Development and validation of a high-speed video system for measuring saccadic eye movement

被引:0
作者
Jeffrey S. Brooks
William J. Smith
Brandon M. Webb
Matthew D. Heath
James P. Dickey
机构
[1] Western University,School of Kinesiology, Faculty of Health Sciences
来源
Behavior Research Methods | 2019年 / 51卷
关键词
Antisaccades; Prosaccades; Subconcussive impacts; Oculomotor; Brain; Eye tracker; Reaction time;
D O I
暂无
中图分类号
学科分类号
摘要
Laboratory-based retroreflective and magnetic scleral search-coil technologies are the current standards for collecting saccadometric data, but such equipment is costly and cumbersome. We have validated a novel, portable, high-speed video camera-based system (Exilim EX-FH20, Casio, Tokyo, Japan) for measuring saccade reaction time (RT) and error rate in a well-lit environment. This system would enable measurements of pro- and antisaccades in athletes, which is important because antisaccade metrics provide a valid tool for concussion diagnosis and determining an athlete’s safe return to play. A total of 529 trials collected from 15 participants were used to compare saccade RT and error rate measurements of the high-speed camera system to a retroreflective video-based eye tracker (Eye-Trac 6: Applied Sciences Laboratories, Bedford, MA). Bland–Altman analysis revealed that the RT measurements made by the high-speed video system were 11 ms slower than those made by the retroreflective system. Error rate measurements were identical between the two systems. An excellent degree of reliability was found between the system measurements and in the ratings of independent researchers examining the video data. A strong association (r = .97) between the RTs determined via the retroreflective and high-speed camera systems was observed across all trials. Our high-speed camera system is portable and easily set up, does not require extensive equipment calibration, and can be used in a well-lit environment. Accordingly, the camera-based capture of saccadometric data may provide a valuable tool for neurological assessment following a concussive event and for the continued monitoring of recovery.
引用
收藏
页码:2302 / 2309
页数:7
相关论文
共 200 条
[1]  
Allison RS(1996)Combined head and eye tracking system for dynamic testing of the vestibular system IEEE Transactions on Biomedical Engineering 43 1073-1082
[2]  
Eizenman M(2010)Sampling frequency and eye-tracking measures: How speed affects durations, latenciesand more Journal of Eye Movement Research 3 6-5
[3]  
Cheung BS(2013)An internationally standardised antisaccade protocol Vision Research 84 1-48
[4]  
Andersson R(2016)Oculomotor, vestibular, and reaction time tests in mild traumatic brain injury PLoS ONE 11 e0162168-310
[5]  
Nyström M(2015)Fitting linear mixed-effects models using lme4 Journal of Statistical Software 67 1-436
[6]  
Holmqvist K(1986)Statistical methods for assessing agreement between two methods of clinical measurement The Lancet 327 307-934
[7]  
Antoniades C(1997)The Psychophysics Toolbox Spatial Vision 10 433-28
[8]  
Ettinger U(2014)Vertical saccades in children: A developmental study Experimental Brain Research 232 927-343
[9]  
Gaymard B(2015)Differential eye movements in mild traumatic brain injury versus normal controls Journal of Head Trauma Rehabilitation 30 21-268
[10]  
Gilchrist I(1994)Frontal cortex: Choosing where to look Current Biology 4 341-IN5