Platinum determination by inductively coupled plasma–sector field mass spectrometry (ICP–SFMS) in different matrices relevant to human biomonitoring

被引:0
作者
S. Hann
G. Koellensperger
K. Kanitsar
G. Stingeder
M. Brunner
B. Erovic
M. Müller
C. Reiter
机构
[1] University of Agricultural Sciences,Institute of Chemistry, BOKU
[2] University of Vienna Medical School,Vienna
[3] University of Vienna,Department of Clinical Pharmacology, Division of Clinical Pharmacokinetics
来源
Analytical and Bioanalytical Chemistry | 2003年 / 376卷
关键词
Platinum; Inductively coupled plasma–sector field mass spectrometry; Urine; Serum; Clinical microdialysis; Lung tissue;
D O I
暂无
中图分类号
学科分类号
摘要
The analytical challenges of Pt determination by ICP–SFMS posed by different human tissues and fluids have been critically assessed. Investigated samples were (1) urine, (2) serum of cancer patients sampled during chemotherapy with carboplatin, (3) microdialysates (20 µL sample volume) collected from tumor and non-tumor tissue, and, finally—for the first time—(4) human lung tissue to study background concentrations of inhaled platinum. Sample preparation involved microwave digestion and open vessel treatment or simple dilution (microdialysates). Depending on the sample preparation and introduction systems used (microconcentric nebulization, ultrasonic nebulization with and without membrane desolvation) excellent procedural detection limits (3s criterion) of 0.35 pg g−1 for urine, 420 pg g−1 for serum, 400 pg g−1 for lung tissue and 13 pg g−1 for microdialysates could be obtained. Ultratrace concentrations of 1–40 pg g−1, and 1000–3000 pg g−1 were measured in urine and human lung tissue, respectively, as typical for samples in environmental studies. Quantification was carried out by IDMS and standard addition in the case of urine samples. Internal standardization could not correct for non-spectral interferences in external calibration. In the serum and microdialysates of patients during chemotherapy with carboplatin, elevated Pt levels ranging between 0.01 and 10 µg g−1 were determined by external calibration (195Pt isotope). For all investigated samples spectral interferences could be excluded by following different strategies. High-resolution control measurements (194Pt, 195Pt) were performed in the case of elevated Pt levels, i.e. for microdialysates and serum samples. An Hf/Pt ratio of 0.4 was determined in human lung samples. An HfO formation ratio of 0.2% was assessed for standard solutions at the present experimental conditions, revealing that the contribution of 179Hf16O, 178Hf17O, 177Hf18O to the 195Pt isotope signal used for quantification was not significant.
引用
收藏
页码:198 / 204
页数:6
相关论文
共 19 条
  • [1] Barefoot undefined(1999)undefined Talanta 49 1-undefined
  • [2] Rao undefined(2000)undefined Trends Anal Chem 19 565-undefined
  • [3] Haverland undefined(1991)undefined Int J Rad Appl Instr 42 775-undefined
  • [4] Parry undefined(1992)undefined J Geochem Explor 44 321-undefined
  • [5] Artem'ev undefined(1993)undefined J Radioanal Nucl Chem 173 125-undefined
  • [6] Hoppstock undefined(1989)undefined Fresenius J Anal Chem 335 813-undefined
  • [7] Helmers undefined(1998)undefined Fresenius J Anal Chem 362 522-undefined
  • [8] Krachler undefined(1998)undefined Anal Chim Acta 363 1-undefined
  • [9] Bergerov undefined(1997)undefined J Anal At Spectrom 12 1095-undefined
  • [10] Köllensperger undefined(2000)undefined J Anal Atom Spectrom 15 1553-undefined