Bone plug length and loading angle determine the primary stability of patellar tendon-bone grafts in press-fit ACL reconstruction

被引:0
作者
R. Schmidt-Wiethoff
J. Dargel
M. Gerstner
T. Schneider
J. Koebke
机构
[1] Clinic for Orthopedics and Sports Traumatology,Department of Biomechanics and Orthopaedics
[2] German Sport University of Cologne,Institute II for Anatomy
[3] University of Cologne,undefined
来源
Knee Surgery, Sports Traumatology, Arthroscopy | 2006年 / 14卷
关键词
Knee; ACL reconstruction; Patellar tendon; Graft fixation; Press-fit;
D O I
暂无
中图分类号
学科分类号
摘要
The present study was conducted to evaluate the influence of bone-plug length on the primary stability of patellar tendon-bone grafts, using a femoral press-fit fixation technique. Forty-eight human PTB grafts with a patellar bone-plug length of 15 mm and 25 mm were obtained from 24 human cadavers (mean age 72 years) and implanted to porcine femora in a press-fit fixation technique. Tensile loading was performed at 10 mm/s until failure at varying loading angles of 0°, 30° and 60°. Compared to 25 mm, a significant decrease of primary graft stability was recorded in the testing of 15-mm bone plugs. For both plug lengths, the ultimate load to failure increased with rising the loading angles. While axial graft loading exclusively caused plug dislocation, the predominant mode of failure was tendon rupture at 60° loading angle. We conclude that 15-mm bone plugs do not result in a sufficient stability for early aggressive rehabilitation and therefore 25-mm bone plugs are recommended for the femoral press-fit technique.
引用
收藏
页码:108 / 111
页数:3
相关论文
共 76 条
  • [1] Aune AK(1998)Interference screw fixation of hamstring vs patellar tendon grafts for anterior cruciate ligament reconstruction Knee Surg Sports Traumatol 6 99-102
  • [2] Ekeland A(1989)Potential pitfalls of Kurosaka screw interference fixation for ACL surgery Am J Knee Surg 2 76-82
  • [3] Cowley PW(1994)The effects of donor age and strain rate on the biomschanical properties of bone-patellar tendon-bone allografts Am J Sports Med 22 328-333
  • [4] Bach RB(1997)Arthroskopische femorale Press-fit-Fixation des Lig.-patellae-Transplantats beim Ersatz des vorderen Kreuzbands Arthroskopie 10 126-132
  • [5] Blevins FT(2000)Graft fixation in cruciate ligament reconstruction Am J Sports Med 28 761-774
  • [6] Hecker AT(1999)All-Press-Fit, eine neue Operationsmethode zum vorderen Kreuzbandersatz mit gleichzeitiger femoraler und tibialer Press-fit-Verankerung Arthroskopie 12 299-304
  • [7] Bigler GT(2002)Patellar tendon anterior cruciate ligament reconstruction with conical press-fit femoral fixation: 5-year results in athletes population Knee Surg Sports Traumatol Arthrosc 10 73-79
  • [8] Boland AL(1985)Complications of anterior cruciate ligament surgery Orthop Clin North Am 16 237-240
  • [9] Hayes WC(1993)Effect of varying angles on pullout strength of interference screw fixation Arthroscopy 9 417-424
  • [10] Boszotta H(2003)Analysis of initial fixation strength of press-fit fixation technique in anterior cruciate ligament reconstruction Knee Surg Sports Traumatol Arthrosc 11 91-98