Performance of independent dose calculation in helical tomotherapy: implementation of the MCSIM code

被引:0
|
作者
Eric Dzintars
Sotirios Stathakis
Panayiotis Mavroidis
Amir Sadeghi
Nikos Papanikolaou
机构
[1] University of Texas Health Science Center,Department of Radiological Sciences
[2] Karolinska Institute and Stockholm University,Department of Medical Radiation Physics
来源
Australasian Physical & Engineering Sciences in Medicine | 2012年 / 35卷
关键词
Independent dose calculation; Helical tomotherapy; Implementation; MCSIM code;
D O I
暂无
中图分类号
学科分类号
摘要
Currently, a software-based second check dose calculation for helical tomotherapy (HT) is not available. The goal of this study is to evaluate the dose calculation accuracy of the in-house software using EGS4/MCSIM Monte Carlo environment against the treatment planning system calculations. In-house software was used to convert HT treatment plan information into a non-helical format. The MCSIM dose calculation code was evaluated by comparing point dose calculations and dose profiles against those from the HT treatment plan. Fifteen patients, representing five treatment sites, were used in this comparison. Point dose calculations between the HT treatment planning system and the EGS4/MCSIM Monte Carlo environment had percent difference values below 5 % for the majority of this study. Vertical and horizontal planar profiles also had percent difference values below 5 % for the majority of this study. Down sampling was seen to improve speed without much loss of accuracy. EGS4/MCSIM Monte Carlo environment showed good agreement with point dose measurements, compared to the HT treatment plans. Vertical and horizontal profiles also showed good agreement. Significant time saving may be obtained by down-sampling beam projections. The dose calculation accuracy of the in-house software using the MCSIM code against the treatment planning system calculations was evaluated. By comparing point doses and dose profiles, the EGS4/MCSIM Monte Carlo environment was seen to provide an accurate independent dose calculation.
引用
收藏
页码:423 / 438
页数:15
相关论文
共 50 条
  • [41] High-dose pre-operative helical tomotherapy (54 Gy) for retroperitoneal liposarcoma
    Sargos, Paul
    Dejean, Catherine
    de Figueiredo, Benedicte Henriques
    Brouste, Veronique
    Binh Nguyen Bui
    Italiano, Antoine
    Stoeckle, Eberhard
    Kantor, Guy
    RADIATION ONCOLOGY, 2012, 7
  • [42] High-dose Radiotherapy Using Helical Tomotherapy for Vertebral Metastasis: Early Clinical Outcomes and Cord Dose Specification
    Lee, Dong Soo
    Kwak, Yoo Kang
    Jeong, Song Mi
    Song, Jin Ho
    Kang, Young Nam
    Lee, Sang Nam
    Jang, Hong Seok
    Kim, Yeon Sil
    Yoon, Sei Chul
    Choi, Byung Ock
    JAPANESE JOURNAL OF CLINICAL ONCOLOGY, 2013, 43 (06) : 646 - 653
  • [43] Correlation of thickness changes and daily dose detection of delivery analysis (DA) software in helical tomotherapy
    Wijarn, Nattawut
    Watcharawipha, Anirut
    Nobnop, Wannapha
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2024, 23
  • [44] Dose delivery accuracy on helical tomotherapy for 4-dimensional tumor motion - a phantom study
    Holla, Raghavendra
    Khanna, David
    Narayanan, V. K. Sathiya
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2021, 26 (03) : 380 - 388
  • [45] TLD measurement to investigate the bolus effect on skin dose in total body irradiation with helical tomotherapy
    Alveroglu, Gokce Ucar
    Akbas, Canan Koksal
    Celtik, Inci Kingir
    Becerir, Hatice Bilge
    RADIATION PHYSICS AND CHEMISTRY, 2024, 219
  • [46] CT-Myelography for High-Dose Irradiation of Spinal and Paraspinal Tumors with Helical Tomotherapy
    Uhl, Matthias
    Sterzing, Florian
    Habl, Gregor
    Schubert, Kai
    Sroka-Perez, Gabriele
    Debus, Juergen
    Herfarth, Klaus
    STRAHLENTHERAPIE UND ONKOLOGIE, 2011, 187 (07) : 416 - 420
  • [47] Using dose volume histogram (DVH) predictions to improve the plan quality of helical tomotherapy (HT)
    Kang, Zheng
    MEDICAL DOSIMETRY, 2024, 49 (04) : 363 - 371
  • [48] Impact of the spinal cord position uncertainty on the dose received during head and neck helical tomotherapy
    Piotrowski, Tomasz
    Kazmierska, Joanna
    Sokolowski, Adam
    Skorska, Malgorzata
    Jodda, Agata
    Ryczkowski, Adam
    Cholewinski, Witold
    Bak, Bartosz
    JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2013, 57 (04) : 503 - 511
  • [49] Dose accumulation with deformable image registration method using helical tomotherapy images for prostate cancer
    Thongsuk, Warit
    Chitapanarux, Imjai
    Wanwilairat, Somsak
    Nobnop, Wannapha
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2020, 19 (01) : 25 - 29
  • [50] THE DOSIMETRIC EFFECT OF ZIPPER ARTIFACTS ON TOMOTHERAPY ADAPTIVE DOSE CALCULATION-A PHANTOM STUDY
    Geng, Hui
    Yu, Siu-Ki
    Lam, Wai-Wang
    Wong, Wing-Kei Rebecca
    Ho, Yick-Wing
    Liu, Sau-Fan
    MEDICAL DOSIMETRY, 2011, 36 (03) : 306 - 312