Effect of Gaussian Smoothing Filter Size for CT-Based Attenuation Correction on Quantitative Assessment of Bone SPECT/CT: A Phantom Study

被引:0
|
作者
Yoya Tomita
Yasutaka Ichikawa
Kengo Hashizume
Hajime Sakuma
机构
[1] Mie University Hospital,Central Division of Radiology
[2] Mie University Hospital,Department of Radiology
关键词
Gaussian filter; CT-based attenuation correction; Attenuation coefficient; Bone quantitative single photon emission computed tomography/computed tomography (SPECT/CT);
D O I
暂无
中图分类号
学科分类号
摘要
This study aims to determine the effect of Gaussian filter size for CT-based attenuation correction (CTAC) on the quantitative assessment of bone SPECT. An experiment was performed using a cylindrical phantom containing six rods, of which one was filled with water and five were filled with various concentrations of K2HPO4 solution (120–960 mg/cm3) to simulate different bone densities. 99mTc-solution of 207 kBq/ml was also included within the rods. SPECT data were acquired at 120 views for 30 s/view. CT for attenuation correction were obtained at 120 kVp and 100 mA. Sixteen different CTAC maps processed with different Gaussian filter sizes (ranging from 0 to 30 mm in 2 mm increments) were generated. SPECT images were reconstructed for each of the 16 CTAC maps. Attenuation coefficients and radioactivity concentrations in the rods were compared with those in the water-filled rod without K2HPO4 solution as a reference. Gaussian filter sizes below 14–16 mm resulted in an overestimation of radioactivity concentrations for rods with high concentrations of K2HPO4 (≥ 666 mg/cm3). The overestimation of radioactivity concentration measurement was 3.8% and 5.5% for 666 mg/cm3 and 960 mg/cm3 K2HPO4 solutions, respectively. The difference in radioactivity concentration between the water rod and the K2HPO4 rods was minimal at 18–22 mm. The use of Gaussian filter sizes smaller than 14–16 mm caused an overestimation of radioactivity concentration in regions of high CT values. Setting the Gaussian filter size to 18–22 mm enables radioactivity concentration to be measured with the least influence on bone density.
引用
收藏
页码:2313 / 2321
页数:8
相关论文
共 50 条
  • [1] Effect of Gaussian Smoothing Filter Size for CT-Based Attenuation Correction on Quantitative Assessment of Bone SPECT/CT: A Phantom Study
    Tomita, Yoya
    Ichikawa, Yasutaka
    Hashizume, Kengo
    Sakuma, Hajime
    JOURNAL OF DIGITAL IMAGING, 2023, 36 (05) : 2313 - 2321
  • [2] Effect of Gaussian smoothing filter for CT-based attenuation correction mapon quantitative assessment in bone SPECT/CT: a phantom study
    Tomita, Yoya
    Ichikawa, Yasutaka
    Hashizume, Kengo
    Kamigiri, Akira
    Yamada, Tsuyoshi
    Maki, Hiroaki
    Sakuma, Hajime
    JOURNAL OF NUCLEAR MEDICINE, 2018, 59
  • [3] CT-based SPECT attenuation correction and assessment of infarct size: results from a cardiac phantom study
    Alexander Stephan Kroiss
    Stephan Gerhard Nekolla
    Georg Dobrozemsky
    Thomas Grubinger
    Barry Lynn Shulkin
    Markus Schwaiger
    Annals of Nuclear Medicine, 2017, 31 : 764 - 772
  • [4] CT-based SPECT attenuation correction and assessment of infarct size: results from a cardiac phantom study
    Kroiss, Alexander Stephan
    Nekolla, Stephan Gerhard
    Dobrozemsky, Georg
    Grubinger, Thomas
    Shulkin, Barry Lynn
    Schwaiger, Markus
    ANNALS OF NUCLEAR MEDICINE, 2017, 31 (10) : 764 - 772
  • [5] Effects of CT-based attenuation correction on quantitative SPECT/CT of jawbone
    Tsuchitani, Tatsuya
    Kitajima, Kazuhiro
    Takahashi, Yoshiyuki
    Yoshida, Rika
    Kiada, Hayato
    Tsurusaki, Masakatsu
    Minami, Tetsuya
    Kotura, Noriko
    HELLENIC JOURNAL OF NUCLEAR MEDICINE, 2023, 26 (02) : 108 - 113
  • [6] Periprosthetic artifacts on SPECT images processed by CT-based attenuation correction: a phantom study
    Nishimura, Takuma
    Tomita, Yoya
    Ichikawa, Yasutaka
    Maki, Hiroaki
    Sakuma, Hajime
    JOURNAL OF NUCLEAR MEDICINE, 2019, 60
  • [7] Effect of the difference in tube voltage on attenuation coefficients and quantitative assessment of bone SPECT/CT: a phantom study
    Hashizume, Kengo
    Tomita, Yoya
    Ichikawa, Yasutaka
    Kamigiri, Akira
    Yamada, Tsuyoshi
    Kubooka, Naoya
    Maki, Hiroaki
    Sakuma, Hajime
    JOURNAL OF NUCLEAR MEDICINE, 2018, 59
  • [8] Implementation and performance evaluation of CT-based scatter and attenuation correction in SPECT: A phantom study.
    Yokoi, T
    Takahashi, M
    Kishi, H
    Inaoka, Y
    Kitamura, K
    Yamamoto, T
    Shinohara, H
    JOURNAL OF NUCLEAR MEDICINE, 2002, 43 (05) : 220P - 221P
  • [9] Effect of Kilovoltage and Quality Reference mAs on CT-Based Attenuation Correction in 177Lu SPECT/CT Imaging: a Phantom Study
    Ramirez, M. Salas
    Lassmann, M.
    Tran-Gia, J.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2022, 49 (SUPPL 1) : S265 - S265
  • [10] Effect of kilovoltage and quality reference mAs on CT-based attenuation correction in 177Lu SPECT/CT imaging: a phantom study
    Salas-Ramirez, Maikol
    Leube, Julian
    Lassmann, Michael
    Tran-Gia, Johannes
    EJNMMI PHYSICS, 2024, 11 (01)