Evaluation of a transcutaneous Jaundice Meter following hospital discharge in term and near-term neonates

被引:53
作者
Engle W.D. [1 ]
Jackson G.L. [1 ]
Stehel E.K. [1 ]
Sendelbach D.M. [1 ]
Manning M.D. [1 ]
机构
[1] Department of Pediatrics, Division of Neonatal-Perinatal Medicine, University of Texas Southwestern Medical Center at Dallas, Dallas, TX 75390-9063
关键词
D O I
10.1038/sj.jp.7211333
中图分类号
学科分类号
摘要
Objectives: To evaluate performance of the Minolta JM-103 Jaundice Meter (JM) as a predictor of total serum bilirubin (TSB) in outpatient neonates during the first week postnatal, and to estimate the number of TSB determinations that might be avoided in clinical use. Study design: In neonates evaluated posthospital discharge. JM and TSB results were comopared using linear regression and a Bland-Altman plot, and predictive indices were calculated for various JM cutoff values. Utilizing the 2004 American Academy of Pediatrics (AAP) guidelines, the ability of JM to predict risk zone status was determined. Results: Overall correlation between JM and TSB was 0.77 (p<0.001; n = 121). When TSB was >17 mg/dl, a cutoff value for JM of 13 mg/dl had a sensitivity of 1.0, and 50% of TSB determinations would be avoided. Conclusions: JM may facilitate outpatient management of hyperbilirubinemia by reducing the number of TSB determinations required; however, it does not provide a reliable substitute for laboratory measurement of TSB. © 2005 Nature Publishing Group All rights reserved.
引用
收藏
页码:486 / 490
页数:4
相关论文
共 24 条
  • [1] Sentinel Event Alert: Kernicterus threatens healthy newborns, (2001)
  • [2] Kernicterus in full-term infants, United States, 1994-1998, MMWR Morbid. Mortal Wkly. Rep., 50, pp. 491-494, (2001)
  • [3] A national framework for healthcare quality measurement and reporting: A consensus report
  • [4] Management of hyperbilirubinemia in the newborn infant 35 or more weeks of gestation, Pediatrics, 114, pp. 297-316, (2004)
  • [5] Tayaba R., Gribetz D., Gribetz I., Holzman I.R., Noninvasive estimation of serum bilirubin, Pediatrics, 102, 3, (1998)
  • [6] Maisels M.J., Newman T.B., Jaundice in full-term and near-term babies who leave the hospital within 36 hours, Clin. Perinatol., 25, pp. 295-302, (1998)
  • [7] Robertson A., Kazmierczak S., Vos P., Improved transcutaneous bilirubinometry: Comparison of SpectRx BiliCheck and Minolta Jaundice Meter JM-102 for estimating total serum bilirubin in a normal newborn population, J. Perinatol., 22, pp. 12-14, (2002)
  • [8] Schumacher R.E., Noninvasive measurements of bilirubin in the newborn, Clin. Perinatol., 17, pp. 417-435, (1990)
  • [9] Rubaltelli F.F., Gourley G.R., Loskamp N., Et al., Transcutaneous bilirubin measurement: A multicenter evaluation of a new device, Pediatrics, 107, pp. 1264-1271, (2001)
  • [10] Kazmierczak S.C., Robertson A.F., Briley K.P., Kreamer B., Gourley G.R., Transcutaneous measurement of bilirubin in newborns: Comparison with an automated Jendrassik-Grof procedure and HPLC, Clin. Chem., 50, pp. 433-435, (2004)