Sub-4-micron full-field optical coherence tomography on a budget

被引:0
作者
HARI NANDAKUMAR
ADITHYA KOKKODU SUBRAMANIA
SHAILESH SRIVASTAVA
机构
[1] Sri Sathya Sai Institute of Higher Learning,Department of Physics
[2] Prasanthi Nilayam,undefined
来源
Sādhanā | 2018年 / 43卷
关键词
Optical coherence tomography; coherence imaging; interferometric imaging;
D O I
暂无
中图分类号
学科分类号
摘要
We evaluate the construction and performance of an ultra-low-cost full-field optical coherence tomography (FF-OCT) instrument, also known as an optical coherence microscope. Although the cost of construction of the instrument from off-the-shelf parts is at least ten times lower than those of commercial products, sub-4-micron axial and lateral resolutions are obtained, albeit at the cost of higher acquisition times. Standard test samples are imaged and the performance of the instrument is evaluated. The device is found to be useful in measuring length, dispersion and group refractive index as well. Suggestions for bettering performance are discussed.
引用
收藏
相关论文
共 75 条
  • [1] Swanson EA(1992)High-speed optical coherence domain reflectometry Opt. Lett. 17 151-153
  • [2] Huang D(1998)Eye-length measurement by interferometry with partially coherent light Opt. Lett. 13 186-188
  • [3] Lin CP(1992)Measurement of corneal thickness by laser Doppler interferometry Invest. Ophthalmol. Vis. Sci. 33 98-103
  • [4] Puliafito CA(2015)Multiple-reference optical coherence tomography for smartphone applications SPIE Newsroom 37 4820-4822
  • [5] Hee MR(2012)Integrated-optics-based swept-source optical coherence tomography Opt. Lett. 23 244-246
  • [6] Fujimoto JG(1998)Full-field optical coherence microscopy Opt. Lett. 41 805-812
  • [7] Fercher AF(2002)High-resolution full-field optical coherence tomography with a Linnik microscope Appl. Opt. 7 236-48
  • [8] Mengedoht K(2017)Full-field optical coherence tomography as a diagnosis tool: recent progress with multimodal imaging Appl. Sci. 254 1178-342
  • [9] Werner W(1991)Optical coherence tomography Science 117 43-1708
  • [10] Hitzenberger CK(1995)Measurement of intraocular distances by backscattering spectral interferometry Opt. Commun. 22 340-6295