Thermal stability and transition studies with a hybrid warming device for neonates

被引:6
作者
Greenspan J.S. [1 ]
Cullen A.B. [2 ]
Touch S.M. [1 ]
Wolfson M.R. [2 ]
Shaffer T.H. [2 ]
机构
[1] Thomas Jefferson University Hospital, Thomas Jefferson University, Philadelphia, PA
[2] Temple University, School of Medicine, Philadelphia, PA
关键词
D O I
10.1038/sj.jp.7200513
中图分类号
学科分类号
摘要
Objective: The use of both warmer beds and incubators is common in neonatal intensive care units (NICU), and transferring between these two warming devices is a routine and necessary event. This study was designed to evaluate the efficacy of a new hybrid-warming device, the Versalet™, in transitioning a preterm animal from a warmer bed to an incubator mode and back. Study Design: Nine premature lambs were randomized, following delivery, to receive thermal support from a conventional warming bed and an incubator (control group), or from the Versalet™ (study group) in the warmer bed and incubator modes. Core and various surface temperatures, as well as physiological parameters were measured first during warming in the radiant warmer bed mode, Versalet™ or Resuscitaire™ and then during transition to the incubator mode, Versalet™ or Isolette™, and then back to the warmer bed mode. Results: The animals remained stable during all the transitions. Despite careful planning, adverse events occurred in the control group during transfers. There were no significant differences in the temperature or physiologic profiles during any of the transitions in either group. Conclusion: Compared with the standard warming technique used in NICUs (separate warmer bed and incubator), the Versalet provides similar thermal and cardiovascular stability without adverse events during transition to different modes of warming. The degree to which this device would contribute to ease of management and improved outcomes in humans needs to be evaluated in a clinical trial.
引用
收藏
页码:167 / 173
页数:6
相关论文
共 10 条
[1]  
Bethod P., La couveuse et le gavage a la maternite de Paris, (1887)
[2]  
Bell E.F., Weinstein M.R., Oh W., Heat balance in premature infants: Comparative effects of convectively heated incubator and radiant warmer, with and without plastic heat shield, J Pediatr, 96, 3, pp. 460-465, (1980)
[3]  
Chessex P., Blouet S., Vaucher J., Environmental temperature control in very low birth weight infants (less than 1000 grams) cared for in double-walled incubators, J Pediatr, 113, 2, pp. 373-380, (1988)
[4]  
Wyatt J.S., Thoresen M., Hypothermia treatment and the newborn, Pediatrics, 100, pp. 1028-1030, (1997)
[5]  
Simbruner G., Nanz S., Fleischhacker E., Derganc M., Brain temperature discriminates between neonates with damaged, hypoperfused, and normal brains, Am J Perinatol, 11, pp. 137-143, (1994)
[6]  
Wheldon A.E., Rutter N., The heat balance of small babies nursed in incubators and under radiant warmers, Early Hum Dev, 6, 2, pp. 131-143, (1982)
[7]  
Mok Q., Bass C.A., Et al., Temperature instability during nursing procedures in preterm neonates, Arch Dis Child, 66, 7, pp. 783-786, (1991)
[8]  
Shaffer T.H., Delivoria-Papadopoulos M., Arcinue E., Paez P., Dubois A.B., Pulmonary mechanics in premature lambs during the first few hours of life, Respir Physiol, 28, pp. 179-188, (1976)
[9]  
Shaffer T.H., Wolfson M.R., Forman D., Developmental alterations in pulmonary function of the lamb, Respiration, 47, pp. 129-137, (1985)
[10]  
Wolfson M.R., Kechner N.E., Roache R.F., Roache J.P., Dechandarevian, Friss H.E., Rubenstein S.D., Shaffer T.H., Perfluorochemical rescue of natural surfactant treated preterm lambs: Effect of perflubron dose and ventilatory frequency during gas ventilation, J Appl Physiol, 84, 2, pp. 624-640, (1998)