Characterization of aqueous rose bengal dye solution for the measurement of low doses of gamma radiation

被引:0
|
作者
Hasan Mahmood Khan
Ashfaq A. Khan
机构
[1] University of Peshawar,Radiation Chemistry Laboratory, National Centre of Excellence in Physical Chemistry
来源
Journal of Radioanalytical and Nuclear Chemistry | 2010年 / 284卷
关键词
Dosimetry; Rose bengal aqueous solution; Low dose radiation; Absorbance; Gamma radiation;
D O I
暂无
中图分类号
学科分类号
摘要
Aqueous solution of rose bengal dye has been studied spectrophotometrically as a gamma-ray dosimeter for the measurement of low doses of radiation. The useful dose range was found to be from 50 to 1000 Gy when the measurements were made at 549 nm. The effects of temperature and light conditions on the stability of response during post-irradiation storage were also investigated. When stored in dark at room temperature, the dosimetric solutions showed a stable response up to 22 days. The storage of irradiated solutions in diffused sunlight showed a stable response only up to 6 days. When exposed to direct sunlight, very prominent and fast bleaching of dye solution occurred. At low storage temperature (ca. 11 °C), dosimetric response was found to be stable up to 22 days while at higher temperature (ca. 30 °C), the response of dosimetric solution was stable only up to 6 days. The rose bengal aqueous solution showed promising characteristics as a low dose radiation dosimeter when stored at lower temperatures (<25 °C) in dark.
引用
收藏
页码:37 / 42
页数:5
相关论文
共 38 条
  • [1] Characterization of aqueous rose bengal dye solution for the measurement of low doses of gamma radiation
    Khan, Hasan Mahmood
    Khan, Ashfaq A.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2010, 284 (01) : 37 - 42
  • [2] A New Fluorometric Dosimetry for Low-medium Gamma Radiation Doses
    Ece Ergun
    Journal of Fluorescence, 2021, 31 : 941 - 950
  • [3] A New Fluorometric Dosimetry for Low-medium Gamma Radiation Doses
    Ergun, Ece
    JOURNAL OF FLUORESCENCE, 2021, 31 (04) : 941 - 950
  • [4] Degradation of antineoplastic cytarabine in aqueous solution by gamma radiation
    Ocampo-Perez, R.
    Rivera-Utrilla, J.
    Sanchez-Polo, M.
    Lopez-Penalver, J. J.
    Leyva-Ramos, R.
    CHEMICAL ENGINEERING JOURNAL, 2011, 174 (01) : 1 - 8
  • [5] The impact of low doses of gamma radiation on Echis coloratus venom and its fractions
    Samy, Esraa M.
    Shaaban, Esmat A.
    Kenawy, Sanaa A.
    Abd Elfattah, Mai A.
    Salama, Walaa H.
    RADIATION PHYSICS AND CHEMISTRY, 2018, 150 : 145 - 150
  • [6] Optimisation of radiolysis of Reactive Red 120 dye in aqueous solution using ionising 60Co gamma radiation by response surface methodology
    Padmanaban, V. C.
    Nandagopal, M. S. Giri
    Achary, Anant
    Vasudevan, V. N.
    Selvaraju, N.
    WATER SCIENCE AND TECHNOLOGY, 2016, 73 (12) : 3041 - 3048
  • [7] Xenon-filled diode performance under influence of low doses of gamma radiation
    Pejovic, Milic M.
    Zivanovic, Emilija N.
    Stojanovic, Milan D.
    APPLIED RADIATION AND ISOTOPES, 2022, 184
  • [8] Assessment of reaction intermediates of gamma radiation-induced degradation of ofloxacin in aqueous solution
    Changotra, Rahil
    Guin, Jhimli Paul
    Varshney, Lalit
    Dhir, Amit
    CHEMOSPHERE, 2018, 208 : 606 - 613
  • [9] The effect of high-energy radiation on aqueous solution of Acid Red 1 textile dye
    Foeldvary, Cs. M.
    Wojnarovits, L.
    RADIATION PHYSICS AND CHEMISTRY, 2007, 76 (8-9) : 1485 - 1488
  • [10] Fecundity changes induced by low-doses of gamma radiation on Biomphalaria straminea (Dunker, 1848)
    daMotta, MA