Calibration accuracy of hospital-based non-invasive blood pressure measuring devices

被引:0
作者
A de Greeff
I Lorde
A Wilton
P Seed
A J Coleman
A H Shennan
机构
[1] Maternal and Fetal Research Unit,Division of Reproduction and Endocrinology
[2] King’s College London,Medical Physics Department
[3] Guy’s and St Thomas’ NHS Foundation Trust,undefined
来源
Journal of Human Hypertension | 2010年 / 24卷
关键词
calibration; blood pressure measurement; device; accuracy; non-invasive;
D O I
暂无
中图分类号
学科分类号
摘要
Accurate blood pressure (BP) measurement is dependent on a trained observer using validated and properly maintained equipment. BP devices should be checked regularly to ensure that their calibration remains within the European Standard specification of ±3 mm Hg. This study assessed the air leakage rates and calibration accuracy of BP devices in use at a large teaching hospital, using a calibrated electronic pressure gauge as reference. Air leakage rates were recorded over 1 min and static pressures were recorded at 250/200/150/100/50/0 mm Hg for computer download and analysis. A total of 127 devices were assessed (18 mercury, 62 aneroid and 47 automated). In total, 22 different models of devices were available, of which 11 were automated and only 4 had published evidence of a validation using a recognized protocol (British Hypertension Society, Association for the Advancement of Medical Instrumentation or International Protocol). Only 3% (n=4) of devices had an air leakage rate within 4 mm Hg per min and 25% (n=32) of devices failed to meet the European calibration standard of ±3 mm Hg. Respective failure rates were 6% (1/18) for mercury, 31% (19/62) for aneroid and 26% (12/47) for automated devices. Inaccurate BP measurement of only 3 mm Hg can have detrimental effects in the patient. This study shows a quarter of devices currently in use at a large teaching hospital to have an unacceptable calibration error. Regular maintenance and calibration checks are vital in ensuring that BP is measured as accurately as possible.
引用
收藏
页码:58 / 63
页数:5
相关论文
共 82 条
[1]  
Chobanian AV(2003)2003 European Society of Hypertension-European Society of Cardiology guidelines for the management of arterial hypertension J Hypertens 21 1011-1053
[2]  
Bakris GL(2003)The Seventh Report of the Joint National Committee on prevention, detection, evaluation, and treatment of high blood pressure: the JNC 7 report JAMA 289 2560-2572
[3]  
Black HR(1993)The British Hypertension Society protocol for the evaluation of blood pressure measuring devices J Hypertens 11 S43-S62
[4]  
Cushman WC(2002)Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults Blood Press Monit 7 3-17
[5]  
Green LA(2006)Lack of sphygmomanometer calibration causes over- and under-detection of hypertension: a computer simulation study J Hypertens 24 1931-1938
[6]  
Izzo JL(2002)Practice audits: reliability of sphygmomanometers and blood pressure recording bias J Hum Hypertens 16 359-361
[7]  
O’Brien E(2001)The extent and implications of sphygmomanometer calibration error in primary care J Hum Hypertens 15 587-591
[8]  
Petrie J(2001)Sphygmomanometers in use in general practice: an overlooked aspect of quality in patient care J Hum Hypertens 15 681-684
[9]  
Littler W(2005)Accuracy of the pressure scale of sphygmomanometers in clinical use within primary care Blood Press Monit 10 181-188
[10]  
de Swiet M(1988)Blood pressure measurement in family physician offices Fam Med 20 25-27