Application of Design of Experiments® Approach-Driven Artificial Intelligence and Machine Learning for Systematic Optimization of Reverse Phase High Performance Liquid Chromatography Method to Analyze Simultaneously Two Drugs (Cyclosporin A and Etodolac) in Solution, Human Plasma, Nanocapsules, and Emulsions

被引:0
作者
Syed Nazrin Ruhina Rahman
Oly Katari
Datta Maroti Pawde
Gopi Sumanth Bhaskar Boddeda
Abhinab Goswami
Srinivasa Rao Mutheneni
Tamilvanan Shunmugaperumal
机构
[1] National Institute of Pharmaceutical Education and Research (NIPER),Departments of Pharmaceutics
[2] CSIR-Indian Institute of Chemical Technology,Bioinformatics Group, Applied Biology Division
来源
AAPS PharmSciTech | / 22卷
关键词
analytical quality by design; face-centered central composite design; artificial neural network; Levenberg–Marquardt algorithm; cyclosporin A; etodolac; human plasma;
D O I
暂无
中图分类号
学科分类号
摘要
The objectives of current investigation are (1) to find out wavelength of maximum absorbance (λmax) for combined cyclosporin A and etodolac solution followed by selection of mobile phase suitable for the RP-HPLC method, (2) to define analytical target profile and critical analytical attributes (CAAs) for the analytical quality by design, (3) to screen critical method parameters with the help of full factorial design followed by optimization with face-centered central composite design (CCD) approach-driven artificial neural network (ANN)-linked with the Levenberg–Marquardt (LM) algorithm for finding the RP-HPLC conditions, (4) to perform validation of analytical procedures (trueness, linearity, precision, robustness, specificity and sensitivity) using combined drug solution, and (5) to determine drug entrapment efficiency value in dual drug-loaded nanocapsules/emulsions, percentage recovery value in human plasma spiked with two drugs and solution state stability analysis at different stress conditions for substantiating the double-stage systematically optimized RP-HPLC method conditions. Through isobestic point and scouting step, 205 nm and ACN:H2O mixture (74:26) were selected respectively as the λmax and mobile phase. The ANN topology (3:10:4) indicating the input, hidden and output layers were generated by taking the 20 trials produced from the face-centered CCD model. The ANN-linked LM model produced minimal differences between predicted and observed values of output parameters (or CAAs), low mean squared error and higher correlation coefficient values in comparison to the respective values produced by face-centered CCD model. The optimized RP-HPLC method could be applied to analyze two drugs concurrently in different formulations, human plasma and solution state stability checking.
引用
收藏
相关论文
共 85 条
  • [1] Fruman DA(1992)Calcineurin phosphatase activity in T lymphocytes is inhibited by FK 506 and cyclosporin A Proc Natl Acad Sci 89 3686-3690
  • [2] Klee CB(2007)Keratoconjunctivitis sicca associated with administration of etodolac in dogs: 211 cases (1992–2002) J Am Vet Med Assoc 230 541-547
  • [3] Bierer BE(2019)Studies on olive-and silicone-oils-based Janus macroemulsions containing ginger to manage primary dysmenorrhealpain Mater Sci Eng C Mater Biol Appl 100 276-285
  • [4] Burakoff SJ(2009)Oral cyclosporine A—the current picture of its liposomal and other delivery systems Cell Mol Biol Lett 14 139-152
  • [5] Klauss GI(2019)UV Spectrophotometric and stability indicating RP-HPLC assay methods for the estimation of etodolac in bulk and tablet dosage form P Int J Pharma Investi 9 53-58
  • [6] Giuliano EA(2019)Quality by design based development and validation of HPLC method for simultaneous estimation of paclitaxel and vinorelbine tartrate in dual drug loaded liposomes Microchem J 149 103982-63
  • [7] Moore CP(2018)Quality by design approach for development and validation of a RP-HPLC method for simultaneous determination of co-administered levetiracetam and pyridoxine HCl in prepared tablets Microchem J 143 55-S322
  • [8] Puri D(2016)Quality by Design (QbD) approach to develop HPLC method for eberconazole nitrate: application to hydrolytic, thermal, oxidative and photolytic degradation kinetics J Saudi Chem Soc 20 S313-416
  • [9] Khatik GL(2015)The applications of genetic algorithms in medicine Oman Med J 30 406-1752
  • [10] Tamilvanan SS(2007)The use of artificial neural networks in materials science based research Mater Des 28 1747-16