Biomechanical finite element analysis of bone cemented hip crack initiation according to stem design

被引:0
|
作者
Byeongsoo Kim
Byungyoung Moon
Junghong Park
机构
[1] Inje University,Department of Mechanical and Automobile Engineering
[2] Pusan National University,Industrial Liaison Innovation Cluster
[3] Pusan National University,Department of Mechanical Design Engineering
来源
Journal of Mechanical Science and Technology | 2006年 / 20卷
关键词
Total Hip Replacement; Bone Cement; Femoral Stem; Crack Propagation; Finite Element Method; Biomechanics;
D O I
暂无
中图分类号
学科分类号
摘要
The purpose of this investigation was to determine the specific fracture mechanics response of cracks that initiate at the stem-cement interface and propagate into the cement mantle. Two-dimensional finite element models of idealized stem-cement-bone cross-sections from the proximal femur were developed for this study. Two general stem types were considered; Rectangular shape and Charnley type stem designs. The FE results showed that the highest principal stress in the cement mantle for each case occurred in the upper left and lower right regions adjacent to the stem-cement interface. There was also a general decrease in maximum tensile stress with increasing cement mantle thickness for both Rectangular and Charnley-type stem designs. The cement thickness is found to be one of the important fatigue failure parameters which affect the longevity of cemented femoral components, in which the thinner cement was significantly associated with early mechanical failure for shot-time period.
引用
收藏
页码:2168 / 2177
页数:9
相关论文
共 50 条
  • [41] A computational tool for the probabilistic finite element analysis of an uncemented total hip replacement considering variability in bone-implant version angle
    Dopico-Gonzalez, Carolina
    New, Andrew M.
    Browne, Martin
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2010, 13 (01) : 1 - 9
  • [42] Biomechanical aspects of the insect wing: an analysis using the finite element method
    Kesel, AB
    Philippi, U
    Nachtigall, W
    COMPUTERS IN BIOLOGY AND MEDICINE, 1998, 28 (04) : 423 - 437
  • [43] Biomechanical analysis and design of a dynamic spinal fixator using topology optimization: a finite element analysis
    Lin, Hung-Ming
    Liu, Chien-Lin
    Pan, Yung-Ning
    Huang, Chang-Hung
    Shih, Shih-Liang
    Wei, Shun-Hwa
    Chen, Chen-Sheng
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2014, 52 (05) : 499 - 508
  • [44] Finite element analysis of a novel design approach to resisting total hip dislocation
    Scifert, CF
    Brown, TD
    Lipman, JD
    CLINICAL BIOMECHANICS, 1999, 14 (10) : 697 - 703
  • [45] Optimizing Stem Length in Conversion Total Hip Arthroplasty: An Expanded Finite Element Analysis
    Shimasaki, Koshiro
    Nishino, Tomofumi
    Yoshizawa, Tomohiro
    Watanabe, Ryunosuke
    Hirose, Fumi
    Yasunaga, Shota
    Mishima, Hajime
    JOURNAL OF CLINICAL MEDICINE, 2025, 14 (04)
  • [46] Use of a Genetic Algorithm for Multiobjective Design Optimization of the Femoral Stem of a Cemented Total Hip Arthroplasty
    Ishida, Toshimasa
    Nishimura, Ikuya
    Tanino, Hiromasa
    Higa, Masaru
    Ito, Hiroshi
    Mitamura, Yoshinori
    ARTIFICIAL ORGANS, 2011, 35 (04) : 404 - 410
  • [47] Cemented fixation of the acetabular cup in total hip replacement: Improving the distribution of anchorage holes by finite element method
    Mootanah, R
    New, A
    Pearmain, D
    Ingle, P
    Dowell, J
    Cheah, K
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING - 3, 2001, : 33 - 38
  • [48] Design and non-linear finite element analysis of titanium-based femoral hip-stem for Indian population
    Ankit D. Oza
    Nakul Gupta
    Rajesh Singh
    International Journal on Interactive Design and Manufacturing (IJIDeM), 2023, 17 : 2489 - 2493
  • [49] Design and non-linear finite element analysis of titanium-based femoral hip-stem for Indian population
    Oza, Ankit D.
    Gupta, Nakul
    Singh, Rajesh
    INTERNATIONAL JOURNAL OF INTERACTIVE DESIGN AND MANUFACTURING - IJIDEM, 2023, 17 (05): : 2489 - 2493
  • [50] Biomechanical analysis of fixation methods for bone flap repositioning after lateral orbitotomy approach: A finite element analysis
    Sang, Zexi
    Ren, Zhangjun
    Yu, Jinhai
    Wang, Yaohua
    Liao, Hongfei
    JOURNAL OF STOMATOLOGY ORAL AND MAXILLOFACIAL SURGERY, 2024, 125 (05)