Parallel imaging: Is GRAPPA a useful acquisition tool for MR imaging intended for volumetric brain analysis?

被引:11
作者
Lindholm T.L. [1 ,2 ]
Botes L. [1 ,2 ]
Engman E.-L. [2 ]
Frank A. [1 ,2 ]
Jonsson T. [1 ,2 ]
Svensson L. [1 ,2 ]
Julin P. [2 ,3 ,4 ]
机构
[1] Department of Diagnostic Medical Physics, Karolinska University Hospital Huddinge, Stockholm
[2] SMILE Image Laboratory, Karolinska Institute, Stockholm
[3] Astra Zeneca Research and Development, Södertälje
[4] Department of Rehabilitation Medicine, Danderyd Hospital, Karolinska Institute, Stockholm
关键词
Cerebral Spinal Fluid; Parallel Imaging; Acceleration Factor; Mild Cognitively Impaired Patient; Young Healthy Volunteer;
D O I
10.1186/1471-2342-9-15
中图分类号
学科分类号
摘要
Background: The work presented here investigates parallel imaging applied to T1-weighted high resolution imaging for use in longitudinal volumetric clinical studies involving Alzheimer's disease (AD) and Mild Cognitive Impairment (MCI) patients. This was in an effort to shorten acquisition times to minimise the risk of motion artefacts caused by patient discomfort and disorientation. The principle question is, "Can parallel imaging be used to acquire images at 1.5 T of sufficient quality to allow volumetric analysis of patient brains?" Methods: Optimisation studies were performed on a young healthy volunteer and the selected protocol (including the use of two different parallel imaging acceleration factors) was then tested on a cohort of 15 elderly volunteers including MCI and AD patients. In addition to automatic brain segmentation, hippocampus volumes were manually outlined and measured in all patients. The 15 patients were scanned on a second occasion approximately one week later using the same protocol and evaluated in the same manner to test repeatability of measurement using images acquired with the GRAPPA parallel imaging technique applied to the MPRAGE sequence. Results: Intraclass correlation tests show that almost perfect agreement between repeated measurements of both segmented brain parenchyma fraction and regional measurement of hippocampi. The protocol is suitable for both global and regional volumetric measurement dementia patients. Conclusion: In summary, these results indicate that parallel imaging can be used without detrimental effect to brain tissue segmentation and volumetric measurement and should be considered for both clinical and research studies where longitudinal measurements of brain tissue volumes are of interest. © 2009 Lindholm et al; licensee BioMed Central Ltd.
引用
收藏
相关论文
共 50 条
  • [21] A parallel MR imaging method using multilayer perceptron
    Kwon, Kinam
    Kim, Dongchan
    Park, HyunWook
    MEDICAL PHYSICS, 2017, 44 (12) : 6209 - 6224
  • [22] Evaluation of motion effects on parallel MR imaging with precalibration*
    Guo, Jun-Yu
    Kholmovski, Eugene G.
    Zhang, Ling
    Parker, Dennis L.
    MAGNETIC RESONANCE IMAGING, 2007, 25 (08) : 1130 - 1137
  • [23] Routine MR examination of the knee using parallel imaging
    Niitsu, M
    Ikeda, K
    CLINICAL RADIOLOGY, 2003, 58 (10) : 801 - 807
  • [24] High-pass GRAPPA: An image support reduction technique for improved partially parallel imaging
    Huang, Feng
    Li, Yu
    Vijayakumar, Sathya
    Hertel, Sarah
    Duensing, George R.
    MAGNETIC RESONANCE IN MEDICINE, 2008, 59 (03) : 642 - 649
  • [25] Fast method for 1D non-Cartesian parallel imaging using GRAPPA
    Heidemann, Robin M.
    Griswold, Mark A.
    Seiberlich, Nicole
    Nittka, Mathias
    Kannengiesser, Stephan A. R.
    Kiefer, Berthold
    Jakob, Peter M.
    MAGNETIC RESONANCE IN MEDICINE, 2007, 57 (06) : 1037 - 1046
  • [26] MR breast imaging : a comparative analysis of conventional and parallel imaging acquisitionRM delle mammelle: confronto tra tecnica convenzionale ed imaging parallelo
    A. Orlacchio
    F. Bolacchi
    A. Rotili
    E. Cossu
    I. Tanga
    V. Cozzolino
    G. Simonetti
    La radiologia medica, 2008, 113 : 465 - 476
  • [27] MR imaging of the cervical spine: assessment of image quality with parallel imaging compared to non-accelerated MR measurements
    Noebauer-Huhmann, Iris-M.
    Glaser, Christian
    Dietrich, Olaf
    Wallner, Claus-Peter
    Klinger, Wolfgang
    Imhof, Herwig
    Schoenberg, Stefan O.
    EUROPEAN RADIOLOGY, 2007, 17 (05) : 1147 - 1155
  • [28] MR imaging of the cervical spine: assessment of image quality with parallel imaging compared to non-accelerated MR measurements
    Iris-M. Noebauer-Huhmann
    Christian Glaser
    Olaf Dietrich
    Claus-Peter Wallner
    Wolfgang Klinger
    Herwig Imhof
    Stefan O. Schoenberg
    European Radiology, 2007, 17 : 1147 - 1155
  • [29] Using GRAPPA to improve autocalibrated coil sensitivity estimation for the SENSE family of parallel imaging reconstruction algorithms
    Hoge, W. Scott
    Brooks, Dana H.
    MAGNETIC RESONANCE IN MEDICINE, 2008, 60 (02) : 462 - 467
  • [30] MR Physics in Practice How to Optimize Acquisition Quality and Time for Cardiac MR Imaging
    Saloner, David
    Liu, Jing
    Haraldsson, Henrik
    MAGNETIC RESONANCE IMAGING CLINICS OF NORTH AMERICA, 2015, 23 (01) : 1 - +