High-resolution entry and exit surface dosimetry in a 1.5 T MR-linac

被引:0
|
作者
E. Patterson
P. Stokes
D. Cutajar
A. Rosenfeld
J. Baines
P. Metcalfe
M. Powers
机构
[1] University of Wollongong,Centre of Medical and Radiation Physics
[2] University of Wollongong,Illawarra Health Medical Research Institute
[3] Ingham Institute for Applied Medical Research,Townsville Cancer Centre
[4] Townsville Hospital and Health Service,College of Science and Engineering
[5] James Cook University,undefined
来源
Physical and Engineering Sciences in Medicine | 2023年 / 46卷
关键词
MR-linac; Elekta Unity; MOSFET; Skin dose; Surface dosimetry; Monte Carlo;
D O I
暂无
中图分类号
学科分类号
摘要
The magnetic field of a transverse MR-linac alters electron trajectories as the photon beam transits through materials, causing lower doses at flat entry surfaces and increased doses at flat beam-exiting surfaces. This study investigated the response of a MOSFET detector, known as the MOSkin™, for high-resolution surface and near-surface percentage depth dose measurements on an Elekta Unity. Simulations with Geant4 and the Monaco treatment planning system (TPS), and EBT-3 film measurements, were also performed for comparison. Measured MOSkin™ entry surface doses, relative to Dmax, were (9.9 ± 0.2)%, (10.1 ± 0.3)%, (11.3 ± 0.6)%, (12.9 ± 1.0)%, and (13.4 ± 1.0)% for 1 × 1 cm2, 3 × 3 cm2, 5 × 5 cm2, 10 × 10 cm2, and 22 × 22 cm2 fields, respectively. For the investigated fields, the maximum percent differences of Geant4, TPS, and film doses extrapolated and interpolated to a depth suitable for skin dose assessment at the beam entry, relative to MOSkin™ measurements at an equivalent depth were 1.0%, 2.8%, and 14.3%, respectively, and at a WED of 199.67 mm at the beam exit, 3.2%, 3.7% and 5.7%, respectively. The largest measured increase in exit dose, due to the electron return effect, was 15.4% for the 10 × 10 cm2 field size using the MOSkin™ and 17.9% for the 22 × 22 cm2 field size, using Geant4 calculations. The results presented in the study validate the suitability of the MOSkin™ detector for transverse MR-linac surface dosimetry.
引用
收藏
页码:787 / 800
页数:13
相关论文
共 50 条
  • [1] High-resolution entry and exit surface dosimetry in a 1.5 T MR-linac
    Patterson, E.
    Stokes, P.
    Cutajar, D.
    Rosenfeld, A.
    Baines, J.
    Metcalfe, P.
    Powers, M.
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2023, 46 (02) : 787 - 800
  • [2] Surface and near-surface dose measurements at beam entry and exit in a 1.5 T MR-Linac using optically stimulated luminescence dosimeters
    Kim, Anthony
    Lim-Reinders, Stephanie
    Ahmad, Syed Bilal
    Sahgal, Arjun
    Keller, Brian M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (04)
  • [3] Performance of the HYPERSCINT scintillation dosimetry research platform for the 1.5 T MR-linac
    Uijtewaal, Prescilla
    Cote, Benjamin
    Foppen, Thomas
    de Vries, Wilfred
    Woodings, Simon
    Borman, Pim
    Lambert-Girard, Simon
    Therriault-Proulx, Francois
    Raaymakers, Bas
    Fast, Martin
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (04)
  • [4] Electron streaming dose measurements and calculations on a 1.5 T MR-Linac
    Patterson, Elizabeth
    Powers, Marcus
    Metcalfe, Peter E.
    Cutajar, Dean
    Oborn, Bradley M.
    Baines, John A.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024, 25 (07):
  • [5] The Quality Assurance of a 1.5 T MR-Linac
    Riis, Hans Lynggaard
    Chick, Joan
    Dunlop, Alex
    Tilly, David
    SEMINARS IN RADIATION ONCOLOGY, 2024, 34 (01) : 120 - 128
  • [6] Relative dosimetry in a 1.5 T magnetic field: an MR-linac compatible prototype scanning water phantom
    Smit, K.
    Sjoholm, J.
    Kok, J. G. M.
    Lagendijk, J. J. W.
    Raaymakers, B. W.
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (15) : 4099 - 4109
  • [7] Dosimetry with the TruView gel on a 0.35 T MR-Linac: A feasibility study
    Ermeneux, L.
    Petitfils, A.
    Marage, L.
    Gschwind, R.
    Huet, C.
    RADIATION MEASUREMENTS, 2024, 175
  • [8] Performance of a cylindrical diode array for use in a 1.5 T MR-linac
    Houweling, A. C.
    de Vries, J. H. W.
    Wolthaus, J.
    Woodings, S.
    Kok, J. G. M.
    van Asselen, B.
    Smit, K.
    Bel, A.
    Lagendijk, J. J. W.
    Raaymakers, B. W.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (03) : N80 - N89
  • [9] Elekta Unity MR-linac commissioning: mechanical and dosimetry tests
    Tsuneda, Masato
    Abe, Kota
    Fujita, Yukio
    Ikeda, Yohei
    Furuyama, Yoshinobu
    Uno, Takashi
    JOURNAL OF RADIATION RESEARCH, 2023, 64 (01) : 73 - 84
  • [10] Dosimetric evaluation of off-axis fields and angular transmission for the 1.5 T MR-linac
    van den Dobbelsteen, M.
    Hackett, S. L.
    de Vries, J. H. W.
    van Asselen, B.
    Oolbekkink, S.
    Woodings, S. J.
    Raaymakers, B. W.
    PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (20)