Homogeneous and Heterogeneous Modeling of Patient-Specific Hip Implant Under Static and Dynamic Loading Condition Using Finite Element Analysis

被引:5
|
作者
Ravikant [1 ]
Mittal V.K. [1 ]
Gupta V. [2 ]
机构
[1] Department of Mechanical Engineering, National Institute of Technology, Kurukshetra
[2] Department of Mechanical Engineering, Ch. Devi Lal State Institute of Engineering & Technology, Panniwala Mota
基金
英国科研创新办公室;
关键词
Finite element analysis; Gait motion; Hip prosthesis; Implant; Stress shielding;
D O I
10.1007/s40033-023-00447-0
中图分类号
学科分类号
摘要
Stress and strain shielding are the most common phenomenon shown by the metallic implant which reduces the implant life due to modulus mismatch. It is necessary to have a sufficient knowledge about the induced stress and strain distribution in order to find the suitable implant design with appropriate material combinations. In the present work, a three-dimensional implant model has been developed using real-life patient-specific computed tomographic (CT) data. Two different material models, viz. homogeneous (Ti–6Al–4V) and heterogeneous (Ti–6Al–4V and Ti-35Nb–5Ta-7Zr–0.4O), have been taken into account to demonstrate their mechanical properties like stress, strain and deformation. In addition, the boundary condition has been applied at distal end using ASTM F2996-1 and two different types of loading conditions have been applied (1) static loading using ISO 7206–4:2010(E) standard and (2) dynamic loading using patient-specific constraints, i.e., forces and torsional moment. For dynamic analysis, a patient-specific walking and going downstairs gait motions have been considered to obtain the essential mechanical characteristics. The final outcome reveals that heterogeneous model demonstrates lower von mises stress and approximate similar strain value compared to homogeneous model. It has been observed that the heterogeneous material model demonstrates less von mises stress at peak curve, i.e., 4.7% and 18.81% lesser for walking and going downstairs, respectively. Finally, the validation of the present study with the results of Joshi et al. (ASME Open J Eng 1:011001, 2022) reveals that heterogeneous material model could be preferred over homogeneous material for hip prosthesis. © The Institution of Engineers (India) 2023.
引用
收藏
页码:1 / 20
页数:19
相关论文
共 50 条
  • [41] Patient-specific three-dimensional evaluation of interface micromotion in two different short stem designs in cementless total hip arthroplasty: a finite element analysis
    Kanaizumi, Arata
    Suzuki, Daisuke
    Nagoya, Satoshi
    Teramoto, Atsushi
    Yamashita, Toshihiko
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2022, 17 (01)
  • [42] Finite element analysis of dynamic stability of skeletal structures under periodic loading
    THANA Hemantha Kumar
    AMEEN Mohammed
    Journal of Zhejiang University(Science A:An International Applied Physics & Engineering Journal), 2007, (02) : 245 - 256
  • [43] Finite Element Analysis of Bone Stress for Miniscrew Implant Proximal to Root Under Occlusal Force and Implant Loading
    Shan, Li-Hua
    Guo, Na
    Zhou, Guan-jun
    Qie, Hui
    Li, Chen-Xi
    Lu, Lin
    JOURNAL OF CRANIOFACIAL SURGERY, 2015, 26 (07) : 2072 - 2076
  • [44] Finite Element Modeling and Analysis of Perforated Steel Members under Blast Loading
    Nawar, Mahmoud T.
    El-Zohairy, Ayman
    Arafa, Ibrahim T.
    MODELLING, 2023, 4 (04): : 628 - 649
  • [45] Determining the changes in morphology and loading status following medial displacement calcaneal osteotomy for flatfoot using patient-specific finite element models
    Kobayashi, Yumiko
    Ikoma, Kazuya
    Maki, Masahiro
    Imai, Kan
    Kido, Masamitsu
    Okubo, Naoki
    Sotozono, Yasutaka
    Wang, Zhongkui
    Hirai, Shinichi
    Tanaka, Masaki
    Takahashi, Kenji
    SCIENTIFIC REPORTS, 2024, 14 (01):
  • [46] Assessment of Intramural Segment Compression in Anomalous Coronary Arteries through Patient-Specific Finite Element Modeling
    Rosato, Antonio
    Lo Rito, Mauro
    Anglese, Serena
    Ceserani, Valentina
    Pascaner, Ariel Fernando
    Secchi, Francesco
    Conti, Michele
    APPLIED SCIENCES-BASEL, 2023, 13 (20):
  • [47] Influence of implant abutment type on stress distribution in bone under various loading conditions using finite element analysis
    Chun, HJ
    Shin, HS
    Han, CH
    Lee, SH
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2006, 21 (02) : 195 - 202
  • [48] Predisposing Factors for Orthodontic Mini-Implant Failure Defined by Bone Strains in Patient-Specific Finite Element Models
    Mhd Hassan Albogha
    Toru Kitahara
    Mitsugu Todo
    Hiroto Hyakutake
    Ichiro Takahashi
    Annals of Biomedical Engineering, 2016, 44 : 2948 - 2956
  • [49] Finite Element Modeling Using Patient-Specific Geometry to Predict Aortic Valve Insufficiency During Percutaneous Pulmonary Valve Implantation
    Donahue, Carly L.
    Aggarwal, Varun
    Barocas, Victor H.
    PROCEEDINGS OF THE 2022 DESIGN OF MEDICAL DEVICES CONFERENCE (DMD2022), 2022,
  • [50] Predisposing Factors for Orthodontic Mini-Implant Failure Defined by Bone Strains in Patient-Specific Finite Element Models
    Albogha, Mhd Hassan
    Kitahara, Toru
    Todo, Mitsugu
    Hyakutake, Hiroto
    Takahashi, Ichiro
    ANNALS OF BIOMEDICAL ENGINEERING, 2016, 44 (10) : 2948 - 2956