Influence of the tibial stem design on bone density after cemented total knee arthroplasty: a prospective seven-year follow-up study

被引:0
作者
Daniel Hernandez-Vaquero
Manuel A. Garcia-Sandoval
Jose M. Fernandez-Carreira
Richard Gava
机构
[1] School of Medicine,
[2] University of Oviedo,undefined
[3] Department of Orthopaedic Surgery,undefined
[4] Hospital San Agustin,undefined
[5] Clinical Research Unit,undefined
[6] Hospital San Agustin,undefined
来源
International Orthopaedics | 2008年 / 32卷
关键词
Bone Mineral Density; Total Knee Arthroplasty; Tibial Component; Bone Mineral Density Change; Bone Mineral Density Decrease;
D O I
暂无
中图分类号
学科分类号
摘要
We prospectively measured the changes in bone mineral density (BMD) in the proximal tibia of 20 total knee arthroplasties, ten with cruciform stems and ten with cylindrical stems. The measurements were made one, four and seven years after surgery. We observed a uniform density decrease in three regions of interest from one to seven years of follow-up. Cylindrical stems showed an asymmetrical density decrease between the three regions of interest, with no change in the central region, a slight decrease in the lateral region, and large decrease in the medial region. Multivariate analysis with general linear model showed the stem type factor as statistically significant for medial region of interest (p = 0.006). The cylindrical stem produces heterogeneous BMD changes under the tibial platform in knee arthroplasties, and this could be a potential risk factor for asymmetrical subsidence of this component.
引用
收藏
页码:47 / 51
页数:4
相关论文
共 61 条
[11]  
Eckardt A(1999)Preoperative bone mineral density of the proximal tibia and migration of the tibial component after uncemented total knee arthroplasty J Arthroplasty 14 77-81
[12]  
Heine J(1999)Bone mineral and migratory patterns in uncemented total knee arthroplasties: a randomized 5-year follow-up study of 38 knees Acta Orthop Scand 70 603-608
[13]  
Lee RW(2000)Measurement of bone density around total knee arthroplasty using fan-beam dual energy X-ray absorptiometry Calcif Tissue Int 67 267-272
[14]  
Volz RG(2004)Periprosthetic tibial bone mineral density changes after total knee arthroplasty: one-year follow-up study of 69 patients Acta Orthop Scand 75 600-605
[15]  
Sheridan DC(1997)The effect of tibial stem design on component micromotion in knee arthroplasty Clin Orthop 345 44-52
[16]  
Levitz CL(1998)Bone assessment after total knee arthroplasty by dual-energy X-ray absorptiometry: analysis protocol and reproducibility Calcif Tissue Int 62 359-361
[17]  
Lotke PA(undefined)undefined undefined undefined undefined-undefined
[18]  
Karp JS(undefined)undefined undefined undefined undefined-undefined
[19]  
Li MG(undefined)undefined undefined undefined undefined-undefined
[20]  