Measurement of naturally occurring radioactive materials, 238U and 232Th-part 3: is efficiency calibration necessary for quantitative measurement of ultra-low level NORM?

被引:0
作者
Nabanita Naskar
Susanta Lahiri
Punarbasu Chaudhuri
Alok Srivastava
机构
[1] University of Calcutta,Department of Environmental Science
[2] Saha Institute of Nuclear Physics,Department of Chemistry
[3] Panjab University,undefined
来源
Journal of Radioanalytical and Nuclear Chemistry | 2017年 / 314卷
关键词
Naturally occurring radioactive materials; Gamma-spectrometry; Efficiency calibration; Comparator method;
D O I
暂无
中图分类号
学科分类号
摘要
A well-calibrated and highly shielded HPGe detector with considerable relative efficiency is an efficient tool for measurement of naturally occurring radioactive material (NORM) present at ultra low-level in the environmental samples. Efficiency calibration is standard protocol required for quantitative estimation of the activity of any radioactive sample. In the present work it has been shown that in the case of ultra-low level NORM measurement, one can comfortably by pass efficiency calibration by using comparator method i.e., comparing the peak area counts of different photopeaks of standard and samples provided they are counted in same geometry and comparable composition.
引用
收藏
页码:507 / 511
页数:4
相关论文
共 107 条
[1]  
Durec F(2008)Intercomparison exercise on the determination of radionuclides in sediment from the Dudvah River Appl Radiat Isot 66 1706-1710
[2]  
Betti M(2016)Measurement of naturally occurring radioactive material, J Radioanal Nucl Chem 310 1381-1396
[3]  
Durecova A(2017)U and J Radioanal Nucl Chem 312 161-171
[4]  
Naskar N(2008)Th: anomalies in photo-peak selection J Radioanal Nucl Chem 277 579-585
[5]  
Lahiri S(1996)Measurement of naturally occurring radioactive material, Nucl Instr Methods Phys Res Sect A 382 495-502
[6]  
Chaudhuri P(2001)U and Nucl Instr Methods Phys Res Sect A 470 520-532
[7]  
Srivastava A(2002)Th: part 2-optimization of counting time Nucl Instr Methods Phys Res Sect A 491 152-162
[8]  
Naskar N(2006)Practical considerations regarding high resolution gamma-spectrometry measurements of naturally occurring radioactive samples Appl Radiat Isot 64 1319-1322
[9]  
Lahiri S(2003)A method for the determination of counting efficiencies in γ-spectrometric measurements with HPGe detectors Nucl Instr Methods Phys Res Sect A 511 360-381
[10]  
Chaudhuri P(2004)Efficiency calibration of a HPGe detector in the [46.54–2000] keV energy range for the measurement of environmental samples Nucl Instr and Methods Phys Res Sect A 518 764-774