Evaluation of the Long-Term Stability and Temperature Coefficient of Dew-Point Hygrometers

被引:0
作者
R. Benyon
T. Vicente
P. Hernández
L. De Rivas
F. Conde
机构
[1] Instituto Nacional de Técnica Aeroespacial,
[2] Torrejón de Ardoz,undefined
来源
International Journal of Thermophysics | 2012年 / 33卷
关键词
Calibration; Dew-point temperature; Drift; Humidity; Hygrometer; Traceability;
D O I
暂无
中图分类号
学科分类号
摘要
The continuous quest for improved specifications of optical dew-point hygrometers has raised customer expectations on the performance of these devices. In the absence of a long calibration history, users with a limited prior experience in the measurement of humidity, place reliance on manufacturer specifications to estimate long-term stability. While this might be reasonable in the case of measurement of electrical quantities, in humidity it can lead to optimistic estimations of uncertainty. This article reports a study of the long-term stability of some hygrometers and the analysis of their performance as monitored through regular calibration. The results of the investigations provide some typical, realistic uncertainties associated with the long-term stability of instruments used in calibration and testing laboratories. Together, these uncertainties can help in establishing initial contributions in uncertainty budgets, as well as in setting the minimum calibration requirements, based on the evaluation of dominant influence quantities.
引用
收藏
页码:1758 / 1770
页数:12
相关论文
共 2 条
[1]  
Benyon R.(2008)undefined Int. J. Thermophys. 29 1623-undefined
[2]  
Mitter H.(undefined)undefined undefined undefined undefined-undefined