Optimal Material Selection for Total Hip Implant: A Finite Element Case Study

被引:0
作者
Abdullah Tahir Şensoy
Murat Çolak
Irfan Kaymaz
Fehim Findik
机构
[1] Bayburt University,Technical Sciences Vocational School
[2] Bayburt University,Mechanical Engineering Department
[3] Erzurum Technical University,Engineering Faculty
[4] Sakarya Applied Sciences University,Technology Faculty
[5] Sakarya University,BIMAS
来源
Arabian Journal for Science and Engineering | 2019年 / 44卷
关键词
Hip prosthesis; Optimal material; Selection; CES; Finite element analysis; Implant;
D O I
暂无
中图分类号
学科分类号
摘要
The selection of most proper materials in engineering design is known as an important stage of the design process. In order to successfully complete this stage, it is necessary to have sufficient knowledge about the structure of materials, density, melting point, thermal expansion coefficient, tensile and yield strength, elongation, modulus of elasticity, hardness and many other properties. There are several selection systems that help the design engineer to choose most suitable material that meet the required properties. In the field of bioengineering, the selection of materials and the development of new materials for the clinical needs are increasingly important. In this study, the cases of optimal implant stabilization were investigated, material alternatives for hip prosthesis were evaluated, and optimal materials were determined. Using computerized tomography data with MIMICS software, virtual surgery was applied the hip bone and the implant was attached to bone. Boundary conditions and material properties have been defined, and finite element model has been created. FEA investigation of the mechanical behavior of the hip implant for various material alternatives determined by the CES software showed that the best material candidate is austenitic, annealed and biodurable stainless steel in terms of the micromotions at the implant–bone cement interface regarding osseointegration. This candidate showed 20.69% less strain value than the most commercially used hip implant material, Ti6Al4V. Therefore, the findings of this study suggest that the use of some specific stainless steel materials for implants may reduce the operation cost and increase the operation success for the total hip arthroplasty.
引用
收藏
页码:10293 / 10301
页数:8
相关论文
共 50 条
[31]   Finite element analysis of interfacial crack behaviour in cemented total hip arthroplasty [J].
Achour, T. ;
Tabeti, M. S. H. ;
Bouziane, M. M. ;
Benbarek, S. ;
Bouiadjra, B. Bachir ;
Mankour, A. .
COMPUTATIONAL MATERIALS SCIENCE, 2010, 47 (03) :672-677
[32]   Effects of cementless fixation of implant prosthesis: A finite element study [J].
Lee, Hyeonjong ;
Park, Soyeon ;
Kwon, Kung-Rock ;
Noh, Gunwoo .
JOURNAL OF ADVANCED PROSTHODONTICS, 2019, 11 (06) :341-349
[33]   Effect of Implant Design on Stress Distribution: A Finite Element Study [J].
Houshmand B. ;
Ardakani M.T. ;
Moscowchi A. ;
Oskoui I.Z. .
Journal of Long-Term Effects of Medical Implants, 2022, 32 (04) :39-45
[34]   Bone remodeling after total hip arthroplasty with a short stemmed metaphyseal loading implant: Finite element analysis validated by a prospective DEXA investigation [J].
Lerch, Matthias ;
Kurtz, Agnes ;
Stukenborg-Colsman, Christina ;
Nolte, Ingo ;
Weigel, Nelly ;
Bouguecha, Anas ;
Behrens, Bernd A. .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2012, 30 (11) :1822-1829
[35]   A Review of Biomaterials and Associated Performance Metrics Analysis in Pre-Clinical Finite Element Model and in Implementation Stages for Total Hip Implant System [J].
Soliman, Md Mohiuddin ;
Chowdhury, Muhammad E. H. ;
Islam, Mohammad Tariqul ;
Musharavati, Farayi ;
Nabil, Mohammad ;
Hafizh, Muhammad ;
Khandakar, Amith ;
Mahmud, Sakib ;
Nezhad, Erfan Zal ;
Shuzan, Md Nazmul Islam ;
Abir, Farhan Fuad .
POLYMERS, 2022, 14 (20)
[36]   Effect of Short Dental Implant Material on Bone Stress: An In Vitro Finite Element Analysis [J].
Al Qahtani, Waleed M. S. .
JOURNAL OF BIOMATERIALS AND TISSUE ENGINEERING, 2022, 12 (09) :1781-1786
[37]   EFFECTS OF IMPLANT MATERIAL PROPERTIES AND IMPLANT-BONE BONDING ON STRESS SHIELDING IN CEMENTLESS TOTAL HIP-ARTHROPLASTY [J].
MCNAMARA, BP ;
TONI, A ;
TAYOR, D .
ADVANCES IN ENGINEERING MATERIALS, 1995, 99-1 :309-314
[38]   Glenoid implant orientation and cement failure in total shoulder arthroplasty: a finite element analysis [J].
Yongpravat, Charlie ;
Kim, H. Mike ;
Gardner, Thomas R. ;
Bigliani, Louis U. ;
Levine, William N. ;
Ahmad, Christopher S. .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2013, 22 (07) :940-947
[39]   Influence of sidewall retention and interference fit in total ankle replacement on implant-bone micromotion: A finite element study [J].
Johnson, Joshua E. ;
Clarke, Gabriel A. ;
de Cesar Netto, Cesar ;
Anderson, Donald D. .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2024, 42 (07) :1536-1544
[40]   The effect of abutment material stiffness on the mechanical behavior of dental implant assemblies: A 3D finite element study [J].
Tonin, Bruna S. H. ;
Fu, Jing ;
He, Yiting ;
Ye, Ning ;
Chew, Hooi Pin ;
Fok, Alex .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2023, 142