Assessment of flatness and symmetry of megavoltage x-ray beam with an electronic portal imaging device (EPID)

被引:15
作者
Liu G. [1 ]
Van Doorn T. [1 ]
Bezak E. [1 ]
机构
[1] Department of Medical Physics, Royal Adelaide Hospital, Adelaide
来源
Australasian Physics & Engineering Sciences in Medicine | 2002年 / 25卷 / 2期
关键词
Electronic portal imaging device; Quality assurance;
D O I
10.1007/BF03178467
中图分类号
学科分类号
摘要
The input/output characteristics of the Wellhofer BIS 710 electronic portal imaging device (EPID) have been investigated to establish its efficacy for periodic quality assurance (QA) applications. Calibration curves have been determined for the energy fluence incident on the detector versus the pixel values. The effect of the charge coupled device (CCD) camera sampling time and beam parameters (such as beam field size, dose rate, photon energy) on the calibration have been investigated for a region of interest (ROI) around the central beam axis. The results demonstrate that the pixel output is a linear function of the incident exposure, as expected for a video-based electronic portal imaging system. The field size effects of the BIS 710 are similar to that of an ion chamber for smaller field sizes up to 10 × 10 cm2. However, for larger field sizes the pixel value increases more rapidly. Furthermore, the system is slightly sensitive to dose rate and is also energy dependent. The BIS 710 has been used in the current study to develop a QA procedure for measurements of flatness and symmetry of a linac x-ray beam. As a two-dimensional image of the radiation field is obtained from a single exposure of the BIS 710, a technique has been developed to calculate flatness and symmetry from a defined radiation area. The flatness and symmetry values obtained are different from those calculated conventionally from major axes only (inplane, crossplane). This demonstrates that the technique can pick up the "cold" and "hot" spots in the analysed area, providing thus more information about the radiation beam. When calibrated against the water tank measurements, the BIS 710 can be used as a secondary device to monitor the x-ray beam flatness and symmetry.
引用
收藏
页码:58 / 66
页数:8
相关论文
共 19 条
  • [1] Van Herk M., Meertens H., A matrix ionization chamber for on-line patient setup verification during radiotherapy, Radiother. Oncol., 11, pp. 369-378, (1988)
  • [2] Bijhold J., Van Herk M., Vijlbrief R., Lebesque J.V., Fast evaluation of patient setup during radiotherapy by aligning features in portal and simulator images, Phys. Med. Biol., 36, pp. 1665-1679, (1991)
  • [3] El A., Vos P.H., Rodrigus P.T.R., Creutzberg C.L., Visser A.G., Stroom J.C., Lebesque J.V., High-precision prostate cancer irradiation by clinical application of an off-line patient setup verification procedure, using portal imaging, Int. J. Radiat. Oncol. Biol. Phys., 35, pp. 321-332, (1996)
  • [4] Meertens H., Van Herk M., Bijhold J., Bartelink H., First clinical experience with a newly developed electronic portal imaging device, Int. J. Radiat. Oncol. Biol. Phys., 18, pp. 1173-1181, (1990)
  • [5] Michalski J.M., Wong J.W., Gerber R.L., Yan D., Cheng A., Graham M.V., Renna M.A., Sawyer P.J., Perez C.A., The use of on-line image verification to estimate the verification in radiation therapy dose delivery, Int. J. Radiat. Oncol. Biol. Phys., 27, pp. 707-716, (1993)
  • [6] Kirby M.C., Williams P.C., The use of an electronic portal imaging device for exit dosimetry and quality control measurement, Int. J. Radiat. Oncol. Biol. Phys., 31, (1995)
  • [7] Beollaard R., Van Herk M., Minjinheer B.J., The dose response relationship of a liquid-filled electronic portal imaging device, Med. Phys., 23, 7, pp. 1601-1611, (1996)
  • [8] McNutt T.R., Mackie T.R., Reckwerdt P., Pawliwal B.R., Modelling dose distributions from portal dose images using the convolution/superposition method, Med. Phys., 23, pp. 1381-1392, (1996)
  • [9] Yin F.F., Schell M.C., Rubin P., A technique of automatic compensator design for lung inhomogeneity correction using an electronic portal imaging device, Med. Phys., 21, pp. 1792-1732, (1994)
  • [10] Roback D.M., Gerbi B.J., Evaluation of electronic portal imaging devices for missing tissue compensator design and verification, Med. Phys., 22, pp. 2029-2034, (1994)