Twelve years of experience with the ATS mechanical heart valve prostheses

被引:7
|
作者
Taniguchi S. [1 ]
Hashizume K. [1 ]
Ariyoshi T. [1 ]
Hisata Y. [1 ]
Tanigawa K. [1 ]
Miura T. [1 ]
Odate T. [1 ]
Matsukuma S. [1 ]
Nakaji S. [1 ]
Eishi K. [1 ]
机构
[1] Department of Cardiovascular Surgery, Graduate School of Medicine, Nagasaki University, Nagasaki 852-8501
关键词
ATS mechanical valve; Morbidity; Mortality; Valve replacement;
D O I
10.1007/s11748-012-0124-1
中图分类号
学科分类号
摘要
Objective The ATS Open Pivot mechanical heart valve has been implanted routinely at our institution since 1999. The objective of this study is to retrospectively analyze our 12-year clinical results with ATS prostheses. Methods ATS Open Pivot mechanical valves were implanted in 268 adult patients between May 1999 and August 2010. We selected 259 subjects who could be adequately followed (follow-up rate 96.6 %). Aortic valve replacement was performed in 157 patients, mitral valve replacement (MVR) in 71, and double (aortic and mitral) valve replacements (DVR) in 31. Mean age at the time of implant was 58.8 ± 10.6 years. The gender ratio was 128 males/131 females. Mean follow-up was 4.4 ± 7.8 years, and the cumulative follow-up was 1144 patient-years (pt-yr). Results Early death within 30 days after the operation occurred in 5 (2.5 %) patients. Late death occurred in 27 patients including valve-related deaths in 13. The 10-year survival rate after the operation was 82.7 ± 2.9 %. The rate of freedom from valve-related death was 92.2 ± 2.2 %. The incidence of valve-related complications was 2.19 %/pt-yr. Of these, the incidence of thromboembolic events and that of bleeding complications were 1.22 and that 0.87 %/pt-yr. The incidence of valve thrombosis was 0.09 %/pt-yr. No structural valve deterioration was observed in any of the three operative procedure groups. Conclusions Our 12-year experience with aortic and MVR using the ATS mechanical heart valve demonstrated low incidences of thromboembolic events, bleeding complications, and valve thrombosis. © The Japanese Association for Thoracic Surgery 2012.
引用
收藏
页码:561 / 568
页数:7
相关论文
共 50 条
  • [41] MECHANICAL DISINSERTIONS OF VALVE PROSTHESES
    CORMIER, B
    GUERINON, J
    BERNARD, M
    HOUDART, E
    LAFLECHE, A
    ACAR, J
    ANNALES DE CHIRURGIE, 1987, 41 (02): : 130 - 134
  • [42] HEART VALVE PROSTHESES AND PREGNANCY
    SOMMER, A
    ROMANO, JP
    GOUDARD, JR
    JOUVE, A
    SERMENT, H
    ANNALES DE CARDIOLOGIE ET D ANGEIOLOGIE, 1973, 22 (03): : 183 - 187
  • [43] Effect of valve welder flexibility on cavitation initiation with mechanical heart valve prostheses: An in vitro study
    Lee, CS
    Aluri, S
    Chandran, KB
    JOURNAL OF HEART VALVE DISEASE, 1996, 5 (01): : 104 - 113
  • [44] Cardiac valve replacement with mechanical prostheses in patients aged 65 years and over
    Aksöyek, A
    Ulus, AT
    Tütün, U
    Budak, B
    Parlar, A
    Korkmaz, K
    Çolak, A
    Katircioglu, SF
    JOURNAL OF HEART VALVE DISEASE, 2004, 13 (04): : 641 - 650
  • [45] Flexible fiberoptic bronchoscopy in children with heart diseases:: A twelve years experience
    Cerda, Jaime
    Chacon, Jorge
    Reichhard, Cristina
    Bertrand, Pablo
    Holmgren, Nils L.
    Claveria, Cristian
    Sanchez, Ignacio
    PEDIATRIC PULMONOLOGY, 2007, 42 (04) : 319 - 324
  • [46] COMPARISON OF PORCINE VALVE XENOGRAFTS WITH MECHANICAL PROSTHESES - A 71/2 YEAR EXPERIENCE
    JOYCE, LD
    NELSON, RM
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1984, 88 (01): : 102 - 113
  • [47] New anticoagulants: the disappointed hopes of dabigatran in patients with mechanical heart valve prostheses
    Danchin, N.
    ANNALES DE CARDIOLOGIE ET D ANGEIOLOGIE, 2013, 62 : 10 - 11
  • [48] SUPPRESSION OF COAGULATION BY ORAL ANTICOAGULANTS IN PATIENTS WITH MECHANICAL HEART-VALVE PROSTHESES
    IWADE, K
    AOSAKI, M
    OHKI, K
    UETSUKA, Y
    HOKARI, T
    SAITO, K
    NAKANO, K
    HOSODA, S
    THROMBOSIS AND HAEMOSTASIS, 1993, 69 (06) : 677 - 677
  • [49] Mechanical Load Analysis of Transcatheter Heart Valve Prostheses in Accelerated Wear Testers
    Roggenkamp, Jan
    Struth, Alexander
    Schmitz, Christoph
    Schmitz-Rode, Thomas
    Steinseifer, Ulrich
    CARDIOLOGY, 2016, 134 (02) : 246 - 247
  • [50] DYNAMIC ANALYSIS OF FLUTTER IN DISK TYPE MECHANICAL HEART-VALVE PROSTHESES
    PRABHU, AA
    HWANG, NHC
    JOURNAL OF BIOMECHANICS, 1988, 21 (07) : 585 - 590