Quantitation of specific binding ratio in 123I-FP-CIT SPECT: accurate processing strategy for cerebral ventricular enlargement with use of 3D-striatal digital brain phantom

被引:1
作者
Furuta A. [1 ,2 ]
Onishi H. [2 ]
Amijima H. [3 ]
机构
[1] Department of Radiological Technology, Hiroshima Citizens Asa Hospital, 1-1, Kabeminami, Asakita-ku, Hiroshima, 731-0293, Hiroshima
[2] Program in Health and Welfare, Graduate School of Comprehensive Scientific Research, Prefectural University of Hiroshima, 1-1, Gakuenmachi, Mihara, 723-0053, Hiroshima
[3] Graduate School of Nursing, Hyogo University of Health Sciences, 1-3-6, Minatojima, Chuo-ku, Kobe, 650-8530, Hyogo
关键词
Cerebrospinal fluid mask processing; Evans Index; Normal pressure hydrocephalus;
D O I
10.1007/s12194-018-0459-0
中图分类号
学科分类号
摘要
This study aimed to evaluate the effect of ventricular enlargement on the specific binding ratio (SBR) and to validate the cerebrospinal fluid (CSF)-Mask algorithm for quantitative SBR assessment of 123I-FP-CIT single-photon emission computed tomography (SPECT) images with the use of a 3D-striatum digital brain (SDB) phantom. Ventricular enlargement was simulated by three-dimensional extensions in a 3D-SDB phantom comprising segments representing the striatum, ventricle, brain parenchyma, and skull bone. The Evans Index (EI) was measured in 3D-SDB phantom images of an enlarged ventricle. Projection data sets were generated from the 3D-SDB phantoms with blurring, scatter, and attenuation. Images were reconstructed using the ordered subset expectation maximization (OSEM) algorithm and corrected for attenuation, scatter, and resolution recovery. We bundled DaTView (Southampton method) with the CSF-Mask processing software for SBR. We assessed SBR with the use of various coefficients (f factor) of the CSF-Mask. Specific binding ratios of 1, 2, 3, 4, and 5 corresponded to SDB phantom simulations with true values. Measured SBRs > 50% that were underestimated with EI increased compared with the true SBR and this trend was outstanding at low SBR. The CSF-Mask improved 20% underestimates and brought the measured SBR closer to the true values at an f factor of 1.0 despite an increase in EI. We connected the linear regression function (y = − 3.53x + 1.95; r = 0.95) with the EI and f factor using root-mean-square error. Processing with CSF-Mask generates accurate quantitative SBR from dopamine transporter SPECT images of patients with ventricular enlargement. © 2018, Japanese Society of Radiological Technology and Japan Society of Medical Physics.
引用
收藏
页码:219 / 227
页数:8
相关论文
共 23 条
[1]  
Gunther I., Hall H., Halldin C., Swahn C.G., Farde L., Sedvall G., [125I] beta-CIT-FE and [125I] beta-CIT-FP are superior to [125I] beta-CIT for dopamine transporter visualization: auto-radiographic evaluation in human brain, Nucl Med Biol, 24, pp. 629-634, (1997)
[2]  
Piggot M., Perry E., Marshall E.F., McKeith I.G., Johnson M., Melrose H.L., Court J.A., Lloyd S., Fairbairn A., Brown A., Thompson P., Perry R.H., Nigrostriatal dopaminergic activities in dementia with Lewy bodies in relation to neuroleptic sensitivity: comparisons with Parkinson’s disease, Soc Bio Psychiatry, 44, pp. 765-774, (1998)
[3]  
Booij J., Tissingh G., Boer G.J., Speelman J.D., Stoof J.C., Janssen A.G., Wolters E.C., van Roven E.A., [123I] FP-CIT SPECT shows a pronounced decline of striatal dopamine transporter labelling in early and advanced Parkinson’s disease, J Neurol Neurosurg Psychiatry, 62, pp. 133-140, (1997)
[4]  
Booij J., Speelman J.D., Horstink M.W., Wolters E.C., The clinical benefit of imaging striatal dopamine transporters with [123I] FP-CIT SPECT in differentiating patients with presynaptic Parkinsonism from those with other forms of Parkinsonism, Eur J Nucl Med, 28, pp. 266-272, (2001)
[5]  
Antonini A., Benti R., De Notaris R., Tesei S., Zecchinelli A., Sacilotto G., Meucci N., Canesi M., Mariani C., Pezzoli G., Gerundini P., 123I-loflupane/SPECT binding to striatal dopamine transporter (DAT) uptake in patients with Parkinson’s disease, multiple system atrophy, and progressive supranuclear palsy, Neurol Sci, 24, pp. 149-150, (2003)
[6]  
Tossici-Bolt L., Hoffmann S.M., Kemp P.M., Mehta R.L., Fleming J.S., Quantification of [<sup>123</sup>I] FP-CIT SPECT brain images: an accurate technique for measurement of the specific binding ratio, Eur J Nucl Med Mol Imaging, 33, pp. 1491-1499, (2006)
[7]  
Fleming J.S., Bolt L., Stratford J.S., Kemp P.M., The specific uptake size index for quantifying radiopharmaceutical uptake, Phys Med Biol, 49, pp. 227-234, (2004)
[8]  
Calvini P., Rodriquez G., Inguglia F., Mignone A., Guerra U.P., Nobili F., The basal ganglia matching tools package for striatal uptake semi-quantification: description and validation, Eur J Nucl Med Mol Imaging, 34, pp. 1240-1253, (2007)
[9]  
Matsuda H., Mizumura S., Soma T., Takemura N., Conversion of brain SPECT images between different collimators and reconstruction processes for analysis using statistical parametric mapping, Nucl Med Commun, 25, pp. 67-74, (2004)
[10]  
Evans W.A., An encephalographic ratio for estimating the size of the cerebral ventricles, Arch Neurol Psychiatry, 47, 6, pp. 931-937, (1942)