The accuracy of implantation is an accepted prognostic factor for the long-term survival of unicompartmental knee prostheses (UKP). We developed a non-image-guided navigation system for UKP implantation without any extramedullary or intramedullary guiding device. The 30 patients operated on with the navigation system (group A) were matched to 30 patients operated on with the conventional technique (group B) using age, sex, body mass index, preoperative coronal mechanical femorotibial angle, and severity of the preoperative degenerative changes. All patients had a complete radiological examination in the first 3 months after the index procedure, with anteroposterior and lateral plain knee radiographs and anteroposterior and lateral long leg radiographs. Coronal femorotibial mechanical angle and both coronal and sagittal orientations of the femoral and tibial components were measured. There were no significant differences in the mean numerical values of all measured angles except for the sagittal orientation of the tibial component, with a significant excessive posterior tibial slope in group B. There was a significant increase in the rate of prostheses implanted in the desired angular range for all criteria except the coronal mechanical femorotibial angle in group A. An optimal implantation with all optimal items was obtained by 18 cases in group A and 6 cases in group B. Navigated implantation of a UKP with the used, non-image-based system improved the accuracy of the radiological implantation without any significant inconvenience and with little change in the conventional operative technique. The only inconvenience was a 20-min longer operative time. This improvement could be related to a longer survival of such implanted prostheses.
引用
收藏
页码:40 / 45
页数:5
相关论文
共 18 条
  • [1] Ahlback S(1968)Osteoarthrosis of the knee. A radiographic investigation Acta Radiol Diagn (Stockh) [Suppl] 277 1-72
  • [2] Argenson P(2001)Prothèses unicompartimentales du genou Aubaniac JM a-undefined
  • [3] Bert G(1998)undefined J Arthroplasty 13 901-undefined
  • [4] Cartier undefined(1996)undefined J Arthroplasty 11 782-undefined
  • [5] Delp undefined(1998)undefined Clin Orthop 354 49-undefined
  • [6] Hernigou undefined(1996)undefined Rev Chir Orthop Reparatrice Appar Mot 82 23-undefined
  • [7] Deschamps undefined(2001)undefined Rev Chir Orthop Reparatrice Appar Mot 87 645-undefined
  • [8] Jenny undefined(1997)undefined Acta Orthop Belg 63 94-undefined
  • [9] Lootvoet undefined(1998)undefined J Bone Joint Surg Br 80 862-undefined
  • [10] Newman undefined(1995)undefined J Arthroplasty 10 615-undefined