Prediction of femoral strength using 3D finite element models reconstructed from DXA images: validation against experiments

被引:0
作者
Lorenzo Grassi
Sami P. Väänänen
Matti Ristinmaa
Jukka S. Jurvelin
Hanna Isaksson
机构
[1] Lund University,Department of Biomedical Engineering
[2] University of Eastern Finland,Department of Applied Physics
[3] Kuopio University Hospital,Department of Orthopaedics, Traumatology and Hand Surgery
[4] Lund University,Division of Solid Mechanics
[5] Kuopio University Hospital,Diagnostic Imaging Center
来源
Biomechanics and Modeling in Mechanobiology | 2017年 / 16卷
关键词
Statistical shape model; Statistical appearance model; Finite element; Proximal femur; Validation;
D O I
暂无
中图分类号
学科分类号
摘要
Computed tomography (CT)-based finite element (FE) models may improve the current osteoporosis diagnostics and prediction of fracture risk by providing an estimate for femoral strength. However, the need for a CT scan, as opposed to the conventional use of dual-energy X-ray absorptiometry (DXA) for osteoporosis diagnostics, is considered a major obstacle. The 3D shape and bone mineral density (BMD) distribution of a femur can be reconstructed using a statistical shape and appearance model (SSAM) and the DXA image of the femur. Then, the reconstructed shape and BMD could be used to build FE models to predict bone strength. Since high accuracy is needed in all steps of the analysis, this study aimed at evaluating the ability of a 3D FE model built from one 2D DXA image to predict the strains and fracture load of human femora. Three cadaver femora were retrieved, for which experimental measurements from ex vivo mechanical tests were available. FE models were built using the SSAM-based reconstructions: using only the SSAM-reconstructed shape, only the SSAM-reconstructed BMD distribution, and the full SSAM-based reconstruction (including both shape and BMD distribution). When compared with experimental data, the SSAM-based models predicted accurately principal strains (coefficient of determination >0.83, normalized root-mean-square error <16%) and femoral strength (standard error of the estimate 1215 N). These results were only slightly inferior to those obtained with CT-based FE models, but with the considerable advantage of the models being built from DXA images. In summary, the results support the feasibility of SSAM-based models as a practical tool to introduce FE-based bone strength estimation in the current fracture risk diagnostics.
引用
收藏
页码:989 / 1000
页数:11
相关论文
共 117 条
[1]  
Bayraktar HH(2004)Comparison of the elastic and yield properties of human femoral trabecular and cortical bone tissue J Biomech 37 27-35
[2]  
Morgan EF(2014)Image-based vs. mesh-based statistical appearance models of the human femur: implications for finite element simulations Med Eng Phys 36 1626-1635
[3]  
Niebur GL(2009)Use of a statistical model of the whole femur in a large scale, multi-model study of femoral neck fracture risk J Biomech 42 2171-2176
[4]  
Bonaretti S(2007)Incidence and economic burden of osteoporosis-related fractures in the United States, 2005–2025 J Bone Miner Res 22 465-475
[5]  
Seiler C(2014)Statistical shape and appearance models in osteoporosis Curr Osteoporos Rep 12 163-173
[6]  
Boichon C(1999)Femoral strength is better predicted by finite element models than QCT and DXA J Biomech 32 1013-1020
[7]  
Bryan R(2006)BMD and risk of hip and nonvertebral fractures in older men: a prospective study and comparison with older women J Bone Miner Res 21 1550-1556
[8]  
Nair PB(2016)Experimental validation of DXA-based finite element models for prediction of femoral strength J Mech Behav Biomed Mater 63 17-25
[9]  
Taylor M(2015)Clinical use of quantitative computed tomography (QCT) of the hip in the management of osteoporosis in adults: the 2015 ISCD official positions-Part I J Clin Densitom 18 338-358
[10]  
Burge R(2014)Comprehensive evaluation of PCA-based finite element modelling of the human femur Med Eng Phys 36 1246-1252