Homogeneous and Heterogeneous Modeling of Patient-Specific Hip Implant Under Static and Dynamic Loading Condition Using Finite Element Analysis

被引:5
|
作者
Ravikant [1 ]
Mittal V.K. [1 ]
Gupta V. [2 ]
机构
[1] Department of Mechanical Engineering, National Institute of Technology, Kurukshetra
[2] Department of Mechanical Engineering, Ch. Devi Lal State Institute of Engineering & Technology, Panniwala Mota
基金
英国科研创新办公室;
关键词
Finite element analysis; Gait motion; Hip prosthesis; Implant; Stress shielding;
D O I
10.1007/s40033-023-00447-0
中图分类号
学科分类号
摘要
Stress and strain shielding are the most common phenomenon shown by the metallic implant which reduces the implant life due to modulus mismatch. It is necessary to have a sufficient knowledge about the induced stress and strain distribution in order to find the suitable implant design with appropriate material combinations. In the present work, a three-dimensional implant model has been developed using real-life patient-specific computed tomographic (CT) data. Two different material models, viz. homogeneous (Ti–6Al–4V) and heterogeneous (Ti–6Al–4V and Ti-35Nb–5Ta-7Zr–0.4O), have been taken into account to demonstrate their mechanical properties like stress, strain and deformation. In addition, the boundary condition has been applied at distal end using ASTM F2996-1 and two different types of loading conditions have been applied (1) static loading using ISO 7206–4:2010(E) standard and (2) dynamic loading using patient-specific constraints, i.e., forces and torsional moment. For dynamic analysis, a patient-specific walking and going downstairs gait motions have been considered to obtain the essential mechanical characteristics. The final outcome reveals that heterogeneous model demonstrates lower von mises stress and approximate similar strain value compared to homogeneous model. It has been observed that the heterogeneous material model demonstrates less von mises stress at peak curve, i.e., 4.7% and 18.81% lesser for walking and going downstairs, respectively. Finally, the validation of the present study with the results of Joshi et al. (ASME Open J Eng 1:011001, 2022) reveals that heterogeneous material model could be preferred over homogeneous material for hip prosthesis. © The Institution of Engineers (India) 2023.
引用
收藏
页码:1 / 20
页数:19
相关论文
共 50 条
  • [1] Finite element modeling of hip implant static loading
    Colic, Katarina
    Sedmak, Aleksandar
    Grbovic, Aleksandar
    Tatic, Uros
    Sedmak, Simon
    Djordjevic, Branislav
    INTERNATIONAL CONFERENCE ON MANUFACTURING ENGINEERING AND MATERIALS, ICMEM 2016, 2016, 149 : 257 - 262
  • [2] Evaluation of Patient-Specific Cranial Implant Design Using Finite Element Analysis
    Huys, Stijn E. F.
    Van Gysel, Anke
    Mommaerts, Maurice Y.
    Vander Sloten, Jos
    WORLD NEUROSURGERY, 2021, 148 : 198 - 204
  • [3] Finite element analysis of hip implant with varying in taper neck lengths under static loading conditions
    Chethan, K. N.
    Bhat, N. Shyamasunder
    Zuber, Mohammad
    Shenoy, B. Satish
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2021, 208
  • [4] Finite Element Analysis of Femoral Implant Under Static Load
    Vulovic, Aleksandra
    Sustersic, Tijana
    Filipovic, Nenad
    2017 IEEE 17TH INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOENGINEERING (BIBE), 2017, : 559 - 562
  • [5] Finite element modeling of K-Monel bolts under static loading and dynamic shock loading
    Behan, Kevin
    Guzas, Emily
    Milburn, Jeffrey
    Moss, Stacy
    SHOCK AND VIBRATION, 2013, 20 (03) : 575 - 589
  • [6] Patient-specific finite element modeling for femoral bone augmentation
    Basafa, Ehsan
    Armiger, Robert S.
    Kutzer, Michael D.
    Belkoff, Stephen M.
    Mears, Simon C.
    Armand, Mehran
    MEDICAL ENGINEERING & PHYSICS, 2013, 35 (06) : 860 - 865
  • [7] Static, dynamic, and fatigue life investigation of a hip prosthesis for walking gait using finite element analysis
    Reginald, Jonathan
    Kalayarasan, M.
    Chethan, K. N.
    Dhanabal, P.
    INTERNATIONAL JOURNAL OF MODELLING AND SIMULATION, 2023, 43 (05) : 797 - 811
  • [8] Validation of patient-specific flatfoot models on finite element analysis
    Kobayashi, Yumiko
    Ikoma, Kazuya
    Maki, Masahiro
    Imai, Kan
    Kido, Masamitsu
    Okubo, Naoki
    Sotozono, Yasutaka
    Wang, Zhongkui
    Hirai, Shinichi
    Tanaka, Masaki
    Takahashi, Kenji
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2024,
  • [9] Finite Element Analysis of Different Artificial Hip Stem Designs Based on Fenestration under Static Loading
    Ikhsan
    Prabowo, Aditya Rio
    Sohn, Jung Min
    Triyono, Joko
    6TH INTERNATIONAL E-CONFERENCE ON INDUSTRIAL, MECHANICAL, ELECTRICAL AND CHEMICAL ENGINEERING (ICIMECE 2020) SPECIAL SYMPOSIUM - INTEGRITY OF MECHANICAL STRUCTURE AND MATERIAL, 2020, 27 : 101 - 108
  • [10] Biomechanical Analysis of Patient-Specific Temporomandibular Joint Implant and Comparison with Natural Intact Jaw Bone Using Finite Element Method
    Tiwari, Anubhav
    Gupta, Vijay Kumar
    Haldkar, Rakesh Kumar
    Parinov, Ivan A.
    APPLIED SCIENCES-BASEL, 2022, 12 (06):