Early aortic dissection after aortic valve replacement with a Perceval sutureless bioprosthesis

被引:0
作者
Giosuè Falcetta
Stefano Pratali
Giacomo Ravenni
Uberto Bortolotti
机构
[1] University Hospital of Pisa,Section of Cardiac Surgery, Cardiothoracic and Vascular Department
来源
Indian Journal of Thoracic and Cardiovascular Surgery | 2019年 / 35卷
关键词
Aortic valve replacement; Aortic dissection; Sutureless bioprostheses;
D O I
暂无
中图分类号
学科分类号
摘要
We report a patient who developed ascending aorta dissection after aortic valve replacement with a Perceval sutureless pericardial bioprosthesis, a previously not reported complication with this device. At reoperation, the intimal tear was located at the aortotomy suture line; the tips of the nitinol stent were far from the aortic lesion which therefore was not attributable to the presence of this specific prosthetic model. Nevertheless, a previous report of aortic dissection due to erosion of the aortic wall by the strut of a transcatheter prosthesis may raise the suspicion that such complication could also occur with sutureless prostheses of similar design.
引用
收藏
页码:510 / 512
页数:2
相关论文
共 33 条
[1]  
Shrestha M(2014)Aortic valve replacement and concomitant procedures with the Perceval valve: results of European trials Ann Thorac Surg 98 1294-1300
[2]  
Folliguet TA(2016)European multicentre experience with sutureless Perceval valve: clinical and haemodynamic outcomes up to 5 years in over 700 patients Eur J Cardio-thorac Surg 49 234-241
[3]  
Pfeiffer S(1998)Ascending aorta dissection after aortic valve replacement J Heart Valve Dis 7 75-80
[4]  
Shrestra M(2012)A supra-annular malposition of the Perceval S sutureless aortic valve: the ‘ℵ-movement’ removal technique and subsequent reimplantation Interact Cardiovasc Thorac Surg 15 280-281
[5]  
Fischlein T(2016)Clinical and haemodynamic outcomes in 658 patients receiving the Perceval sutureless aortic valve: early results from a prospective European multicentre study (the Cavalier Trial) Eur J Cardiothorac Surg 49 978-986
[6]  
Meuris B(2016)Sutureless, rapid deployment valves and stented bioprosthesis in aortic valve replacement: recommendations of an International Expert Consensus Panel Eur J Cardiothorac Surg 49 709-718
[7]  
Milano A(1993)Risk of aortic dissection after aortic valve replacement Am J Cardiol 72 1043-1047
[8]  
Pratali S(2013)Delayed aortic dissection after transcatheter aortic valve implantation J Heart Valve Dis 22 701-703
[9]  
De Carlo M(undefined)undefined undefined undefined undefined-undefined
[10]  
Borzoni G(undefined)undefined undefined undefined undefined-undefined