Improving tribological and anti-bacterial properties of titanium external fixation pins through surface ceramic conversion

被引:0
作者
Huan Dong
Tatiana Mukinay
Maojun Li
Richard Hood
Sein Leung Soo
Simon Cockshott
Rachel Sammons
Xiaoying Li
机构
[1] Royal Derby Hospital,School of Metallurgy and Materials
[2] University of Birmingham,School of Engineering
[3] University of Birmingham,School of Dentistry
[4] University of Birmingham,undefined
来源
Journal of Materials Science: Materials in Medicine | 2017年 / 28卷
关键词
Plasma Electrolytic Oxidation; Surface Oxide Layer; Insertion Force; Wear Groove; Coupon Sample;
D O I
暂无
中图分类号
学科分类号
摘要
In this study, an advanced ceramic conversion surface engineering technology has been applied for the first time to self-drilling Ti6Al4V external fixation pins to improve their performance in terms of biomechanical, bio-tribological and antibacterial properties. Systematic characterisation of the ceramic conversion treated Ti pins was carried out using Scanning electron microscope, X-ray diffraction, Glow-discharge optical emission spectroscopy, nano- and micro-indentation and scratching; the biomechanical and bio-tribological properties of the surface engineered Ti pins were evaluated by insertion into high density bone simulation material; and the antibacterial behaviour was assessed with Staphylococcus aureus NCTC 6571. The experimental results have demonstrated that the surfaces of Ti6Al4V external fixation pins were successfully converted into a TiO2 rutile layer (~2 μm in thickness) supported by an oxygen hardened case (~15 μm in thickness) with very good bonding due to the in-situ conversion nature. The maximum insertion force and temperature were reduced from 192N and 31.2 °C when using the untreated pins to 182N and 26.1 °C when the ceramic conversion treated pins were tested. This is mainly due to the significantly increased hardness (more than three times) and the effectively enhanced wear resistance of the cutting edge of the self-drilling Ti pins following the ceramic conversion treatment. The antibacterial tests also revealed that there was a significantly reduced number of bacteria isolated from the ceramic conversion treated pins compared to the untreated pins of around 50 % after 20 h incubation, P < 0.01 (0.0024). The results reported are encouraging and could pave the way towards high-performance anti-bacterial titanium external fixation pins with reduced pin-track infection and pin loosing.
引用
收藏
相关论文
共 98 条
[31]  
Klenke FM(undefined)undefined undefined undefined undefined-undefined
[32]  
Jennison T(undefined)undefined undefined undefined undefined-undefined
[33]  
McNally M(undefined)undefined undefined undefined undefined-undefined
[34]  
Pandit H(undefined)undefined undefined undefined undefined-undefined
[35]  
Visai L(undefined)undefined undefined undefined undefined-undefined
[36]  
De Nardo L(undefined)undefined undefined undefined undefined-undefined
[37]  
Punta C(undefined)undefined undefined undefined undefined-undefined
[38]  
Melone L(undefined)undefined undefined undefined undefined-undefined
[39]  
Cigada A(undefined)undefined undefined undefined undefined-undefined
[40]  
Imbriani M(undefined)undefined undefined undefined undefined-undefined