Repeatability, reproducibility, and comparison of ocular biometry using a new optical coherence tomography-based system and another device

被引:0
|
作者
Adam Wylęgała
Bartłomiej Bolek
Robert Mazur
Edward Wylęgała
机构
[1] Railway Hospital,Ophthalmology Department
[2] Zabrze Medical University of Silesia,II School of Medicine with the Division of Dentistry
来源
关键词
D O I
暂无
中图分类号
学科分类号
摘要
Precise measurement of axial length before cataract surgery is necessary for the proper lens implantation. We included 100 eyes of 56 patients in the study. The mean age was 41 (15–84 years). Measurements of axial length, anterior chamber depth (ACD) carried out with the new Revo NX were compared with those obtained with the IOLMaster 500. Interoperator testing was performed on 56 eyes of 56 participants. To test repeatability, axial length, ACD, central corneal thickness, and lens thickness were used. Inter-operator reproducibility was also assessed. The results were evaluated using Bland–Altman analyses. The mean ACD measured with the IOLMaster and Revo NX was 3.41 and 3.40 mm p = 0.467, respectively. The interclass correlation was excellent at the level of 0.975. ICC for axial length values was 0.999, and the mean was equal to 23.47 for the new device and 23.48 for IOLMaster. ICC for Inter-operator repeatability was higher than 0.99 in all parameters. Revo NX showed a very high level of repeatability with ICC ranging from 0.9929 for central corneal thickness to 0.9997 for axial length. Both devices showed excellent agreement and differences that are insignificant, which suggests that they can be used interchangeably.
引用
收藏
相关论文
共 50 条
  • [1] Repeatability, reproducibility, and comparison of ocular biometry using a new optical coherence tomography-based system and another device
    Wylegala, Adam
    Bolek, Bartlomiej
    Mazur, Robert
    Wylegala, Edward
    SCIENTIFIC REPORTS, 2020, 10 (01)
  • [2] Repeatability and reproducibility of ocular biometry using a new noncontact optical low-coherence interferometer
    Huang, Jinhai
    Savini, Giacomo
    Wu, Fan
    Yu, Xinxin
    Yang, Jing
    Yu, Ayong
    Yu, Ye
    Wang, Qinmei
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (10): : 2233 - 2241
  • [3] Comparison of ocular biometry and refractive outcomes using two swept-source optical coherence tomography-based biometers
    Park, Hansol
    Yoo, Young-Sik
    Shin, Eunhae
    Song, Won Seok
    Won, Yeokyoung
    Chung, Tae-Young
    Lim, Dong Hui
    PLOS ONE, 2024, 19 (12):
  • [4] Repeatability and reproducibility of optical biometry implemented in a new optical coherence tomographer and comparison with a optical low-coherence reflectometer
    Kanclerz, Piotr
    Hoffer, Kenneth J.
    Rozema, Jos J.
    Przewlocka, Katarzyna
    Savini, Giacomo
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2019, 45 (11): : 1619 - 1624
  • [5] Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography-based biometer versus partial coherence interferometry and optical low-coherence reflectometry
    Kunert, Kathleen S.
    Peter, Monika
    Blum, Marcus
    Haigis, Wolfgang
    Sekundo, Walter
    Schuetze, Juliane
    Bueehren, Tobias
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (01): : 76 - 83
  • [6] Comparison of 2 swept-source optical coherence tomography-based biometry devices
    Fisus, Andreea D.
    Hirnschall, Nino D.
    Findl, Oliver
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2021, 47 (01): : 87 - 92
  • [7] Reproducibility of a Long-Range Swept-Source Optical Coherence Tomography Ocular Biometry System and Comparison with Clinical Biometers
    Grulkowski, Ireneusz
    Liu, Jonathan J.
    Zhang, Jason Y.
    Potsaid, Benjamin
    Jayaraman, Vijaysekhar
    Cable, Alex E.
    Duker, Jay S.
    Fujimoto, James G.
    OPHTHALMOLOGY, 2013, 120 (11) : 2184 - 2190
  • [8] Small Animal Ocular Biometry using Optical Coherence Tomography
    Ruggeri, Marco
    Kocaoglu, Omer
    Uhlhorn, Stephen
    Borja, David
    Urs, Raksha
    Chou, Tsung-Han
    Porciatti, Vittorio
    Parel, Jean-Marie
    Manns, Fabrice
    OPHTHALMIC TECHNOLOGIES XX, 2010, 7550
  • [9] Comparison of a new optical biometry with an optical low-coherence reflectometry for ocular biometry
    Guler, Emre
    Kulak, Ali Ender
    Totan, Yuksel
    Yuvarlak, Armagan
    Hepsen, Ibrahim Feyzi
    CONTACT LENS & ANTERIOR EYE, 2016, 39 (05): : 336 - 341
  • [10] Biometry with a new swept-source optical coherence tomography biometer: Repeatability and agreement with an optical low-coherence reflectometry device
    Kurian, Mathew
    Negalur, Nikhil
    Das, Sudeep
    Puttaiah, Narendra K.
    Haria, Dhawal
    Tejal, S. J.
    Thakkar, Maithil M.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (04): : 577 - 581