Accuracy of the OMRON M4 automatic blood pressure measuring device

被引:0
|
作者
JE Naschitz
L Loewenstein
R Lewis
D Keren
L Gaitini
A Tamir
D Yeshurun
机构
[1] Bnai Zion Medical Center and the Bruce Rappaport Faculty of Medicine,Department of Internal Medicine A
[2] Technion Israel Institute of Technology,Department of Anesthesiology
[3] Bnai Zion Medical Center and the Bruce Rappaport Faculty of Medicine,Department of Epidemiology
[4] Technion Israel Institute of Technology,undefined
[5] Bnai Zion Medical Center and the Bruce Rappaport Faculty of Medicine,undefined
[6] Technion Israel Institute of Technology,undefined
来源
关键词
validation; blood pressure measurement; tilt test;
D O I
暂无
中图分类号
学科分类号
摘要
Despite widespread use of the automatic blood pressure (BP) measuring device OMRON M4, there is little formal validation on its accuracy. A study was conducted to assess the accuracy of the OMRON M4 compared with the true indirect BP measured by mercury sphygmomanometer (MS). A rapid method for the evaluation of automatic blood pressure measurement devices (READ) was applied for this study. The READ is based on numerous BP measurements at rest and during a standardised postural challenge in a small number of subjects who exhibit a wide range of BPs. Blood pressure measurements were done in supine position for 10 min followed by head-up tilt for 30 min and again supine for 10 min. The automatic device (AU) and the MS were connected to one arm-cuff by means of a T connector. A stethoscope with dual sets of ear-pieces was used for duplicate MS measurements (MS1 and MS2). The MS1, MS2 and AU measurements were taken simultaneously in a blinded manner. Three units of the automatic instrument were evaluated. An average of 111 measurements per unit were performed, every BP category being present in ⩾15 MS measurements. The differences between MS1 and MS2 measurements (ΔBP: MS1–MS2) were utilized to assess the consistency of true indirect BP and the differences between AU and MS measurements (ΔBP:AU-MS2) were utilized to assess the accuracy of the AU. The following characteristics of the OMRON M4 were assessed: (1) grade of accuracy, (2) aberration pattern, (3) consistency of the aberration-pattern, and (4) correlation between levels of BP and ΔBP: AU-MS. For MS paired readings, 92–100% of systolic and 99–100% of diastolic readings fell within 5 mm Hg difference range, that is consistent with a British Hypertension Society grade A of both. For AU compared to MS2, 29–64% of systolic and 73–94% of diastolic readings fell within 5 mm Hg and 49–86% of systolic and 86–99% of diastolic readings fell within 10 mm Hg and the devices qualified C, D and C, respectively. All devices exhibited irregular and inconsistent aberration patterns, making the design of correction formulas impractical. In conclusion, the OMRON M4 device did not meet the requirements of the British Hypertension Society and, therefore, cannot be recommended for clinical use.
引用
收藏
页码:423 / 427
页数:4
相关论文
共 50 条
  • [1] Accuracy of the OMRON M4 automatic blood pressure measuring device
    Naschitz, JE
    Loewenstein, L
    Lewis, R
    Keren, D
    Gaitini, L
    Tamir, A
    Yeshurun, D
    JOURNAL OF HUMAN HYPERTENSION, 2000, 14 (07) : 423 - 427
  • [2] Accuracy of the Omron RX-M, an automated wrist blood pressure measuring device
    Braam, R
    Aslan, B
    Thien, T
    JOURNAL OF HYPERTENSION, 2003, 21 : S22 - S23
  • [3] ACCURACY OF AN AUTOMATIC BLOOD-PRESSURE DEVICE - THE OMRON-HEM403C
    NETEA, RT
    SMITS, P
    THIEN, T
    JOURNAL OF HUMAN HYPERTENSION, 1995, 9 (09) : 783 - 784
  • [4] Accuracy of the Omron RX-M, an automated blood pressure measuring device, measuring blood pressure at the wrist, according to a modified British Hypertension Society protocol
    Braam, RL
    Aslan, B
    Thien, T
    BLOOD PRESSURE MONITORING, 2004, 9 (01) : 25 - 30
  • [5] ACCURACY OF AN OSCILLOMETRIC AUTOMATIC BLOOD-PRESSURE DEVICE - THE OMRON-HEM403C
    OBRIEN, E
    ATKINS, N
    JOURNAL OF HUMAN HYPERTENSION, 1995, 9 (09) : 781 - 782
  • [6] ACCURACY OF AN OSCILLOMETRIC AUTOMATIC BLOOD-PRESSURE DEVICE - THE OMRON HEM403C
    WALMA, EP
    VANDOOREN, C
    VANDERDOES, E
    PRINS, A
    MULDER, P
    HOES, AW
    JOURNAL OF HUMAN HYPERTENSION, 1995, 9 (03) : 169 - 174
  • [7] Validation of the omron M6 upper arm blood pressure measuring device according to the international protocol in adults
    Altunkan, Sekip
    Iliman, Nevzat
    Kayaturk, Nur
    Altunkan, Erkan
    JOURNAL OF HYPERTENSION, 2006, 24 : 213 - 213
  • [8] Evaluation and effects of the Omron 725 CIC device for measuring blood pressure in a hypertension clinic
    Fonseca-Reyes, Salvador
    Cervantes-Munguia, Ramon
    Garcia de Alba-Garcia, Javier
    Parra-Carrillo, Jose Z.
    Lopez-Maldonado, Francisco
    Montes-Casillas, Mayra
    BLOOD PRESSURE MONITORING, 2007, 12 (05) : 321 - 327
  • [9] THE ACCURACY OF THE AUTOMATIC BLOOD-PRESSURE MEASURING DEVICE TONOMED - A COMPARISON OF DIRECT AND INDIRECT MEASUREMENTS
    WEBER, F
    ANLAUF, M
    HERZ KREISLAUF, 1982, 14 (05): : 279 - 283
  • [10] Home blood pressure measurement with an automatic device (OMRON IC): acceptance and comparison with office blood pressure measurement.
    Gomes, MAM
    Plerin, A
    Alavarce, DC
    Mion, D
    AMERICAN JOURNAL OF HYPERTENSION, 1999, 12 (04) : 165A - 165A