Health Service,College of Science and Engineering
[5] James Cook University,undefined
来源
Physical and Engineering Sciences in Medicine | 2023年 / 46卷
关键词
MR-linac; Elekta Unity; MOSFET; Skin dose; Surface dosimetry; Monte Carlo;
D O I
暂无
中图分类号
学科分类号
摘要
The magnetic field of a transverse MR-linac alters electron trajectories as the photon beam transits through materials, causing lower doses at flat entry surfaces and increased doses at flat beam-exiting surfaces. This study investigated the response of a MOSFET detector, known as the MOSkin™, for high-resolution surface and near-surface percentage depth dose measurements on an Elekta Unity. Simulations with Geant4 and the Monaco treatment planning system (TPS), and EBT-3 film measurements, were also performed for comparison. Measured MOSkin™ entry surface doses, relative to Dmax, were (9.9 ± 0.2)%, (10.1 ± 0.3)%, (11.3 ± 0.6)%, (12.9 ± 1.0)%, and (13.4 ± 1.0)% for 1 × 1 cm2, 3 × 3 cm2, 5 × 5 cm2, 10 × 10 cm2, and 22 × 22 cm2 fields, respectively. For the investigated fields, the maximum percent differences of Geant4, TPS, and film doses extrapolated and interpolated to a depth suitable for skin dose assessment at the beam entry, relative to MOSkin™ measurements at an equivalent depth were 1.0%, 2.8%, and 14.3%, respectively, and at a WED of 199.67 mm at the beam exit, 3.2%, 3.7% and 5.7%, respectively. The largest measured increase in exit dose, due to the electron return effect, was 15.4% for the 10 × 10 cm2 field size using the MOSkin™ and 17.9% for the 22 × 22 cm2 field size, using Geant4 calculations. The results presented in the study validate the suitability of the MOSkin™ detector for transverse MR-linac surface dosimetry.
引用
收藏
页码:787 / 800
页数:13
相关论文
共 50 条
  • [31] Feasibility of using a dual isocentre technique for treating cervical cancer on the 1.5 T MR-Linac
    Chuter, Robert W.
    Brewster, Frank
    Retout, Louise
    Cree, Anthea
    Akturk, Nesrin
    Hales, Rosie
    Benson, Rebecca
    Hoskin, Peter
    McWilliam, Alan
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (02)
  • [32] Experimental dosimetric verification of the intra-fraction drift correction on the 1.5 T MR-linac
    van den Dobbelsteen, Madelon
    Hackett, Sara L.
    Oolbekkink, Stijn
    van Asselen, Bram
    Uijtewaal, Prescilla
    Fast, Martin F.
    Raaymakers, Bas W.
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S4152 - S4155
  • [33] Seminal vesicle intrafraction motion during the delivery of radiotherapy sessions on a 1.5 T MR-Linac
    De, D. M. Muinck Keizer
    Willigenburg, T.
    Van, J. R. N. der Voort van Zyp
    Raaymakers, B. W.
    Lagendijk, J. J. W.
    De, J. C. J. Boer
    RADIOTHERAPY AND ONCOLOGY, 2021, 162 : 162 - 169
  • [34] Two-dimensional EPID dosimetry for an MR-linac: Proof of concept
    Torres-Xirau, Iban
    Olaciregui-Ruiz, Igor
    van der Heide, Uulke A.
    Mans, Anton
    MEDICAL PHYSICS, 2019, 46 (09) : 4193 - 4203
  • [35] An improved calibration procedure for accurate plastic scintillation dosimetry on an MR-linac
    van den Dobbelsteen, Madelon
    Lessard, Boby
    Cote, Benjamin
    Hackett, Sara L.
    Mugnes, Jean-Michel
    Therriault-Proulx, Francois
    Lambert-Girard, Simon
    Uijtewaal, Prescilla
    de Vries, Laurie J. M.
    Archambault, Louis
    Bosma, Tom
    van Asselen, Bram
    Raaymakers, Bas W.
    Fast, Martin F.
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (24)
  • [36] Quantitative T1 mapping in the presence of an implantable cardiac device on a 1.5T MR-linac
    Pontre, Beau P.
    Mandija, Stefano
    Akdag, Osman
    Schakel, Tim
    van den Berg, Cornelis A. T.
    Fast, Martin F.
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S3886 - S3889
  • [37] Longitudinal assessment of quality assurance measurements in a 1.5T MR-linac: Part I-Linear accelerator
    Subashi, Ergys
    Lim, Seng Boh
    Gonzalez, Xesus
    Tyagi, Neelam
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (10): : 190 - 201
  • [38] Experimental determination of magnetic field quality conversion factors for eleven ionization chambers in 1.5 T and 0.35 T MR-linac systems
    Orlando, Nathan
    Crosby, Jennie
    Glide-Hurst, Carri
    Culberson, Wesley
    Keller, Brian
    Sarfehnia, Arman
    MEDICAL PHYSICS, 2024, 51 (04) : 2998 - 3009
  • [39] Dosimetric Accuracy of MR-Guided Online Adaptive Planning for Nasopharyngeal Carcinoma Radiotherapy on 1.5 T MR-Linac
    Ding, Shouliang
    Liu, Hongdong
    Li, Yongbao
    Wang, Bin
    Li, Rui
    Huang, Xiaoyan
    FRONTIERS IN ONCOLOGY, 2022, 12
  • [40] Inter-scanner ADC reproducibility on prostate cancer patients: 1.5T MR-Linac vs 1.5T MR-simulator
    Wong, Oi Lei
    Yuan, Jing
    Poon, Darren M. C.
    Chiu, Sin Ting
    Yang, Bin
    Xue, Cindy
    Chiu, George
    Cheung, Kin Yin
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S2331 - S2333