Accuracy of the OMRON M4 automatic blood pressure measuring device

被引:0
|
作者
JE Naschitz
L Loewenstein
R Lewis
D Keren
L Gaitini
A Tamir
D Yeshurun
机构
[1] Bnai Zion Medical Center and the Bruce Rappaport Faculty of Medicine,Department of Internal Medicine A
[2] Technion Israel Institute of Technology,Department of Anesthesiology
[3] Bnai Zion Medical Center and the Bruce Rappaport Faculty of Medicine,Department of Epidemiology
[4] Technion Israel Institute of Technology,undefined
[5] Bnai Zion Medical Center and the Bruce Rappaport Faculty of Medicine,undefined
[6] Technion Israel Institute of Technology,undefined
来源
关键词
validation; blood pressure measurement; tilt test;
D O I
暂无
中图分类号
学科分类号
摘要
Despite widespread use of the automatic blood pressure (BP) measuring device OMRON M4, there is little formal validation on its accuracy. A study was conducted to assess the accuracy of the OMRON M4 compared with the true indirect BP measured by mercury sphygmomanometer (MS). A rapid method for the evaluation of automatic blood pressure measurement devices (READ) was applied for this study. The READ is based on numerous BP measurements at rest and during a standardised postural challenge in a small number of subjects who exhibit a wide range of BPs. Blood pressure measurements were done in supine position for 10 min followed by head-up tilt for 30 min and again supine for 10 min. The automatic device (AU) and the MS were connected to one arm-cuff by means of a T connector. A stethoscope with dual sets of ear-pieces was used for duplicate MS measurements (MS1 and MS2). The MS1, MS2 and AU measurements were taken simultaneously in a blinded manner. Three units of the automatic instrument were evaluated. An average of 111 measurements per unit were performed, every BP category being present in ⩾15 MS measurements. The differences between MS1 and MS2 measurements (ΔBP: MS1–MS2) were utilized to assess the consistency of true indirect BP and the differences between AU and MS measurements (ΔBP:AU-MS2) were utilized to assess the accuracy of the AU. The following characteristics of the OMRON M4 were assessed: (1) grade of accuracy, (2) aberration pattern, (3) consistency of the aberration-pattern, and (4) correlation between levels of BP and ΔBP: AU-MS. For MS paired readings, 92–100% of systolic and 99–100% of diastolic readings fell within 5 mm Hg difference range, that is consistent with a British Hypertension Society grade A of both. For AU compared to MS2, 29–64% of systolic and 73–94% of diastolic readings fell within 5 mm Hg and 49–86% of systolic and 86–99% of diastolic readings fell within 10 mm Hg and the devices qualified C, D and C, respectively. All devices exhibited irregular and inconsistent aberration patterns, making the design of correction formulas impractical. In conclusion, the OMRON M4 device did not meet the requirements of the British Hypertension Society and, therefore, cannot be recommended for clinical use.
引用
收藏
页码:423 / 427
页数:4
相关论文
共 50 条
  • [31] Validation of the Omron® M5-I device for blood pressure measurement according to the International validation Protocol
    El Assaad, M
    Topouchian, J
    Labaki, G
    Asmar, R
    JOURNAL OF HYPERTENSION, 2003, 21 : S23 - S23
  • [32] Validation of the Omron M6 (HEM-7001-E) upper arm blood pressure measuring device according to the International Protocol in elderly patients
    Altunkan, Sekip
    Iman, Nevzat
    Altunkan, Erkan
    BLOOD PRESSURE MONITORING, 2008, 13 (02) : 117 - 122
  • [33] Validation of the Omron M6 (HEM-7001-E) upper arm blood pressure measuring device according to the International Protocol in obese adults
    Altunkan, S.
    Iliman, N.
    Kayaturk, N.
    Altunkan, E.
    JOURNAL OF HYPERTENSION, 2007, 25 : S28 - S28
  • [34] AUTOMATIC DEVICE FOR RECORDING BLOOD PRESSURE
    FERNANDE.H
    ROBINSON, R
    AEROSPACE MEDICINE, 1971, 42 (02): : 209 - &
  • [35] THE ACCURACY OF PORTABLE, AUTOMATIC BLOOD-PRESSURE MONITORS FOR MEASURING HEART-RATE
    HILLS, WL
    GODSEN, RN
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 1985, 17 (02): : 251 - 251
  • [36] An underwater blood pressure measuring device
    Sieber, Arne
    Kuch, Benjamin
    L'Abbate, Antonio
    Wagner, Matthias
    Dario, Paolo
    Bedini, Remo
    DIVING AND HYPERBARIC MEDICINE, 2008, 38 (03) : 128 - 134
  • [37] A PNEUMATIC DEVICE FOR MEASURING BLOOD PRESSURE
    KOVALCIK, V
    DURJAK, J
    HORVATH, K
    KRISKA, M
    PHYSIOLOGIA BOHEMOSLOVACA, 1969, 18 (3-4): : 347 - &
  • [38] Accuracy of wearable electronic device compared to manual and automatic methods of blood pressure determination
    Lins, Lindercy Francisco Tome de Souza
    do Nascimento, Ellany Gurgel Cosme
    da Silva Junior, Jose Antonio
    Fernandes, Thales Allyrio Araujo de Medeiros
    de Andrade, Micassio Fernandes
    Andrade, Cleber de Mesquita
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2023, 61 (10) : 2627 - 2636
  • [39] Accuracy of wearable electronic device compared to manual and automatic methods of blood pressure determination
    Lindercy Francisco Tomé de Souza Lins
    Ellany Gurgel Cosme do Nascimento
    José Antonio da Silva Júnior
    Thales Allyrio Araújo de Medeiros Fernandes
    Micássio Fernandes de Andrade
    Cléber de Mesquita Andrade
    Medical & Biological Engineering & Computing, 2023, 61 : 2627 - 2636
  • [40] The suitability of an automated blood pressure measuring device - the Omron HEM-705CP - in a large multicentre study: the ASCOT study
    O'Neill, H
    O'Brien, E
    Stanton, A
    Hedditch, K
    JOURNAL OF HUMAN HYPERTENSION, 2001, 15 (Suppl 1) : S83 - S85