Comparison of the accuracy of Monte Carlo and Ray Tracing dose calculation algorithms for multiple target brain treatments on CyberKnife

被引:0
|
作者
Godfrey Mukwada
Malgorzata Skorska
Pejman Rowshanfarzad
Martin A. Ebert
机构
[1] Sir Charles Gairdner Hospital,Department of Radiation Oncology
[2] The University of Western Australia,School of Physics, Mathematics and Computing
[3] 5D Clinics,undefined
来源
Physical and Engineering Sciences in Medicine | 2023年 / 46卷
关键词
Multiple brain targets; Monte Carlo; Ray Tracing; CyberKnife;
D O I
暂无
中图分类号
学科分类号
摘要
Single plan multiple brain targets (MBT) stereotactic radiosurgery dose difference between Monte Carlo (MC) and Ray Tracing (RT) algorithms has not been studied. A retrospective study and dose measurements were performed to access factors influencing dose differences. Fifty-three RT treatment plans with a total of 209 brain metastases were extracted from Precision Treatment Planning System (TPS). These plans were generated using fixed cones and were delivered using the CyberKnife M6 system. The same treatment plans were recalculated using MC algorithm and keeping the beam parameters unchanged. MC calculated plan parameters were extracted and dose differences were normalised to MC calculated dose. Correlations were investigated. RT and MC calculated off-centre-ratio (OCR) and tissue-phantom-ratio (TPRs) were exported from the TPS and compared with measured. Plans with 5 gross tumour volumes (GTVs) were created on a phantom and dose measured using a CC04 ionisation chamber and microdiamond detector for comparison with calculated doses. Calculated and measured TPR agreed within ± 1% beyond depth of maximum dose. The OCR showed differences up to 4.3% in the penumbra and out-of-field (OOF) regions. Largest RT and MC calculated GTV mean dose difference was − 5.7%. An increase in the number of GTVs and reduction in the geometric separation of metastases were associated with increased differences between RT and MC calculated doses. In conclusion, calculated dose disagreement in MBT depends on the number of GTVs per plan, number of GTVs within a certain separation distance and plan complexity. MC dose calculation is recommended for complex CyberKnife SRS of MBT.
引用
收藏
页码:1477 / 1487
页数:10
相关论文
共 50 条
  • [1] Comparison of the accuracy of Monte Carlo and Ray Tracing dose calculation algorithms for multiple target brain treatments on CyberKnife
    Mukwada, Godfrey
    Skorska, Malgorzata
    Rowshanfarzad, Pejman
    Ebert, Martin A.
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2023, 46 (04) : 1477 - 1487
  • [2] Comparison of Ray Tracing and Monte Carlo Calculation Algorithms for Spine Lesions Treated With CyberKnife
    Li, Jun
    Zhang, Xile
    Pan, Yuxi
    Zhuang, Hongqing
    Yang, Ruijie
    FRONTIERS IN ONCOLOGY, 2022, 12
  • [3] Comparison of Ray Tracing and Monte Carlo Algorithms for Spine Lesions Treated with CyberKnife
    Li, J.
    Zhang, X.
    Pan, Y.
    Zhuang, H.
    Yang, R.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [4] Assessing accuracy and efficiency of Monte Carlo based dose calculation for MLC CyberKnife treatments
    Mackeprang, P. -H.
    Volken, W.
    Henzen, D.
    Schmidhalter, D.
    Malthaner, M.
    Mueller, S.
    Frei, D.
    Kilby, W.
    Aebersold, D. M.
    Fix, M. K.
    Manser, P.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2019, 195 (06) : 597 - 597
  • [5] Comparison of dose distributions calculated by the cyberknife Monte Carlo and ray tracing algorithms for lung tumors: a phantom study
    Koksal, Canan
    Akbas, Ugur
    Okutan, Murat
    Demir, Bayram
    Sarpun, Ismail Hakki
    THEORETICAL AND EXPERIMENTAL STUDIES IN NUCLEAR APPLICATIONS AND TECHNOLOGY (TESNAT 2015), 2015, 100
  • [6] DOSE CALCULATION ACCURACY OF THE MONTE CARLO ALGORITHM FOR CYBERKNIFE COMPARED WITH OTHER COMMERCIALLY AVAILABLE DOSE CALCULATION ALGORITHMS
    Sharma, Subhash
    Ott, Joseph
    Williams, Jamone
    Dickow, Danny
    MEDICAL DOSIMETRY, 2011, 36 (04) : 347 - 350
  • [7] Comparing the Planned Dose Distributions in CyberKnife Treatments for Intracranial Tumors Between Ray-Tracing and Monte-Carlo Algorithms
    Jiang, W.
    Sharma, S.
    Feng, Y.
    Tonner, B.
    MEDICAL PHYSICS, 2017, 44 (06) : 2967 - 2967
  • [8] Comparison of Monte Carlo and Ray-Tracing Dose Calculation for CyberKnife based Accelerated Partial Breast Irradiation (APBI)
    Xu, Q.
    Chen, Y.
    Grimm, J.
    LaCouture, T.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (02): : S265 - S265
  • [9] Benchmarking Monte-Carlo dose calculation for MLC CyberKnife treatments
    Mackeprang P.-H.
    Vuong D.
    Volken W.
    Henzen D.
    Schmidhalter D.
    Malthaner M.
    Mueller S.
    Frei D.
    Kilby W.
    Aebersold D.M.
    Fix M.K.
    Manser P.
    Radiation Oncology, 14 (1)
  • [10] Comparison of Dose Distributions for Plans Optimized by Using Monte Carlo and Ray-Tracing Algorithms for the Treatment of Lung Tumors with CyberKnife
    Fan, J.
    Hayes, S.
    Jin, L.
    Wang, L.
    Yang, J.
    Ma, C.
    MEDICAL PHYSICS, 2011, 38 (06)