Performance of independent dose calculation in helical tomotherapy: implementation of the MCSIM code

被引:0
|
作者
Eric Dzintars
Sotirios Stathakis
Panayiotis Mavroidis
Amir Sadeghi
Nikos Papanikolaou
机构
[1] University of Texas Health Science Center,Department of Radiological Sciences
[2] Karolinska Institute and Stockholm University,Department of Medical Radiation Physics
来源
Australasian Physical & Engineering Sciences in Medicine | 2012年 / 35卷
关键词
Independent dose calculation; Helical tomotherapy; Implementation; MCSIM code;
D O I
暂无
中图分类号
学科分类号
摘要
Currently, a software-based second check dose calculation for helical tomotherapy (HT) is not available. The goal of this study is to evaluate the dose calculation accuracy of the in-house software using EGS4/MCSIM Monte Carlo environment against the treatment planning system calculations. In-house software was used to convert HT treatment plan information into a non-helical format. The MCSIM dose calculation code was evaluated by comparing point dose calculations and dose profiles against those from the HT treatment plan. Fifteen patients, representing five treatment sites, were used in this comparison. Point dose calculations between the HT treatment planning system and the EGS4/MCSIM Monte Carlo environment had percent difference values below 5 % for the majority of this study. Vertical and horizontal planar profiles also had percent difference values below 5 % for the majority of this study. Down sampling was seen to improve speed without much loss of accuracy. EGS4/MCSIM Monte Carlo environment showed good agreement with point dose measurements, compared to the HT treatment plans. Vertical and horizontal profiles also showed good agreement. Significant time saving may be obtained by down-sampling beam projections. The dose calculation accuracy of the in-house software using the MCSIM code against the treatment planning system calculations was evaluated. By comparing point doses and dose profiles, the EGS4/MCSIM Monte Carlo environment was seen to provide an accurate independent dose calculation.
引用
收藏
页码:423 / 438
页数:15
相关论文
共 50 条
  • [1] Performance of independent dose calculation in helical tomotherapy: implementation of the MCSIM code
    Dzintars, Eric
    Stathakis, Sotirios
    Mavroidis, Panayiotis
    Sadeghi, Amir
    Papanikolaou, Nikos
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2012, 35 (04) : 423 - 438
  • [2] Independent calculation of dose distributions for helical tomotherapy using a conventional treatment planning system
    Klueter, Sebastian
    Schubert, Kai
    Lissner, Steffen
    Sterzing, Florian
    Oetzel, Dieter
    Debus, Juergen
    MEDICAL PHYSICS, 2014, 41 (08) : 130 - 141
  • [3] Application of an independent dose calculation software for estimating the impact of inter-fractional setup shifts in Helical Tomotherapy treatments
    Dzintars, Erik
    Papanikolaou, Nikos
    Mavroidis, Panayiotis
    Sadeghi, Amir
    Stathakis, Sotirios
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (06): : 615 - 623
  • [4] Independent quality assurance of a helical tomotherapy machine using the dose magnifying glass
    Wong, J. H. D.
    Hardcastle, N.
    Tome, W. A.
    Bayliss, A.
    Tolakanahalli, R.
    Lerch, M. L. F.
    Petasecca, M.
    Carolan, M.
    Metcalfe, P.
    Rosenfeld, A. B.
    MEDICAL PHYSICS, 2011, 38 (04) : 2256 - 2264
  • [5] Surface dose measurement for helical tomotherapy
    Snir, Jonatan A.
    Mosalaei, Homeira
    Jordan, Kevin
    Yartsev, Slav
    MEDICAL PHYSICS, 2011, 38 (06) : 3104 - 3107
  • [6] Dosimetric verification of dose calculation algorithm in the lung during total marrow irradiation using helical tomotherapy
    Konstanty, Ewelina
    Malicki, Julian
    Lagodowska, Katarzyna
    Kowalik, Anna
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2017, 13 (01) : 33 - 37
  • [7] Dose verification of virtual bolus application for helical tomotherapy
    Nwe, Saw Yu
    Chitapanarux, Imjai
    Nobnop, Wannapha
    MEDICAL DOSIMETRY, 2023, 48 (01) : 16 - 19
  • [8] Clinical implementation of radiosurgery using the Helical TomoTherapy unit
    Saw, Cheng B.
    Gillette, Carol
    Peters, Christopher A.
    Koutcher, Lawrence
    MEDICAL DOSIMETRY, 2018, 43 (03) : 284 - 290
  • [9] Effective clinical applications of Monte Carlo-based independent secondary dose verification software for helical tomotherapy
    Kosaka, Takahiro
    Takatsu, Jun
    Inoue, Tatsuya
    Hara, Naoya
    Mitsuhashi, Taira
    Suzuki, Michimasa
    Shikama, Naoto
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 104 : 112 - 122
  • [10] Reference dosimetry for helical tomotherapy: Practical implementation and a multicenter validation
    De Ost, B.
    Schaeken, B.
    Vynckier, S.
    Sterpin, E.
    Van den Weyngaert, D.
    MEDICAL PHYSICS, 2011, 38 (11) : 6020 - 6026