Organ dose in cardiac dual-energy computed tomography: a Monte Carlo study

被引:0
作者
Delaram Pakravan
Farshid Babapour Mofrad
Mohammad Reza Deevband
Mahdi Ghorbani
Hamidreza Pouraliakbar
机构
[1] Islamic Azad University,Department of Medical Radiation Engineering, Science and Research Branch
[2] Shahid Beheshti University of Medical Sciences,Biomedical Engineering and Medical Physics Department, School of Medicine
[3] Iran University of Medical Sciences,Rajaie Cardiovascular Medical and Research Center
来源
Physical and Engineering Sciences in Medicine | 2022年 / 45卷
关键词
Monte Carlo simulation; Dual-energy computed tomography; Cardiac; Organ dose; Dose index;
D O I
暂无
中图分类号
学科分类号
摘要
Dual-energy computed tomography (DECT) has appeared as a novel approach with the aim of evaluating artery-related diseases. With the advent of DECT, concerns have been raised about the induction of diseases such as cancer due to high radiation exposure of patients. Therefore, the dose received by patients in DECT should be considered. The parameter most commonly used for patient dosimetry is the effective dose (ED). The purpose of this study is to model and validate a DECT scanner by a developed MCNP Monte Carlo code and to calculate the organ doses, the ED, and the conversion factor (k-factor) used in determining ED in the cardiac imaging protocol. To validate the DECT scanner simulation, a standard dosimetry body phantom was modeled in two radiation modes of single energy CT and DECT. The results of simulated CT dose index (CTDI) were compared with those of ImPACT or measurement data. Then dosimetry phantom was replaced by the male and female ORNL phantoms and the organ doses were calculated. The organ doses were also calculated by ImPACT dose software. In the initial validation stage, the minimum and maximum observed relative differences between results of MNCP simulation and measured were 2.77% and 5.79% for the central CTDI and 1.91% and 5.83% for the averaged peripheral CTDI, respectively. The mean ED of simulation and the ImPACT were 3.23 and 5.55 mSv/100 mAs, and the mean k-factor was 0.016 and 0.032 mSv mGy−1 cm−1 in the male and female phantoms, respectively. The k-factor obtained for males is close to the currently used k-factor, but the k-factor for females is almost twice.
引用
收藏
页码:157 / 166
页数:9
相关论文
共 96 条
[1]  
Cheng W(2010)Detection of myocardial perfusion abnormalities: standard dual-source coronary computed tomography angiography versus rest/stress technetium-99m single-photo emission CT Br J Radiol 83 652-660
[2]  
Zeng M(2008)The measurement, reporting, and management of radiation dose in CT Rep AAPM Task Group 23 1-28
[3]  
Arellano C(2015)Radiation risks: what is to be done? Am J Roentgenol 204 124-127
[4]  
Mafori W(2018)Effective dose estimation from organ dose measurements in fast-kV switch dual energy computed tomography Radiat Prot Dosim 182 352-358
[5]  
Goldin J(2012)Estimation and comparison of effective dose (E) in standard chest CT by organ dose measurements and dose-length-product methods and assessment of the influence of CT tube potential (energy dependency) on effective dose in a dual-source CT Eur J Radiol 81 e507-e512
[6]  
Krishnam M(2010)Dual energy CT of the chest: how about the dose? Invest Radiol 45 347-353
[7]  
Ruehm S(2012)Dual-energy CT: radiation dose aspects Am J Roentgenol 199 S16-S25
[8]  
McCollough C(2009)The development, validation and application of a multi-detector CT (MDCT) scanner model for assessing organ doses to the pregnant patient and the fetus using Monte Carlo simulations Phys Med Biol 54 2699-483
[9]  
Cody D(2011)Validation of radiation dose from a four slice CT scanner using Monte Carlo Arch Appl Sci Res 3 472-1663
[10]  
Edyvean S(2005)A Monte Carlo based method to estimate radiation dose from multidetector CT (MDCT): cylindrical and anthropomorphic phantoms Phys Med Biol 50 3989-20