Ir-192 Calibration in Air with Farmer Chamber for HDR Brachytherapy

被引:0
|
作者
Liyun Chang
Sheng-Yow Ho
Tsair-Fwu Lee
Hueisch-Jy Ding
Pang-Yu Chen
机构
[1] I-Shou University,Department of Medical Imaging and Radiological Sciences
[2] Chang Jung Christian University,Department of Nursing
[3] Chi Mei Medical Center,Department of Radiation Oncology
[4] National Kaohsiung University of Applied Sciences,Medical Physics and Informatics Laboratory, Department of Electronics Engineering
[5] Sinlau Christian Hospital,Department of Radiation Oncology
来源
Journal of Medical and Biological Engineering | 2016年 / 36卷
关键词
Ir-192 calibration; In-air calibration; Farmer-type ion chamber; High dose rate (HDR);
D O I
暂无
中图分类号
学科分类号
摘要
A comprehensive review for the in-air calibration of an Ir-192 high-dose-rate brachytherapy source in terms of air kerma strength (AKS) and reference air kerma rate (RAKR) using the Farmer chamber was conducted. The reviewed calibration methods include the National Physical Laboratory (NPL) calibration standard in the UK, the 7-distance technique with the standard calibration of the National Institute of Standards and Technology and Accredited Dosimetry Calibration Laboratory in the US, the calibration conducted in Australia following recommendations of the International Atomic Energy Agency with the chamber primarily calibrated by the Australian Radiation Protection and Nuclear Safety Agency, the calibration conducted in India following the Deutsche Gesellschaft fur Medizinische Physik recommendation, and the convenient empirical method used in Taiwan. The calibrated AKS (or RAKR) and uncertainty obtained using Farmer chambers are similar to those obtained using well chambers. All reported differences (between measurements obtained using Farmer and well chambers, respectively) and uncertainties (k = 2) were generally less than 1.5 and 2.5 %, respectively. The standard uncertainty of the NPL calibration is approximately half that of all the other proposed approaches, and may become the gold standard calibration procedure. Almost all techniques follow the 7-distance technique basis; however, the services at NPL can calibrate the source with lower uncertainty. Users can calibrate the Ir-192 source more conveniently using the empirical method with only one source-chamber distance.
引用
收藏
页码:145 / 152
页数:7
相关论文
共 4 条
  • [1] Ir-192 Calibration in Air with Farmer Chamber for HDR Brachytherapy
    Chang, Liyun
    Ho, Sheng-Yow
    Lee, Tsair-Fwu
    Ding, Hueisch-Jy
    Chen, Pang-Yu
    JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2016, 36 (02) : 145 - 152
  • [2] An innovative method for 192Ir HDR calibration by farmer chamber, V-film, and solid phantom
    Chang, Liyun
    Ding, Hueisch-Jy
    Ho, Sheng-Yow
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 646 (01): : 192 - 196
  • [3] Calibration of Ir-192 high dose rate brachytherapy source using different calibration procedures
    Bondel, Shwetha
    Ravikumar, Manickham
    Supe, Sanjay Sudhakar
    Reddy, Buchuppudi Rekha
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2014, 19 (03) : 151 - 156
  • [4] Determination of contributions of scatter and distance error to the source strength of Ir-192 HDR brachytherapy source
    BONDEL, Shwetha
    RAVIKUMAR, Manickam
    POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2016, 22 (03) : 55 - 59