摘要
Background: The gamma camera uptake method with Tc-99m-DTPA is simple and less time consuming for the determination of the glomerular filtration rate (GFR). However, its diagnostic accuracy is debated. Gates’ method and predicted creatinine clearance method were compared with plasma clearance method with Tc-99m-DTPA for the measurement of GFR.Materials and Methods: Tc-99m-DTPA renography was performed on 133 patients (69 males and 64 females; age range being 24 to 84 years) with a wide range of renal function. The GFR was determined simultaneously by 3 methods; (1) gamma camera uptake method (modified Gates, Gates); (2) predicted creatinine clearance method (Cockcroft-Gault, CG); (3) single- or two-plasma clearance method (plasma sample clearance method, PSC). The PSC was chosen as a reference.Results: The regression equation of the Gates and the CG against the PSC wasY=11.89 + 1.041X (r=0.790, p<0.001, RMSE=23.55 ml/min/1.73 m2) andY=8.845+0.7899X (r=0.8270, p<0.001, RMSE =16.27 ml/min/1.73 m2), respectively. In comparison with the GFR by PSC, the Gates tended to everestimate the GFR, and contrarily the CG tended to underestimate the GFR.Conclusion: The Gates correlates well with the PSC. However, the Gates is even less precise than the CG. The Gates’ method in Tc-99m-DTPA renography is not suitable for the estimation of GFR in routine practice.
引用
收藏
页码:561 / 565
页数:4
相关论文
共 100 条
[1]  
Shemesh O(1985)Limitations of creatinine as a filtration marker in glomerulopathic patients Kidney Int 28 830-838
[2]  
Golbetz H(1992)Serum creatinine as an index of renal function: new insights into old concepts Clin Chem 38 1933-1953
[3]  
Kriss JP(1999)How to assess glomerular function and damage in humans J Hypertension 17 308-317
[4]  
Myers BP(1997)The search continues—an ideal marker of GFR [editorial] Clin Chem 43 913-914
[5]  
Perrone RD(1999)Consensus report on quality control of quantitative measurements of renal function obtained from the renogram: International consensus committee from the Scientific Committee of Radionuclides in Nephrourology Semin Nucl Med 29 146-159
[6]  
Madias NE(1982)Glomerular filtration rate: Estimation from fractional renal accumulation of AJR 138 565-570
[7]  
Levey AS(1985)Tc-DTPA (stannous) Eur J Nucl Med 10 519-521
[8]  
Rahn KH(1985)Measurement of glomerular filtration rate using Eur J Nucl Med 11 7-12
[9]  
Heidenreich S(1986)Tc-DTPA and the gamma camera: a comparison of methods J Nucl Med 27 1373-1374
[10]  
Bruckner D(1987)Comparative assessment of techniques for estimation of glomerular filtration rate with Tc-99m-DTPA Eur J Nucl Med 12 548-552