Radiation dose measurements of an on-board imager X-ray unit using optically-stimulated luminescence dosimeters

被引:0
|
作者
Leon Smith
Mamoon Haque
Johnny Morales
Robin Hill
机构
[1] The University of Sydney,Sydney Medical School
[2] Chris O’Brien Lifehouse,Department of Radiation Oncology
[3] The University of Sydney,School of Physics, Institute of Medical Physics
来源
Australasian Physical & Engineering Sciences in Medicine | 2015年 / 38卷
关键词
Optically stimulated luminescence; Dosimetry; Cone beam computed tomography; Radiotherapy; On board imaging; Kilovoltage dosimetry;
D O I
暂无
中图分类号
学科分类号
摘要
Cone beam computed tomography (CBCT) is now widely used to image radiotherapy patients prior to treatment for the purpose of accurate patient setup. However each CBCT image delivered to a patient increases the total radiation dose that they receive. The measurement of the dose delivered from the CBCT images is not readily performed in the clinic. In this study, we have used commercially available optically stimulated luminescence (OSLD) dosimeters to measure the dose delivered by the Varian OBI on a radiotherapy linear accelerator. Calibration of the OSLDs was achieved by using a therapeutic X-ray unit. The dose delivered by a head CBCT scan was found to be 3.2 ± 0.3 mGy which is similar in magnitude to the dose of a head computed tomography (CT) scan. The results of this study suggest that the radiation hazard associated with CBCT is of a similar nature to that of conventional CT scans. We have also demonstrated that the OSLDs are suitable for these low X-ray dose measurements.
引用
收藏
页码:665 / 669
页数:4
相关论文
共 32 条
  • [21] Evaluation of a real-time optically stimulated luminescence beryllium oxide (BeO) fibre-coupled dosimetry system with a superficial 140 kVp X-ray beam
    Santos, Alexandre M. Caraca
    Gowda, Raghu
    Bezak, Eva
    Afshar, Shahraam, V
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 65 : 167 - 171
  • [22] Online and offline Radiation-Induced Attenuation measurements on FD-7 radiophotoluminescence dosimeters irradiated at high X-ray doses
    Hasan, A.
    Aguiar, Y.
    Alia, R. Garcia
    Campanella, C.
    Morana, A.
    Alem, A. K.
    Girard, S.
    Mandal, A. Raj
    Ferrari, M.
    RADIATION MEASUREMENTS, 2024, 177
  • [23] Demonstration of 2D dosimetry using Al2O3 optically stimulated luminescence films for therapeutic megavoltage x-ray and ion beams
    Ahmed, M. F.
    Shrestha, N.
    Ahmad, S.
    Schnell, E.
    Akselrod, M. S.
    Yukihara, E. G.
    RADIATION MEASUREMENTS, 2017, 106 : 315 - 320
  • [24] Development of an automated system equipped with a small X-ray irradiator for red/blue thermoluminescence and optically stimulated luminescence measurement from natural minerals
    Hashimoto, T
    Nakagawa, T
    Usuda, H
    Yawata, T
    BUNSEKI KAGAKU, 2002, 51 (08) : 625 - 632
  • [25] Assessment of the radiation dose associated with X-ray examinations using the DoseCal and CALDose_X software packages
    Saeed, M. K.
    Almalki, Y.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2022, 20 (01): : 191 - 197
  • [26] Significant Radiation Dose Reduction in the Hybrid Operating Room Using a Novel X-ray Imaging Technology
    van den Haak, R. F. F.
    Hamans, B. C.
    Zuurmond, K.
    Verhoeven, B. A. N.
    Koning, O. H. J.
    EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2015, 50 (04) : 480 - 486
  • [27] Estimation of dose and elapsed time after unrecognized high-dose radiation exposure using the continuous-wave optically stimulated luminescence from Mg2SiO4:Tb
    Yasuda, Hiroshi
    Discher, Michael
    RADIATION MEASUREMENTS, 2020, 139
  • [28] Assessment of Patient Radiation Dose during Conventional Diagnostic X-ray Examinations in Three Public Hospitals in Northern Jordan using TLDs
    Ababneh, Zaid Q.
    Ababneh, Anas M.
    Hani, Majd S. Bani
    Ababneh, Riad S.
    Aljarrah, Khaled M.
    HEALTH PHYSICS, 2020, 118 (06): : 593 - 599
  • [29] Studies of X-ray Induced Thermally Stimulated Luminescence Properties of Mn2+-Doped MgSO4 by Using Glow-Curve Deconvolution Technique
    Manam, J.
    Das, S.
    PENETRATING RADIATION SYSTEMS AND APPLICATIONS XI, 2010, 7806
  • [30] Practical dosimetry procedure of air kerma for kilovoltage X-ray imaging in radiation oncology using a 0.6-cc cylindrical ionization chamber with a cobalt absorbed dose-to-water calibration coefficient
    Tachibana, Hidenobu
    Takahashi, Ryo
    Kogure, Takayuki
    Nishiyama, Shiro
    Kurosawa, Tomoyuki
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2022, 15 (03) : 264 - 270