Severe damage of the femoral head after dislocation and difficult reduction maneuvers after total hip arthroplasty

被引:0
作者
Alexander Schuh
Wolfram Mittelmeier
Guenther Zeiler
Detlef Behrend
Joern Kircher
Rainer Bader
机构
[1] Orthopädische Klinik Wichernhaus,Orthopädische Klinik und Poliklinik
[2] Universität Rostock,Institut für Biomedizinische Technik
[3] Universität Rostock,undefined
来源
Archives of Orthopaedic and Trauma Surgery | 2006年 / 126卷
关键词
Total hip arthroplasty; Dislocation; Femoral head; Metallic deposits; Scratches;
D O I
暂无
中图分类号
学科分类号
摘要
Introduction: Aseptic loosening associated with osteolysis and release of wear particles is the main reason for revision of total hip arthroplasty (THA). Damage of femoral heads is well known to increase the wear rate at the articulating surface. Instability and dislocation are serious complications and are the second most frequent reason for revision surgery after THA. In the present work the possible damaging of the femoral head as a result of difficult reduction maneuvers of dislocated THA should be investigated. Patients and methods: In three cases of total hip dislocation an open reduction was performed after dislocation of primary THA. Thereby one metal and two ceramic femoral heads were retrieved. Each head was analyzed macroscopically and by scanning electron microscopy (SEM) subsequently. Results: In the first case the SEM analysis revealed severe titanium deposits on the retrieved ceramic head in the case of repeated unsuccessful closed reduction maneuvers. In the second case the retrieved ceramic head showed multiple scratches and a spalling of the surface structure. In the third case the retrieved metal head presented a roughened surface and severe scratches at the articulating area. Discussion: Difficult reduction maneuvers following total hip dislocation can lead to severe damaging of the femoral head associated with an increase of the surface roughness. While repeated reduction maneuvers can lead to metallic deposits and spalling of the surface of ceramic heads, severe scratches can occur at metal heads. The damaging of the femoral head at the articulating surface can lead to increased wear and subsequent aseptic implant loosening. Therefore, open joint reduction has to be considered in cases of difficult or unsuccessful closed reduction maneuvers to prevent severe femoral head damage which may cause an increased release of wear particles.
引用
收藏
页码:134 / 137
页数:3
相关论文
共 72 条
[1]  
Bader R(2005)Treatment of recurrent dislocation associated with impingement after revision total hip arthroplasty Acta Orthop Belg 71 98-101
[2]  
Barbano R(1993)Polyethylene wear in total hip arthroplasty in patient-matched groups. A comparison of stainless steel, cobalt chrome, and titanium-bearing surfaces J Arthroplasty 8 315-322
[3]  
Mittelmeier W(2001)Prevention of wear and osteolysis: alumina-on-alumina bearing Clin Orthop 393 85-93
[4]  
Bankston AB(2000)Ultra-low wear rates for rigid-on-rigid bearings in total hip replacements Proc Inst Mech Eng 214 331-347
[5]  
Faris PM(2004)Wear of a highly cross-linked polyethylene liner associated with metallic deposition on a ceramic femoral head J Arthroplasty 19 532-536
[6]  
Keating EM(2001)Particle disease. Is tribology a topic in revision surgery? Orthopäde 30 258-265
[7]  
Ritter MA(1995)The influence of scratches to metallic counterfaces on the wear of ultra-high molecular weight polyethylene Proc Inst Mech Eng [H] 209 263-264
[8]  
Bizot P(1997)Wear characteristics of ceramic-on-ceramic for hip endoprostheses Biomaterials 18 873-876
[9]  
Nizard R(2002)Alumina-on-alumina total hip arthroplasty: a minimum 18.5-year follow-up study J Bone Joint Surg Am 84 69-77
[10]  
Hamadouche M(2005)Surface roughness of ceramic femoral heads after in vivo transfer of metal: correlation to polyethylene wear J Bone Joint Surg Am 87 577-582