Photon beam softening coefficient determination with slab thickness in small filed size: Monte Carlo study

被引:9
|
作者
Bencheikh M. [1 ]
Maghnouj A. [1 ]
Tajmouati J. [1 ]
机构
[1] LISTA Laboratory, Physics Department, Faculty of Sciences Dhar El-Mahraz, University of Sidi Mohamed Ben Abdellah, Fez
关键词
BEAMDP code; BEAMnrc code; Monte Carlo simulation; Photon beam softening; Slab thickness;
D O I
10.1134/S1547477117060085
中图分类号
学科分类号
摘要
Determination and understanding of photon beam softening using material soften photon beam for clinical usage is important for material study for attenuation and for beam modifier enhancements and linac improvements. Monte Carlo model was used to simulate 6 MeV photon beams produced by Varian Clinac 2100 accelerator with flattening filter thereafter the flattening filter was replaced by a slab of aluminum and copper with different of 0.5, 1, 1.5 and 2 mm. The Monte Carlo geometry was validated by a gamma index acceptance rate of 99% in PDD and 98% in dose profiles, the gamma criteria was 3% for dose difference and 3mm for distance to agreement. The purpose was to investigate the beam softening for small size and beam attenuation as a function of inserted slab thickness of copper and aluminum and also as a function of off-axis distance. For beam softening evaluation, variation amplitude of beam softening coefficient a1 was very high near the beam central axis and decreased with off-axis distance and also it was high for aluminum slab compared to copper slab. For aluminum slab, variation amplitude of beam softening coefficient a1 have a minimum at–0.5 cm–1 and a maximum at 0.5 cm–1 and for copper slab, variation amplitude of beam softening a1 have a minimum at–0.15 cm–1 and a maximum at 0.11 cm–1. Variation amplitude of beam softening coefficient a2 was very high near the beam central axis and decreased with off-axis distance and it was very high for aluminum slab compared to copper slab. For aluminum slab, variation amplitude of beam softening coefficient a2 have a minimum at–0.54 cm–2 and a maximum at 0.44 cm–2 and for copper slab, variation amplitude of beam softening coefficient a2 have a minimum at–0.111 cm–2 and a maximum at 0.0825 cm–2. © 2017, Pleiades Publishing, Ltd.
引用
收藏
页码:963 / 970
页数:7
相关论文
共 28 条
  • [21] Effects of wavelength, beam type and size on cerebral low-level laser therapy by a Monte Carlo study on visible Chinese human
    Li, Ting
    Zhao, Yue
    Sun, Yunlong
    Li, Kai
    JOURNAL OF INNOVATIVE OPTICAL HEALTH SCIENCES, 2015, 8 (01)
  • [22] Determination of Nanocrystal Size Distribution in Magnetic Multicore Particles Including Dipole-Dipole Interactions and Magnetic Anisotropy: a Monte Carlo Study
    Schaller, Vincent
    Wahnstrom, Goran
    Sanz-Velasco, Anke
    Enoksson, Peter
    Johansson, Christer
    8TH INTERNATIONAL CONFERENCE ON THE SCIENTIFIC AND CLINICAL APPLICATIONS OF MAGNETIC CARRIERS, 2010, 1311 : 42 - +
  • [23] Influence of the size and skin thickness of apple varieties on the retrieval of internal optical properties using Vis/NIR spectroscopy: A Monte Carlo-based study
    Vaudelle, Fabrice
    L'Huillier, Jean-Pierre
    COMPUTERS AND ELECTRONICS IN AGRICULTURE, 2015, 116 : 137 - 149
  • [24] Comparison of Flattening Filter (FF) and Flattening-Filter-Free (FFF) 6 MV photon beam characteristics for small field dosimetry using EGSnrc Monte Carlo code
    Sangeetha, S.
    Sureka, C. S.
    RADIATION PHYSICS AND CHEMISTRY, 2017, 135 : 63 - 75
  • [25] Assessment of skin doses in small field radiotherapy for 6 MV photons and beam spectral analysis at skin surface: an EGSnrc based Monte Carlo study
    Attia Gul
    M. Basim Kakakhel
    Sikander M. Mirza
    Radiation and Environmental Biophysics, 2021, 60 : 299 - 308
  • [26] Assessment of skin doses in small field radiotherapy for 6 MV photons and beam spectral analysis at skin surface: an EGSnrc based Monte Carlo study
    Gul, Attia
    Kakakhel, M. Basim
    Mirza, Sikander M.
    RADIATION AND ENVIRONMENTAL BIOPHYSICS, 2021, 60 (02) : 299 - 308
  • [27] Evaluation of dosimetric parameters of small fields of 6 MV flattening filter free photon beam measured using various detectors against Monte Carlo simulation (Mar, 10.1017/S1460396920000114, 2020)
    Mamballikalam, Gopinath
    Senthilkumar, S.
    Jayadevan, P. M.
    Jaon Bos, R. C.
    Basith, P. M. Ahamed
    Inippully, Rohit
    Shine, N. S.
    Clinto, C. O.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2021, 20 (03) : 369 - 369
  • [28] A phantom-based experimental and Monte Carlo study of the suitability of in-vivo diodes and TLD for entrance in-vivo dosimetry in small-to-medium sized 6 MV photon fields
    Ketabi, Ali
    Karbasi, Sareh
    Faghihi, Reza
    Mosleh-Shirazi, Mohammad Amin
    RADIATION PHYSICS AND CHEMISTRY, 2022, 201