Cavitation erosion in bloods

被引:0
作者
Jian-hua Wu
Yu Wang
Fei Ma
Wen-juan Gou
机构
[1] Hohai University,College of Water Conservancy and Hydropower Engineering
[2] Tianjin University,State Key Laboratory of Hydraulic Engineering Simulation and Safety
来源
Journal of Hydrodynamics | 2017年 / 29卷
关键词
Cavitation erosion; blood; concentration; temperature; in vitro;
D O I
暂无
中图分类号
学科分类号
摘要
The cavitation in a mechanical heart valve (MHV) is a serious concern. In most of the investigations of the MHV cavitation in vitro, the tap water, the distilled water, or the glycerin are used as the test liquids, instead of the real blood. Therefore, the effects of the liquid properties on the cavitation can not be well revealed. In this paper, the cavitation erosion in the porcine bloods is experimentally investigated as well as in the tap water and the distilled water by means of a vibratory apparatus. The results show that the blood produces a weaker intensity of the cavitation erosion than the tap water or the distilled water. The cavitation erosion decreases with the decrease of the liquid temperature or with the increase of the concentration of the blood, especially with the increase of the liquid viscosity. It is the viscosity that could be a major dominant factor affecting this erosion. The temperature or the concentration of the blood changes the viscosity, and in turns changes the intensity of the cavitation erosion.
引用
收藏
页码:724 / 727
页数:3
相关论文
共 44 条
[1]  
DeWall R A(2000)Evolution of mechanical heart valves [J] Annals of Thoracic Surgery 69 1612-1621
[2]  
Qasim N(1989)Leaflet fracture in Edwards-Duromedics bileaflet valves [J] Journal of Thoracic and Cardiovascular Surgery 97 90-94
[3]  
Carr L(2003)Leaflet escape in a new bileaflet mechanical valve: Tri technologies [J] Circulation 107 2303-2306
[4]  
Klepetko W(2000)Duromedics original prosthesis: What do we really know about diagnosis and mechanism of leaflet escape? [J] Canadian Journal of Cardiology 16 1269-1272
[5]  
Moritz A(2004)Mechanical heart valve cavitation [J] Expert Review of Medical Devices 1 95-104
[6]  
Mlczoch J(2004)Mechanical heart valve cavitation: Valve specific parameters [J] The International Journal of Artificial Organs 27 855-867
[7]  
Bottio T(2009)A novel study of mechanical heart valve cavitation in a pressurized pulsatile duplicator [J] Asaio Journal 55 455-461
[8]  
Casarotto D(2012)Role of vortices in cavitation formation in the flow at the closure of a bilea-flet mitral mechanical heart valve [J] Journal of Artificial Organs 15 57-64
[9]  
Thiene G(1994)Cavitation damage of pyrolytic carbon in mechanical heart valves [J] Journal of Heart Valve Disease 3 2-7
[10]  
Mastroroberto P(2003)Bubble formation on St. Jude Medical mechanical heart valve: An in vitro study [J] Journal of Heart Valve Disease 12 406-410