Validity of the 3D VECTRA photogrammetric surface imaging system for cranio-maxillofacial anthropometric measurements

被引:82
作者
Metzler P. [1 ]
Sun Y. [1 ]
Zemann W. [1 ]
Bartella A. [1 ]
Lehner M. [1 ]
Obwegeser J.A. [1 ]
Kruse-Gujer A.L. [1 ]
Lübbers H.-T. [1 ]
机构
[1] Department of Cranio-Maxillofacial and Oral Surgery, University of Zurich, Frauenklinikstrasse 24, Zurich
关键词
3D; Three-dimensional; Photography; Photometry; Anthropometric; Accuracy; Validation;
D O I
10.1007/s10006-013-0404-7
中图分类号
学科分类号
摘要
PURPOSE: The use of three-dimensional (3D) photography for anthropometric measurements is of increasing interest, especially in the cranio-maxillofacial field. Before standard implementation, accurate determination of the precision and accuracy of each system is mandatory.; METHODS: A mannequin head was labelled with 52 landmarks, and 28 three-dimensional images were taken using a commercially available five-pod 3D photosystem (3D VECTRA; Canfield, Fairfield, NJ) in different head positions. Distances between the landmarks were measured manually using a conventional calliper and compared with the digitally calculated distances acquired from labelling by two independent observers. The experimental set-up accounted for clinical circumstances by varying the positioning (vertical, horizontal, sagittal) of the phantom.; RESULTS: In the entire calliper measurement data set (n = 410), a significant difference (p = 0.02) between the directly measured and corresponding virtually calculated distances was found. The mean aberration between both modalities covering all data was 7.96 mm. No differences (p = 0.94) between the two groups were found using a cut-off of 10 % (leaving n = 369 distances) due to considerable errors in direct measurements and the necessary manual data translation. The mean diversity of both measurement modalities after cut-off was 1.33 mm (maximum, 6.70 mm). Inter-observer analysis of all 1,326 distances showed no difference (p = 0.99; maximal difference, 0.58 mm) in the digital measurements.; CONCLUSION: The precision and accuracy of this five-pod 3D photosystem suggests its suitability for clinical applications, particularly anthropometric studies. Three-hundred-and-sixty degree surface-contour mapping of the craniofacial region within milliseconds is particularly useful in paediatric patients. Proper patient positioning is essential for high-quality imaging.
引用
收藏
页码:297 / 304
页数:7
相关论文
共 118 条
[1]  
Meintjes EM(2002)A stereo-photogrammetric method to measure the facial dysmorphology of children in the diagnosis of fetal alcohol syndrome Med Eng Phys 24 683-689
[2]  
Douglas TS(1977)Anthropometry of the face in lateral facial dysplasia: the unilateral form Cleft Palate J 14 193-199
[3]  
Martinez F(1985)Craniofacial disproportions in Apert's syndrome: an anthropometric study Cleft Palate J 22 253-265
[4]  
Vaughan CL(1985)Surface morphology in Treacher Collins syndrome: an anthropometric study Cleft Palate J 22 266-274
[5]  
Adams LP(1993)Methods in anthropometric studies Cleft Palate Craniofac J 30 429-431
[6]  
Stekhoven A(2010)Preoperative craniofacial dysmorphology in isolated sagittal synostosis: a comprehensive anthropometric evaluation J Craniofac Surg 21 1404-1410
[7]  
Farkas LG(2011)An epidemiological study of nonsyndromal craniosynostoses J Craniofac Surg 22 47-49
[8]  
James JS(2013)Postoperative cranial vault growth in premature sagittal craniosynostosis J Craniofac Surg 24 146-149
[9]  
Farkas LG(2009)Single sutural craniosynostoses: surgical outcomes and long-term growth Plast Reconstr Surg 123 635-642
[10]  
Kolar JC(1985)Vertical and horizontal proportions of the face in young adult North American Caucasians: revision of neoclassical canons Plast Reconstr Surg 75 328-338